SpO2 Low alarm: After nishing
the high alarm setting, press the right
button and you will see on the screen
the low alarm of SpO
2 setting (refer to
Fig. 10). Press the up or down button
to adjust the number. The low alarm
(SpO
2) setting range is 70-99(%).
7.4 PR Alarm Setting:
After SpO2 low alarm setting, press the right button, and you
will see the screen shown as Fig. 11 for setting PR high alarm.
The high alarm (PR) setting range is: 31-254 (bpm).
Press the up or down button to adjust the number. Then
press the right button for setting the low PR alarm (refer to
Fig. 12). The low alarm (PR) setting range is: 30-253(bpm).
7.5 Alarm on/o
After PR alarm setting, press the right button, you will see the
alarm on/o setting screen. Press the up or down button to
set the alarm on or o.