9.2 Calibration and Verication
The performance should be checked every year and after any
kind of maintenance and repairing.
Required Test Equipment: SpO2 signal simulator.
Notice: The simulator cannot be used to assess the accuracy of
a Pulse Oximeter probe or a Pulse Oximeter.
9.2.1 Control Key Verication
Press Menu key to display the history data.
9.2.2 Sound Verication
a. Set the Pulse Oximeter sound ON.
b. You hear the simulated heart beep sound.
9.2.3 SpO2 & Pulse Rate Measurement Value Verication
a) Connect SpO2 Probe to the SpO2 connector on
the Oximeter.
b) Insert the operator’s nger into the nger sensor, the
SpO
2 measured value of a healthy person should be
from 95% to 99%, and the pulse rate is the same as a
heart rate.
c) If SpO
2 Simulator is available, verify the accuracy of
Oxygen Saturation Value with probes as follows:
Oxygen Saturation Tolerance
96% ±2%
86% ±2%
70% ±3%