Owner's manual
HEMATOFUGE 842 HS
Overall Dimensions (H x W x D): 20 x 28 x 34 cm
Centrifuge Motor: 1/2 H.P. Brushless DC
Protection Breaker: 4 Amp. re–settable
Timer: electronic, with hold
or 0 to 99 minutes, +/- 1%
Weight: 27 lbs (12.25 kgs)
Permitted Environmental Conditions:
Ambient Temperature During Operation: 2 °C - 50 °C
Maximum Relative Air Humidity: 90%
Electrical Requirements:
Power (Watts): 200
Voltage (Volts): 115 (+/- 10)*
230 (+/- 20)*
*Check Serial Number Badge for Your Model Specic Voltage Requirement
653V 115 Volts for US
653V 230V 230 Volts for Europe
Frequency (Hz): 50/60
SPECIFICATIONS:
General Specications for the Hematofuge Horizon 842HS Centrifuge
page 4
RECOMMENDED SPINS
The following spinning times are recommended:
Serum Samples: 2 minutes 12,000 rpm
Urine Samples: 1 minute 1,000 rpm
The following spins are compulsory:
Hematocrit Samples: 5 minutes 12,000 rpm
QBC Samples: 5 minutes 12,000 rpm