Instruction Manual

PART NO. # MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Horizontal (Standard) 7786021
1
75mm tube holders 7713043
1
6 17mm x 75mm 1000 to 4,500 RPM 140 to 2,260 xg
100mm tube holders 7713040
1
6 17mm x 100mm 1000 to 4,500 RPM 140 to 2,800 xg
6-Place Fixed-Angle Rotor 7786047
2
75mm tube holder 7713043
1
6 17mm x 75mm 1000 to 4,000 RPM 140 to 1,500 xg
100mm tube holder 7713040
1
6 17mm x 100mm 1000 to 4,000 RPM 140 to 1,730 xg
125mm tube holder 7713044
2
6 17mm x 125mm 1000 to 4,000 RPM 140 to 1,950 xg
6-Place Performance Plus Horizontal 7786017
3
75mm tube holder 7713056
3
6 13mm x 75mm 1000 to 6,500 RPM 140 to 5,000 xg
12-Place combo Fixed-Angle STAT 7786003
3
75mm tube locations n/a
3
6 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
100mm tube locations n/a
3
6 13mm x 100mm 1000 to 7,000 RPM 140 to 5,700 xg
12-Place Fixed-Angle STAT 7786001
3
75mm tube locations n/a
3
12 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
Overall Dimensions (H x W x D): 8.0 in. x 11.0 in. x 13.5 in.
Centrifuge Motor: 1/2 H.P. Brushless DC
Protection Breaker: 4 Amp. re–settable
Timer: electronic, with hold
or 0 to 99 minutes, +/- 1%
Weight: 23 lbs.
Permitted Environmental Conditions
Ambient Temperature During Operation: 2
o
C - 40
o
C
Maximum Relative Air Humidity: 90%
Electrical Requirements:
Power: 200 Watts
Voltage: North America: 115 Volts (+/- 10%)
Japan: 100 Volts (+/- 10%)
Europe: 230 Volts (+/- 10%)
Frequency: 50/60 Hz
Any use other than those specied by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Specications:
General Specications for the Horizon Model 842VES Centrifuge
page 4
75 mm Tube
Holder
p/n 7713043
1
100 mm
Tube Holder
p/n 7713040
1
125 mm
Tube Holder
p/n 7713044
2
13 x 75mm
Performance Tube
Holder
p/n 7713056
3
Tube Holder
Cap
p/n 7713011
1” Cushion
Spacer
p/n 1525
1
This rotor or rotor accessory is rated for a rotational speed of 5000 RPM and a force of 3000 xg.
2
This rotor or rotor accessory is rated for a rotational speed of 4500 RPM and a force of 2500 xg.
3
This rotor or rotor accessory is rated for a rotational speed of 7000 RPM and a force of 5500 xg.
Refer to serial number label
on rear of centrifuge to
conrm voltage requirement.
NOTE: Contact the Drucker Company for assistance when changing the installed rotor. The installed rotor may
only be changed by qualied personnel.
.5 to 1 ml tube
Adapter
p/n 7713068
13 x 75mm
insert, p/n
7713064
13 x 100 mm
insert, p/n
7713066
1.5 to 2mL
tube Adapter
p/n 7713065