Product Manual
Oximeter SerieSOximeter SerieS
PULSE OXIMETER INSTRUCTION MANUAL
10
• Dyes introduced into the bloodstream, such as
methylene blue, indocyanine green, indigo carmine,
and uorescein, may inuence the accuracy of the SpO
2
reading.
• Any condition that restricts blood ow, such as use of
a blood pressure cu or extremes in systemic vascular
resistance, may cause a failure to determine accurate
pulse rate and SpO
2 readings.
• Remove ngernail polish or articial ngernails before
applying SpO
2 sensors. Fingernail polish or articial
ngernails may cause inaccurate SpO
2 readings.
• Optical cross-talk can occur when two or more sensors are
located in adjoining areas. It can be eliminated by covering
each site with opaque material. Optical cross-talk may
adversely aect the accuracy of the SpO
2 readings.
• Obstructions or dirt on the sensor’s red light or detector
may cause a sensor failure. Make sure there are no
obstructions and the sensor is clean.
• For routine equipment maintenance, please refer to
the service procedures at the associated section as
indicated in the manual.
• As to the other concerns for attention, please carefully
look through the specic chapter in enclosed
instructions.
• This manual describes all features and options. This
model may not have all accessories mentioned in this
manual.
• Federal Law restricts this device to sale by or on the
order of a physician.