Dyes introduced into the bloodstream, such as
methylene blue, indocyanine green, indigo carmine,
and uorescein, may inuence the accuracy of the SpO
2
reading.
Any condition that restricts blood ow, such as use of
a blood pressure cu or extremes in systemic vascular
resistance, may cause a failure to determine accurate
pulse rate and SpO
2 readings.
Remove ngernail polish or articial ngernails before
applying SpO
2 sensors. Fingernail polish or articial
ngernails may cause inaccurate SpO
2 readings.
Optical cross-talk can occur when two or more sensors are
located in adjoining areas. It can be eliminated by covering
each site with opaque material. Optical cross-talk may
adversely aect the accuracy of the SpO
2 readings.
Obstructions or dirt on the sensors red light or detector
may cause a sensor failure. Make sure there are no
obstructions and the sensor is clean.
For routine equipment maintenance, please refer to
the service procedures at the associated section as
indicated in the manual.
As to the other concerns for attention, please carefully
look through the specic chapter in enclosed
instructions.
This manual describes all features and options. This
model may not have all accessories mentioned in this
manual.
Federal Law restricts this device to sale by or on the
order of a physician.