PULSE OXIMETER INSTRUCTION MANUAL
24
5.1.1
Instrument performance depends on the pulsatile character
of the artery. The display would not be considered reliable
and accurate if the following conditions are present during
measurement.
• Shock or cardiac arrest
• Temperature of the unit
• After the administration of a cardiovascular drug
• Anemia
• Evidence of ventilation-perfusion mismatch
5.1.2
Instrument performance depends on the wavelength
absorption for oxyhemoglobin and deoxyhemoglobin. If
there are substances absorbing the same wavelength, this
would induce false or low SpO
2 values. The following may
aect these values:
• carboxyhemoglobin
• methemoglobin
• methylene blue
• Indigo carmine
5.1.3
Extremely high illumination could aect the SpO2 measurement.
Use a semi-translucent or opaque cover to shield the sensor.