Internal calibration menu is part of the Advanced Features menu. 
1.  Press and hold the Menu button for three seconds to access the 
advanced features of the meter. 
2.  Press the Up
 arrow button to scroll to the Internal calibration menu 
(
ICAL) and press Enter. 
3.  The meter displays either current calibration (
C CAL) or voltage 
calibration (
v CAL), according to the input setup. Press Enter to start 
the calibration process. 
Example of Internal Calibration for current input: 
4.  The meter displays low input current message (C lo). Apply the low 
input signal and press Enter. The display flashes for a moment while 
the meter is accepting the low input signal. 
5.  After the display stops flashing, a number is displayed with the 
leftmost digit brighter than the rest. The bright digit is the active digit 
that can be changed by pressing the Up
 arrow button. Press the 
Right
 arrow button to move to the next digit. 
6.  Set the display value to correspond to the input signal being 
calibrated, typically 4.000 mA. 
7.  The display moves to the 
high input calibration (C Hi). Apply the 
high input signal and press Enter. 
8.  Set the display for the high input calibration, in the same way as it 
was set for the low input calibration, typically 20.000 mA. 
The graphic above shows the calibration of the current input. The voltage 
input is calibrated in a similar way. 