7 Alarm Setting
7.1 Alarm Priority
There are three-level priorities for selection.
High priority: the highest level alarm, indicates the patient is
in a very dangerous situation.
Medium priority: indicates the warning should be paid
attention to.
Low priority: indicates the technical alarm caused by the
device itself.
Alarm of this Oximeter includes technical and physiological alarm.
All three priorities are divided by a built-in module and can
not be changed by a user.
Visual Alarm Indicators:
If the alarm is activated through over limitation of
physiological alarm, corresponding data area is ashing. If
the alarm is activated by more than one physiological alarm,
each parameter will be displayed when ashing.