The Company EN ISO 9001:2000 We are an established world force in the design and manufacture of measurement products for industrial process control, flow measurement, gas and liquid analysis and environmental applications. Cert. No. Q 05907 As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Contents Contents 1 Introduction ..................................................................... 2 1.1 System Description .................................................. 2 1.2 PID Control .............................................................. 2 1.3 AX400 Series Analyzer Options ................................ 2 2 Operation ......................................................................... 3 2.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 1 Introduction 1.1 System Description The AX460 single input and AX466 dual input pH/Redox (ORP) analyzers and associated electrode systems have been designed for continuous monitoring and control of pH and Redox (ORP). The electrode system can be standardized to the analyzer using the built-in calibration facility and a single point buffering facility provides easy re-calibration after initial standardization.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2 Operation 2.2.1 Membrane Key Functions 2.1 Powering Up the Analyzer Menu 1 Warning. Ensure all connections are made correctly, especially to the earth stud – see Section 6.3, page 47. Advance to next menu Menu 2 1. Ensure the input sensors are connected correctly. 2. Switch on the power supply to the analyzer.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Use the Menu Key to scroll through the Menus 2 Operation Use the Sidescroll Key to scroll through the Pages within each Menu Section 2.3, Page 6 Section 3.1, Page 12 OPERATING PAGE VIEW SETPOINTS Use the Downscroll Key to scroll through the Frames within each Page Section 3.2, Page 14 Section 3.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Use the Menu Key to scroll through the Menus 2 Operation Use the Sidescroll Key to scroll through the Pages within each Menu Section 5.5, Page 34 CONFIG.ALARMS Use the Downscroll Key to scroll through the Frames within each Page Config. Alarm 1 Config. Alarm 2 Config. Alarm 3 Config. Alarm 4 Config.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3 Operating Page 2.3.1 Single Input pH Measured Values 7.00 24.4 Deg.C pH pH. Temperature. Monitoring pH Measured Millivolts Millivolts. 404 mV Millivolts % Slope and pH Check Value 100.0 % 7.00 pH % slope value. A value between the programmed minimum % slope value (see Set Min Slope – see Section 5.3, page 30) and 105% is displayed. If the value is outside these limits, check the electrode system.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3.2 Dual Input pH 7.00 pH 7.00 pH Measured pH 25.6 24.4 Deg.C Measured Temperature Sensor A. Sensor B. Monitoring pH Deg.C Sensor A. Sensor B. Temperature 404 mV -256 mV Measured Millivolts 100.0 7.00 pH % Slope and pH Check Value – Sensor A Sensor A. Sensor B. Millivolts % % slope value. A value between the programmed minimum % slope value (see Set Min Slope – see Section 5.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3.3 Single Input Redox (ORP) Measured Values 404mV 25.6Deg.C Millivolts. Temperature. Redox (ORP) Offset – Sensor A Displays the offset value for the sensor set in A: Adjust Offset – see Section 4.1.3, page 22. 0mV A: mV Offset VIEW SETPOINTS See Section 3.1, page 12. SENSOR CAL. See Section 4.1, page 19. Wash Function Redox (ORP) 8 A3: Type set to Wash (Section 5.5) – see Section 2.3.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3.4 Dual Input Redox (ORP) Measured Millivolts 404mV -256mV Sensor A. Sensor B. Note. If Probe Type for both Sensor A and Sensor B is set to ORP (Section 5.3), the lower display line shows Dual ORP. Dual Redox Sample Temperature 25.6Deg.C 24.4Deg.C Sensor A. Sensor B. Note. The measured temperature is displayed only if Temp. Sensor is not set to None – see Section 5.3, page 30.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3.5 Dual Input pH and Redox (ORP) Measured pH and Millivolts 7.00 pH -256 mV Sensor A. Sensor B. Note. The Probe Type for Sensors A and B can be set to any combination of pH, Redox or ORP – see Section 5.3, page 30. The display indications change depending on Probe Type settings, e.g. if Sensor A is set to Redox and Sensor B to pH, the lower display shows Redox (ORP)/pH.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 2 Operation 2.3.6 Wash Function Note. The Wash function is available only if A3: Type is set to Wash – see Section 5.5, page 34. ----- Wash Function Manual On Off Off On Wash Function Manual – Wash function off. Lower display line of Operating Page shows WASH INHIBITED. – Wash function controlled automatically. Lower display line of Operating Page shows WASH IN PROGRESS.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views 3 Operator Views 3.1 View Set Points Note. The parameter names and units of measurement displayed in the View Set Points page depend on the Probe Type settings for Sensors A and B – see Section 5.3, page 30. Those shown below are given as examples only. View Set Points This page shows alarm set points.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Note. The menu displayed when pressing the 3 Operator Views key from the Operator View pages depends on analyzer configuration, i.e.: Single Input Analyzers SENSOR CAL. Probe Type set to pH or Probe Type set to Redox or ORP and Enable Cals set to Yes (Section 5.3) – see Section 4.1, page 19. Security Code Probe Type set to Redox or ORP and Enable Cals set to No (Section 5.3) and Alter Sec.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views 3.2 View Outputs Theoretical Analog Output There are up to four analog outputs, each showing information for one sensor. ----- Note. Analog outputs 3 and 4 available only if option board fitted and analog features enabled – see Section 7.3, page 57. VIEW OUTPUTS 12.00mA 50.0% Live current output value being retransmitted.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views 3.4 View Software ----- VIEW SOFTWARE Issue Shows the version number of the operating software. 1.00 AX400/2000 Issue VIEW SOFTWARE VIEW LOGBOOK Option board fitted and analog features enabled (Section 7.3) and Logbook set to On (Section 5.9) – see Section 3.5, page 16. Monitoring pH Redox (ORP) Dual Redox Operating Page (option board not fitted) – see Section 2.3, page 6.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views 3.5 View Logbook Note. The View Logbook function is available only if the option board is fitted and analog features enabled (see Section 7.3, page 57) and Logbook is set to On (see Section 5.9, page 41). The logbook stores data entries for alarm events, sensor errors, power failures and pH calibration information.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views View Logbook Errors Alarms Cals Power Use the ----- and keys to access the Power logbook. Note. If no entries are stored in the Power logbook, the display shows No More Entries. VIEW LOGBOOK Power 1 -----Off 11:09:02 The Power logbook contains up to 2 entries (entry 1 is the most recent), each comprising the power state (On or Off) and the date/time of the occurrence.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 3 Operator Views 3.6 View Clock Note. The View Clock function is available only if the option board is fitted and analog features enabled – see Section 7.3, page 57. ----- VIEW CLOCK Date Shows the current date. ----Date 01:02:04 Time ----Time Shows the current time. 12:00 Monitoring pH Redox (ORP) Dual Redox VIEW CLOCK Operating Page – see Section 2.3, page 6. pH/Redox (ORP) SENSOR CAL.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4 Setup 4.1 Sensor Calibration Note. If Probe Type for either sensor (Sensor A only if single input) is set to Redox or ORP, the sensor can be calibrated only if Enable Cals. for that sensor is set to Yes – see Section 5.3, page 30. Sensor Calibration ----- SENSOR CAL. Sensor Calibration Security Code Note. This frame is displayed only if Alter Cal. Code is not set to zero – see Section 5.8, page 41.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup Set Buffer 1 ----- Set the pH value of the buffer 1 solution – see Appendix A for pH tables. 4.O1pH Set Buffer 1 Set Buffer 2 Set the pH value of the buffer 2 solution. -----9.18pH Set Buffer 2 Note. The solution selected for buffer 2 must be at least 2 ph greater than that selected for buffer 1, e.g. if buffer 1 is set to 7 pH, buffer 2 must be set to at least 9 pH. Set Auto Buffers 20 IM/AX4PH Rev.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4.1.2 Set Up User Defined Buffers (pH Only) Buffer Type set to User (see Section 4.1.1) 15.0 Deg.C 0.00 pH Solution A: Enter point 1 (to 5) Deg.C and Adjust are shown alternately on the upper display line. Using the and keys, adjust the temperature reading (in 5 º increments) to the first of the temperatures on the pH/temperature curve. A: Enter point 1 15.0 Deg.C 3.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4.1.3 Adjust Offset (Redox/ORP Only) Calibrate Sensor A ----- Sensor Cal. A Sensor Cal. B SENSOR CAL. SECURITY CODE CONFIG. DISPLAY A: Adjust Offset A: Buffer Method 0 407mV Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28. Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29. Probe Type set to Redox or ORP (Section 5.3) – continued below.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4.1.4 Automatic, Single- and Two-Point Calibration (pH Only) Sensor A: Buffer Method (pH probes only) Auto Select the type of automatic calibration required: -----1-Pt Auto 1-Pt – Auto 2-Pt – A: Buffer Method 4.00pH 25.0Deg.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup A: Buffer Method set to Auto 2-Pt 9.18pH 20.0Deg.C Calibrate Buffer 2 (Two-Point Calibration only) Immerse Sensor A in the second buffer solution. A: Immerse Buf.2 Press the key to initiate calibration. Note. To abort calibration, press the complete – see previous page. key again at any time before calibration is 128mV The center display line shows the measured sensor output in millivolts.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4.1.5 Manual, Single- and Two-Point Calibration (pH Only) Sensor A: Buffer Method (pH probes only) Man Select the type of manual calibration required: -----1-Pt Man 1-Pt Man 2-Pt A: Buffer Method – – Manual, single-point calibration Manual, two-point calibration Set Buffer Temperature Deg.C (or Deg.F) and Adjust are shown alternately on the center display line. Use the 20.0Deg.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup A: Buffer Method set to Man 2-Pt 9.18pH 20.0Deg.C Calibrate Buffer 2 (Two-Point Calibration only) Immerse Sensor A in the second buffer solution. pH and Adjust are shown alternately on the upper display line. Use the and keys to set the displayed pH value to the temperature-corrected pH value of the chosen solution (see the data sheet provided with the solution). A: Immerse Buf.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 4 Setup 4.1.6 Grab Calibration (pH Only) Sensor A: Buffer Method (pH probes only) Select Grab calibration method. -----Grab A: Buffer Method 9.18pH 25.0Deg.C Adjust Value ph and Adjust are shown alternately on the upper display line. The displayed pH value is the reading sampled by the analyzer as this frame is selected and is held until the display is advanced to the next frame.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5 Programming 5.1 Security Code Note. This frame is displayed only if Alter Sec. Code is not set to zero – see Section 5.8, page 41. 0000 Enter the required code number (between 0000 and 19999), to gain access to the configuration pages. If an incorrect value is entered, access to the configuration pages is prevented and the display reverts to the Operating Page – see Section 2.3, page 6.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.2 Configure Display ----- CONFIG. DISPLAY Set Language Sets the language to be used on all displays. ----- Set Language Language Page Use the ----- and keys to select the required language. English Deutsch Francais Espanol Italiano Set Language Set Temperature Units ----- Set Temp. Units Temperature Units Use the Deg. F Deg. C Off and keys to select the sample temperature display units.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.3 Configure Sensors ----CONFIG. SENSORS Configure Sensor A ----- Config. Sensor A Config. Sensor B Sensor B configuration (dual input analyzers only) is identical to Sensor A configuration. Probe Type ORP Redox pH Select the required probe type. ----- ORP Redox pH A: Probe Type – – – Millivolt display Millivolt display pH display Differential Input Yes No ----- A: Diff.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming A: Probe Type set to pH pH Electrode Type Antim. Glass Select the type of pH electrode used, Glass or Antimony. ----- A: Electrode Temperature Compensation Select Auto to enable the analyzer to compensate automatically for fluctuations in sample temperature. Manual Auto ----- A: Temp. Comp.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming Solution Temperature Compensation Select Yes to enable compensation for solution temperature effects referenced to 25 ºC. Yes No ----- A: Sample Comp. No Yes Sample Coefficient -0.035 If Sample Comp. is set to Yes, enter the temperature coefficient of the sample, in pH/ºC, within the range 0.020 to –0.050 (in –0.001 increments). For ammonia and sodium hydroxide dosed boilers the value is typically –0.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.4 Configure Diagnostics Note. The Configure Diagnostics function is applicable only if Diff. Input for Sensor A and/or Sensor B is set to Yes – see Section 5.3, page 30. ----- CONFIG. DIAGS Configure Sensor A ----- Config. Sensor A Config. Sensor B Sensor B configuration (dual input analyzers only) is identical to Sensor A configuration.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.5 Configure Alarms ----- CONFIG. ALARMS Configure Alarm 1 ----- Config. Alarm 1 Config. Alarm 2 Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and analog features enabled – see Section 7.3, page 57) is identical to Alarm 1 configuration. Alarm 3 can also be configured as a Wash alarm if A3: Type is set to Wash – see following frame.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming Alarm 1 Failsafe ----- Select Yes to enable failsafe action, otherwise select No. Yes No See also Fig. 5.2 to Fig. 5.6 (page 37). A1: Failsafe Alarm 1 Action High Select the alarm action required, High or Low. -----Low See also Fig. 5.2 to Fig. 5.6 (page 37). A1: Action A-B mV Deg.C Deg.F pH Alarm 1 Set Point The Alarm 1 Set Point can be set within the following ranges: 100.0 pH mV Deg.C Deg.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.5.1 Wash Cycle Configuration (Applicable Only to Alarm 3) A3: Type set to Wash Wash Mode ----- Select the wash mode required. Cont. Pulsed Cont. Pulsed Wash Mode – – (continuous) the relay remains energized for the wash duration the relay is switched on and off every second for the duration of the wash, – see Fig. 5.1. Wash Frequency 15 Set the wash frequency required.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming Note. The following examples illustrate High Alarm Actions, i.e. the alarm is activated when the process variable exceeds the defined set point. Low Alarm Actions are the same, except the alarm is activated when the process variable drops below the defined set point.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.6 Configure Outputs ----- CONFIG. OUTPUTS Configure Output 1 ----- Config. Output 1 Config. Output 2 A-B Temp.B Sen.B Temp.A Sen.A Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog features enabled – see Section 7.3, page 57) is identical to Output 1 configuration. Assign Select the sensor and analog output required: ----- AO1: Assign Sen.A Sen.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7.00pH 0.0pH 5 Programming Zero Value pH (or mV or Deg.C or Deg.F or A-B) and Adjust are shown alternately on the center display line. Use the zero value: AO1: Zero Value and keys to adjust the displayed reading to the required Note. The zero value setting plus the minimum differential determines the minimum and maximum values for the span setting, e.g.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.7 Configure Clock Note. The Configure Clock function is available only if the option board is fitted and analog features enabled – see Section 7.3, page 57. ----- CONFIG. CLOCK Set Clock Set the system clock. ----Set Clock? CONFIG. CLOCK CONFIG. SERIAL CONFIG. CONTROL CONFIG. SECURITY Return to main menu. Option board fitted and Serial Communications feature enabled (see Section 7.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.8 Configure Security ----- CONFIG. SECURITY Alter Security Code Set the security code to a value between 0000 and 19999. 0000 Alter Sec. Code Alter Calibration Code Set the sensor calibration access code to a value between 0000 and 19999. 0000 Alter Cal. Code Alter Sec. Code CONFIG. SECURITY Return to main menu. CONFIG. LOGBOOK Option board fitted and analog features enabled (see Section 7.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming 5.10 Test Outputs and Maintenance ----- TEST/MAINTENANCE Test Outputs Displays the output test details for the analog outputs. ----- Note. Outputs 3 and 4 are available only if the option board is fitted and analog features enabled – see Section 7.3, page 57. Test Outputs Test Output 1 frame only is shown; the format of frames for the remaining outputs is identical. Maintenance 4.00mA 20.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 5 Programming Hold Outputs set to Auto. 1Hrs 30Mins Automatic Time If required, set a time period between 1 and 6 hours, in 30 minute increments, for which the outputs are held when Hold Outputs is set to Auto. At the default setting of Off, changes in relay action and analog outputs are inhibited during sensor calibration and released automatically at the end of the procedure.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6 Installation 6.1 Siting Requirements Note. Mount in a location free from excessive vibration, and where the temperature and humidity specification will not be exceeded. Mount away from harmful vapors and/or dripping fluids and ensure that it is suitably protected from direct sunlight, rain, snow and hail.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.2 Mounting 6.2.1 Wall-/Pipe-mount Analyzers Dimensions in mm (in.) Fixing Centers 210 (8.23) R1 192 (7.56) 0( 0.4 ) 96 (3.76) 25 Ø 6. 50 (0 .2 6) 192 (7.56) 150 (5.9) Fixing Centers 175 (6.9) 94 (3.7) Fig. 6.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.2.2 Panel-mount Analyzers Dimensions in mm (in.) 137.50 (5.41) 96 (3.78) 91.60 (3.6) 96 (3.78) 25 (0.98) 92 +0.8 (3.62 +0.03 ) 0 0 Panel Cut-out 92 +0.8 0 (3.62 +0.03 ) 0 5.40 (0.2) Fig. 6.4 Overall Dimensions 1 Cut a hole in the panel (see Fig. 6.4 for dimensions).
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.3 Connections, General Warning. The instrument is not fitted with a switch therefore a disconnecting device such as a switch or circuit breaker conforming to local safety standards must be fitted to the final installation. It must be fitted in close proximity to the instrument within easy reach of the operator and must be marked clearly as the disconnection device for the instrument.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.3.1 Relay Contact Protection and Interference Suppression If the relays are used to switch loads on and off, the relay contacts can become eroded due to arcing. Arcing also generates radio frequency interference (RFI) which can result in analyzer malfunctions and incorrect readings.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.3.2 Cable Entry Knockouts, Wall-/Pipe-mount Analyzer The analyzer is supplied with 7 cable glands, one fitted and six to be fitted, as required, by the user – see Fig. 6.7.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.4 Wall-/Pipe-mount Analyzer Connections 6.4.1 Access to Terminals Release the four captive screws and remove the terminal cover plate Terminal Block C (Option Board) Case Earth (Ground) Stud Terminal Block A Terminal Block B Fig. 6.8 Access to Terminals, Wall-/Pipe-mount Analyzer Note. When refitting the terminal cover plate, tighten the captive screws to a torque of 0.40 Nm (3.5 lbf. in.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.4.2 Connections Analog Output 4 B1 TC Third Lead Common B2 B3 B4 B5 B6 B7 B8 B9 TC Third Lead Common B10 B11 B12 B13 B14 B15 B16 A16 C16 — A15 + Refer to Section 6.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.5 Panel-mount Analyzer Connections 6.5.1 Access to Terminals Terminal Block A Case Earth (Ground) Stud Terminal Block C (Option Board) Terminal Block B Fig. 6.10 Access to Terminals, Panel-mount Analyzers 52 IM/AX4PH Rev.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.5.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.6 pH Sensor Systems Connections 6.6.1 Standard ph Systems Connection – 2867, AP100, AP300, 7650/60, TB5, Non-ABB When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor is OFF.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 6 Installation 6.6.2 Differential ph Systems Connections – Capable of Providing Sensor Diagnostics (AP200, TBX5) When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor is ON. See Section 5.4, page 33 for sensor diagnostics configuration. If diagnostics are not required leave them switched off.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7 Calibration 7 Calibration Note. The analyzer is calibrated by the Company prior to dispatch and the Factory Settings pages are protected by an access code. Routine recalibration is not necessary – high stability components are used in the analyzer's input circuitry and, once calibrated, the Analog-to-Digital converter chip self-compensates for zero and span drift.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7 Calibration 7.3 Factory Settings Use the Sidescroll Key to scroll through the Pages within each Menu Section 7.3, Page 58 FACTORY SETTINGS Use the Menu Key to scroll through the Menus Factory Set Code Use the Downscroll Key to scroll through the Frames within each Page To OPERATING PAGE Cal. Output 2 Cal. Output 3 Cal. Sensor A Cal. Sensor B Cal.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7 Calibration ----- FACTORY SETTINGS Factory Settings Access Code Enter the required code number (between 0000 and 19999) to gain access to the factory settings. If an incorrect value is entered, access to subsequent frames is prevented and the display reverts to the top of the page. 0000 Factory Set Code Calibrate Sensor A Note.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 150.0 Ohms 150.0 Calib Temperature Span (150R) 1000 Ohms 1000 Calib Temperature Zero (1k0) 1500 1500 Calib Temperature Span (1k5) 7 Calibration Set the temperature simulator to 150 The display advances automatically to the next step once a stable and valid value is recorded. A:T.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7 Calibration Calibrate Output 1 Note. When adjusting the 4 and 20 mA outputs, the display reading is unimportant and is used only to indicate that the output is changing when the and keys are pressed. ----- Cal. Output 1 Cal. Output 2 See below. Adjust 4mA Use the 16000 and keys to set the milliammeter reading to 4 mA. Note. The analog output range selected in Configure Outputs (Section 5.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 7 Calibration Option Board Fitted and Analog Features Enabled Calibrate Output 3 Note. ----- Output 3 (and Output 4) calibration is applicable only if the Option board is fitted and analog features enabled – see below. Cal. Output 3 Output 3 calibration is identical to Output 2 calibration. Calibrate Output 4 Note. Output 4 calibration is identical to Output 3 calibration. ----- Cal.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 8 Simple Fault Finding 8 Simple Fault Finding 8.2 Calibration Fail Message or No Response to pH/Redox Changes 8.1 Error Messages The majority of problems are associated with the electrodes and cabling. Replace the electrodes as an initial check – refer to the appropriate instruction manual.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 8 Simple Fault Finding 8.3 Checking the Temperature Input Check the analyzer responds to a temperature input. Disconnect the Pt100/Pt1000/Balco 3K leads and connect a suitable resistance box directly to the analyzer inputs – see Section 6.4, page 50 (wall-/pipe-mount analyzer) or see Section 6.5, page 52 (panel-mount analyzer). Check the analyzer displays the correct values as set on the resistance box – see Table 8.2.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 9 Specification 9 Specification pH/Redox (ORP) – AX460 and AX466 Conductivity – AX416 Only Inputs One or two* pH or mV inputs and solution earth One or two* temperature sensors Range Programmable 0 to 0.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 9 Specification Relay Outputs – On/Off Control Function – AX460 Only Number of relays Three supplied as standard or five with option board fitted Controller Type P, PI, PID (configurable) Number of set points Three supplied as standard or five with option board fitted Set point adjustment Configurable as normal or failsafe high/low or diagnostic alert Control Outputs Output Can be assigned a maximum of two relays, t
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 9 Specification Mechanical Data Environmental Data Wall-/Pipe-mount versions IP65 (not evaluated under UL certification) Operating temperature limits –20 to 55 °C (–4 to 131 °F) Dimensions 192 mm high x 230 mm wide x 94 mm deep (7.56 in. high x 9.06 in. wide x 3.7 in. deep) Storage temperature limits –25 to 75 °C (–13 to 167 °F) Weight 1 kg (2.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Appendix A – Buffer Solutions Appendix A – Buffer Solutions Temp ºF 1.68 pH 4.01 pH 6.86 pH 9.18 pH 0 32 1.666 4.003 6.984 9.464 5 41 1.668 3.999 6.951 9.395 10 50 1.670 3.998 6.923 9.332 15 59 1.672 3.999 6.900 9.276 20 68 1.675 4.002 6.881 9.225 25 77 1.679 4.008 6.865 9.180 30 86 1.683 4.015 6.853 9.139 35 95 1.688 4.024 6.844 9.102 9.18 pH 40 104 1.694 4.
Single and dual input analyzers for pH/Redox (ORP) AX416, AX436, AX460, AX466 & AX468 Temp NIST Buffers Temp US Technical Buffers ºC ºF 4.01 pH 6.86 pH 9.18 pH ºC ºF 4.01 pH 7 pH 10.01 pH 0 32 4.003 6.982 9.460 0 32 4.000 7.118 10.317 5 41 3.998 6.949 9.392 5 41 3.998 7.087 10.245 10 50 3.996 6.921 9.331 10 50 3.997 7.059 10.179 15 59 3.996 6.898 9.276 15 59 3.998 7.036 10.118 20 68 3.999 6.878 9.227 20 68 4.001 7.016 10.062 25 77 4.004 6.
Products and customer support Automation Systems Customer support For the following industries: — Chemical & Pharmaceutical — Food & Beverage — Manufacturing — Metals and Minerals — Oil, Gas & Petrochemical — Pulp and Paper We provide a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre.
ABB Limited Process Automation Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel: +44 1453 826 661 Fax: +44 1453 829 671 ABB Inc. Process Automation 125 E. County Line Road Warminster PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 www.abb.com Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail.