Audible Alarm Indicators:
Audible alarms can be heard in a quiet environment. The
audible alarm has dierent tone pitch and on-o beep
patterns for each alarm priority.
High priority: SpO2 alarm du-du-du ---------- du-du”,
beeps every 8 seconds.
Medium priority: PR alarm du-du-du”. beeps every
8 seconds.
Low priority: Sensor o or nger out du-”, beeps every
20 seconds.
7.2 Character Denition:
: High alarm of SpO2
: Low alarm of SpO2
: High alarm of pulse rate
: Low alarm of pulse rate
: Turn on or turn o the Alarm
7.3 SpO2 Alarm Setting:
SpO2 high alarm: Turn on the Pulse Oximeter and hit
the menu button three times, and you will see the screen
showing high alarm of SpO
2 (refer to Fig. 9). Press the up or
down button to increase or decrease the number. Press the
right button to select other alarms for setting or press menu
button save the alarm setting(s) and enter into another
parameter item. The high alarm (SpO
2) setting range is: 71-
100(%)
Fig. 9