Calibrations
TPS400-5.0.1en
Hz-collimation
V-index
Display adjustment value:
Summarizes saved values.
Procedure:
1. Level instrument with electronic bubble.
2. Aim at a point approximately 100m from the
instrument which is within 5° of the horizontal.
3. [ALL]: Trigger measurement.
4. Change telescope position and aim to the point
again.
For checking the horizontal aiming Hz and V are
displayed.
5. [ALL]: Starts measurement
6. Displays the old and new calculated values.
[SET] Sets new calibration data.
[ESC] Quits program without setting new
calibration data.