Figure 10: Voltage Input Connections 
The meter is capable of accepting any voltage from -10 VDC to +10 VDC. 
Modbus RTU Serial Communications 
Serial communications connection is made to an RJ45 connector labeled 
M-LINK on Figure 6. For interfacing to the APM, use the PMA-01 for RS-
232, or the PMA-03 for RS-485. The same port is used for interfacing 
with all expansion modules (e.g. external relays, digital I/O). 
Use the PMA-02 meter copy cable for meter-to-meter interfacing for 
cloning purposes (i.e. copying settings from one meter to other meters). 
Relay Connections 
Relay connections are made to two six-terminal connectors labeled 
RELAY1 – RELAY4 on Figure 6. Each relay’s C terminal is common only 
to the normally open (NO) and normally closed (NC) contacts of the 
corresponding relay. The relays’ C terminals should not be confused with 
the COM (common) terminal of the INPUT SIGNAL connector. 
Figure 11: Relay Connections
CNONO NC NC C
RELAY4 RELAY3
4365 21
CNONO NC NC C
RELAY2 RELAY1
4365 21