Adjust for the desired maintenance set points when the unit is placed in the maintenance mode. The deadband is
xed at 2% of full scale. To enter or leave the maintenance mode, press and hold the for 8 seconds.
ZERO
and
SPAN
(Calibration of Zero and Span) SUB MENUS
The lower display reads
CAL
in this mode.
ZERO
Calibration
Note: For accurate calibration, do not apply any pressure when performing this function.
With the display reading
ZERO
, press the ENTER key. The upper display will blink. Press ENTER again to complete
the zeroing of the instrument or press the
MENU
key to cancel.
SPAN
Calibration
With the display set to
SPAN
, apply full scale pressure to the unit. Press the ENTER key. The upper display will
blink. Press ENTER again to complete the calibration or press the
MENU
key to cancel.