Manual
Analog Panel Meter  Instruction Manual 
14 
COMPLIANCE INFORMATION 
Safety 
UL & c-UL LISTED 
USA & Canada 
UL 508 Industrial Control Equipment 
UL FILE NUMBER 
E212517 
FRONT PANEL 
UL Type 4X, NEMA 4X, IP65; panel gasket provided 
LOW VOLTAGE 
DIRECTIVE 
EN 61010-1:2001 
Safety requirements for measurement, control, and 
laboratory use 
Electromagnetic Compatibility 
EMISSIONS 
EN 55022:2006/A1:2007 
Class A ITE emissions requirements 
Radiated 
Emissions 
Class A 
AC Mains 
Conducted 
Emissions 
Class A 
IMMUNITY 
EN 61326-1:2006 
Measurement, control, and laboratory equipment 
EN 61000-6-2:2005 
EMC heavy industrial generic immunity standard 
RFI - Amplitude 
Modulated 
80 -1000 MHz 10 V/m 80% AM (1 kHz) 
1.4 - 2.0 GHz 3 V/m 80% AM (1 kHz) 
2.0 - 2.7 GHz 1 V/m 80% AM (1 kHz) 
Electrical Fast 
Transients 
±2kV AC mains, ±1kV other 
Electrostatic 
Discharge 
±4kV contact, ±8kV air 
RFI - Conducted 
10V, 0.15-80 MHz, 1kHz 80% AM 
AC Surge 
±2kV Common, ±1kV Differential 
Surge 
1KV (CM) 
Power-Frequency 
Magnetic Field 
3 A/m 70%V for 0.5 period 
Voltage Dips 
40%V for 5 & 50 periods 
70%V for 25 periods 
Voltage 
Interruptions 
<5%V for 250 periods 