Analog Panel Meter  Instruction Manual 
82 
Tips: 
  Low and high input signals can be any valid values within the 
range of the meter. 
  Observe minimum input span requirements between input 1 
and input 2. 
  Low input should be less than high input signal. 
Error Message (Error) 
An error message indicates that the calibration or scaling process was not 
successful. 
After the error message is displayed, the meter reverts to the input prior 
to the failure during calibration or scaling and to input 1 during internal 
calibration, allowing the appropriate input signal to be applied or 
programmed. 
The error message might be caused by any of the following conditions: 
1.  Input signal is not connected to the proper terminals or it is 
connected backwards. 
2.  Wrong signal selection in 
Setup menu. 
3.  Minimum input span requirements not maintained. 
4.  Input 1 signal inadvertently applied to calibrate input 2. 
Minimum Input Span 
The minimum input span is the minimum difference between input 1 and 
input 2 signals required to complete the calibration or scaling of the meter. 
Input range  Input 1 & input 2 span 
4-20 mA  0.15 mA 
10 VDC 
0.10 VDC 
Meter Copy Function (Copy) 
The Copy function is used to copy (or clone) all the settings from one 
meter to other meters requiring exactly the same setup and 
programming (
i.e. type of input, scaling, decimal pt., filter, bypass, etc.). 
Warning! 
Only the PMA-02 meter copy cable must be used for 
meter-to-meter interfacing. The PMA-02 is a special 
eight-conductor flat cable with two wires swapped. 
Using standard CAT5 or other cable will cause 
damage to both meters. 