User Manual

English
Before Use
Thank you for purchasing this Barloworld
Scientific product. To get the best
performance from the equipment, and for
your own safety, please read these
instructions carefully before use.
If the equipment is not used in the manner
described in this manual the protection
provided may be impaired.
This equipment is designed to operate under the
following conditions:-
For indoor use only
Use in a well ventilated area
Ambient temperature range +5°C to +40°C
Altitude to 2000m
Relative humidity not exceeding 80%
Mains supply fluctuation not exceeding 10%
Overvoltage category II IEC60364-4-443
Pollution degree 2 IEC664
Use with a minimum distance all around of
200mm from walls or other items
Electrical Installation
Before connection please read and
understand this instruction manual and
ensure that the line supply corresponds to
that shown on the rating plate. All models
require a supply rated at: 230V, 50Hz, ~,
single phase.
Power consumption is:
Model Power Model Power
SB300 600W CB300 1200W
SB301 50W CB302 1250W
SB302 650W CB500 2250W
SB500 1500W CR300 900W
SD300 600W CR302 950W
SD500 1500W
THIS INSTRUMENT MUST BE EARTHED
All models are supplied with two mains leads
fitted with IEC plugs for connection to the
instrument. One has a U.K. 3 pin plug and the
other has a 2 pin “Shuko” plug for connection to
the mains. Choose the lead appropriate for your
electrical installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.
IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment’s rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated