4.2 ID Number Setting
1. After powering on, hit the menu button two times. You
can enter into the ID setting screen.
Press the up or down button to set the ID number and
then press the menu button to conrm your setting.
It will turn to the normal screen if there is no button
pressed for 10 seconds. The setting range of ID number
is: 1-99.
2. In normal screen, press the up button, the digital LED
shows the current ID number. It will automatically turn
back to normal screen.
5 Take A Measurement
SpO2 measuring is a non-invasive technique used to
measure the amount of oxygenated hemoglobin and pulse
rate by measuring the absorption of selected wavelengths
of light. The light emitted by the red infrared light-emitting
diodes passes through the tissue and is converted into
electrical signals by a photodiode.