P/N 7711048 HEMATOFUGE 842 HS Operator’s Manual Rev.
TABLE OF CONTENTS Model Description page 3 Supplied Equipment page 3 Intended Use page 3 Warranty Information page 3 Specifications / Recommended spins page 4 Control Panel / Parts of the Centrifuge page 5 Setup Location page 6 Initial Setup Procedure page 6 Operation page 7 Rotor Configurations Sixteen-Place, Microhematocrit/ serum/QBC Rotor page 7 Default user settings page 7 Advanced User Settings page 8 Control Panel Functions Time Adjustment and Timer Operation Spe
WARNING: For the safety of both the operator and service personnel, care should be taken when handling substances that are known to be toxic, radioactive or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II materials are used (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal should be employed. The rotor and rotor accessories should not be considered as Bio-Seals.
HEMATOFUGE 842 HS SPECIFICATIONS: General Specifications for the Hematofuge Horizon 842HS Centrifuge Overall Dimensions (H x W x D): 20 x 28 x 34 cm Centrifuge Motor: 1/2 H.P. Brushless DC Protection Breaker: 4 Amp. re–settable Timer: electronic, with hold or 0 to 99 minutes, +/- 1% Weight: 27 lbs (12.
CONTROL PANEL / PARTS OF THE CENTRIFUGE: Lid Lid Knob Lid Safety Interlock System Control Panel (see below) Cabinet Up/Down RPM Change Buttons RCF Conversion Button Running Indicator Light Locked Indicator Light Unlocked Indicator Light RPM Display Start Button Open / Stop Button Time Display Up/Down Time Change Buttons Brake Display Program Button page 5 Memory Preset Display Memory Button
SETUP LOCATION: 1. Unpack the centrifuge and verify that all of the supplied equipment is present. 2. Choose a setup location which meets the following criteria: a) A bench top clearance height of 40 cm is required in order to open the lid. b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a setup location which will allow for a clearance envelope of at least 61 x 61 cm, (with the centrifuge at the center).
OPERATION: NOTE: Follow the initial setup procedure before initial operation. 1. Press the ‘OPEN / STOP’ button to unlock the lid and then open the centrifuge lid. 2. Turn rotor lid knob counterclockwise, and remove rotor lid. 3. Place the test tube samples into the rotor. Be sure to follow the rules for balanced loads. 4. Put the rotor lid in place and turn the lid knob clockwise until snug, do not over tighten. THE ROTOR LID MUST BE IN PLACE BEFORE OPERATING THE CENTRIFUGE! 5.
CONTROL PANEL FUNCTIONS: Time Adjustment and Timer Operation - The run time may be set from 30 seconds to 99 minutes and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run time. Adjustments may be made prior to a run or after a run begins. A quick tap will adjust the time by 30 seconds. Hold down the button for 1 minute adjustments - continue to hold for 5 minute adjustments. Press the down button once while 00:30 is displayed to access the “hold” feature.
ADVANCED USER SETTINGS (continued): Braking Rate - The user may adjust the braking rate from minimum ( slowest deceleration, 0) to maximum (quickest deceleration, 9) using the following procedure: 1. While the centrifuge is idle, press the PROGRAM button. 2. Use the UP/DOWN arrow buttons next to the speed display until “BRAKE” is displayed in the speed display. 3. The current braking rate is displayed in the brake display.
MEMORY LOCATIONS / STORING AND RECALLING: The Hematofuge 842HS centrifuge is capable of storing up to 10 user-defined presets.These memory presets contain all of the settings needed to define a specific run, (run time, speed, acceleration rate, braking rate, etc.). The user can use these memory locations to quickly configure the centrifuge for a specific test type and ensure that the test is run in the same way each time by recalling this same setting. Store a configuration in a memory location 1. 2. 3.
CARE AND PREVENTATIVE MAINTENANCE: With proper care and maintenance your Horizon centrifuge will provide years of laboratory service. For proper care, the following steps should be taken: 1. Provide Adequate Ventilation: For cooling purposes, the Horizon draws in ambient air through the air intake cover on the top of the lid and exhausts this air in the rear of the base. The centrifuge should be placed on a hard smooth surface for good air circulation. 2.
TROUBLESHOOTING: 1. Problem: Solutions: 2. Problem: Solutions: 3. Problem: Solutions: The rotor does not spin freely. • Make sure that nothing has fallen into the rotor chamber. • If there is nothing obstructing the rotor, the motor may be damaged. Contact your authorised dealer or The Drucker Company for further assistance. There is excessive noise when the machine is running. • Check to see that the load is balanced. • Make sure that nothing has fallen into the rotor chamber.
SAFETY: The Hematofuge 842HS complies with all requirements of CE EN 61000-3-2, -3-3, EN 61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-11, EN 61010-1, -2-020 and UL standard 3101–2–20, Can/CSA C22.2 No. 1010.1 , Can/CSA C22.2 No. 1010.2.20 Horizon Lid Safety Switch: The Horizon lid is secured to the top of the cabinet by a locking knob and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the cabinet opening and prevents the lid from opening.
Contact your authorised dealer or The Drucker Company to order replacement parts or accessories. Replacement Parts: Part No. Description 7751068 7723002 7735052 7717041 3056001 3056003 7751043 7724162 7760004 7760006 7714101 7714103 7712256 7724071 7732018 7786050 7724177 7712234 7732020 Lid Tray Micro-Switch Lid Tray Solenoid Motor, 1/2 H.P. Brushless DC PC Control Board Step Down Transformer (115V) Step Down Transformer (230V) Circuit Breaker Front Panel Label Line Cord (U.K.) Line Cord (U.S.