Instruction Manual
Thank you for purchasing this Barloworld
Scientific product. To get the best
performance from the equipment, and for
your personal safety, please read these
instructions carefully before use.
This equipment is designed to operate under the
following conditions:-
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuations not exceeding 10%
❖ Overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
If the equipment is not used in the manner
described in this manual the protection provided
by the equipment may be impaired.
Electrical installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the
line supply corresponds to that stated on the
rating label.
Power requirements all models:
Voltage & Wattage
230V, 50Hz,~ , 350W
There is an IEC socket at the rear of the
instrument for connection to the mains supply.
The unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2
pin “Shuko” plug for connection to the mains.
Choose the lead appropriate for your electrical
installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. See the enclosed instruction
sheet for advice on how to carry out this
procedure.
Should the mains lead require replacement a
cable of 1mm
2
of harmonised code H05W-F
connected to an IEC 320 plug should be used.
1
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
The mains lead should be connected to the
instrument BEFORE connection to the mains
supply.
Safety Advice Before Use
❖ Never move or carry the unit when in use or
connected to the mains electricity supply.
❖ Do not mix flammable liquids or use the
equipment in hazardous atmospheres.
❖ In the case of mains interrupting, a fault or
mechanical failure, the unit will continue to
operate on removal of fault.
❖ Mechanical energy can lead to breakage of
glass vessels. Use with care.
General description
CAUTION: Keep fingers clear when
operating door.
This range of Stuart incubators contains three
models each fitted with a moving platform
offering a choice of shaking action.
SI50: Has an orbital shaker with variable speed
between 20 and 250rpm and an orbit of 16mm.
The shaker platform is fitted with an robust mat
and adjustable rubber coated bars to secure the
load.
SI70: Has a 3D gyratory rocker with variable
speed between 10 and 70rpm and an angle of
tilt adjustable from 3-12°. The shaker platform is
fitted with an anti-slip mat.
SI80: Has a simple side to side rocking platform
with variable speed between 10 and 70rpm and
an amplitude adjustable from 1-7°. The shaker
platform is fitted with an anti-slip mat.
The incubator compartment is heated by an
efficient sheathed element. Forced air circulation
and precise digital control gives accurate heat
distribution and a quick response to set
temperatures. For visibility of the contents the
walls of the incubator are made from smoked
acrylic.