Analog Panel Meter  Instruction Manual 
41 
Minimum Input Span 
The minimum input span is the minimum difference between input 1 and 
input 2 signals required to complete the calibration or scaling of the meter. 
Input range  Input 1 & input 2 span 
4-20 mA  0.15 mA 
10 VDC 
0.10 VDC 
Calibrating the Meter with External Source (Cal) 
The meter can be calibrated to display the process variable in 
engineering units by applying the appropriate input signal and following 
the calibration procedure. 
The use of a calibrated signal source is strongly recommended to 
calibrate the meter. 
Warm up the meter for at least 15 minutes 
before performing calibration to ensure 
specified accuracy. 
To scale the meter without a signal source 
refer to Scaling the Meter (
SCALE) page 40. 
T
he display flashes
while 
the inputs