User's Manual

Page | 16
Alarm
The alarm set-point value is the numeric value at which the alarm will sound. It will
always be in a default unit which the remote device was calibrated to. When the value
on the display exceeds the set-point, the internal 70dB audible alarm will sound.
Min WLL
This is for indication only and shows the minimum Working Load Limit that the device
can display too.
Max WLL
This is for indication only and shows the maximum Working Load Limit that the remote
device should be used to.
Calibration Date
This is the date of the last re-calibration of the remote device.
Overload Count
This is a counter of how many times the remote device has exceeded the Max WLL
value. There is a hysteresis on the overload count of 10%. For example, if the WLL is
12.00T the overload counter will increment by 1 when the displayed load increases
above 12.00T. However, the load will need to drop down to 10.80T (90%) and then
increase above 12.00T again before the counter increments to 2.
The counter cannot be reset without re-calibration.
Firmware
This is for indication only and shows the current version of firmware installed on the
remote device.