SETUP COMMANDS
OUTX (?) {i} The OUTX command sets the output interface to RS232 (i=0) or GPIB
(i=1). The OUTX i command should be sent before any query com-
mands to direct the responses to the interface in use.
OVRM i In general, every GPIB interface command will put the SR830 into the
REMOTE state with the front panel deactivated. To defeat this feature,
use the OVRM 1 command to overide the GPIB remote. In this mode, the
front panel is not locked out when the unit is in the REMOTE state. The
OVRM 0 command returns the unit to normal remote operation.
KCLK (?) {i} The KCLK command sets or queries the key click On (i=1) or Off (i=0)
state.
ALRM (?) {i} The ALRM command sets or queries the alarm On (i=1) or Off (i=0)
state.
SSET i The SSET i command saves the lock-in setup in setting buffer i (1≤i≤9).
The setting buffers are retained when the power is turned off.
RSET i The RSET i command recalls the lock-in setup from setting buffer i
(1≤i≤9). Interface parameters are not changed when a setting buffer is
recalled with the RSET command. If setting i has not been saved prior to
the RSET i command, then an error will result.