P/N 7711017 NOTE: The programmable run-time has been preset for 10 Minutes Instructions to change this setting are located on page 7. Model 755V Shown Model 755 - 24-Place, 755V - 24-Place, 755 - 12-Place, 755 Microplate Operator’s / Service Manual Rev.
Table of Contents Model Description Supplied Equip. Available Features Specifications Control Panel Setup Location Initial Setup Procedure Additional Setup Procedure Operation Balanced Loads Tube Holder Configurations Force Chart—12 place Care and Preventative Service Rotor Removal and Installation Safety Emergency Rotor Chamber Entry Troubleshooting Available Accessories Warranty pg. 3 pg. 4 pg. 5 pg. 6-7 pg. 8 pg. 9 pg. 9-10 pg. 10-11 pg. 11 pg. 12 pg. 13 pg. 14 pg. 15 pg. 16 pg. 17 pg. 17 pg. 18 pg.
Supplied Equipment*: The following items come standard with each Horizon 755V-24 Centrifuge: 1. One (1) six-place horizontal rotor 2. Six (6) 100 mm four-place tube carriers 3. One (1) Operator's Manual (not shown) 4. One (1) Tutorial CD (not shown) Optional: • 50mL single-place tube carrier • Carrier cap 1. p/n 7786022 p/n 7713023 p/n 7713037 p/n 7713035 2. * The rotor and rotor accessories are rated for a rotation frequency of 3500 RPM (a force of 2100 xg).
Available Features: • Swing-out horizontal rotor design incorporating a unique test tube holder that produces horizontally separated samples while requiring no additional parts, (optional fixed-angle rotors are also available) • Variable-speed (Model 755V, see speed range below) • Cool-Flow air flow design that prevents overheating of samples • Heavy gauge steel construction for safety and durability • Lid safety switch that prevents the centrifuge from operating unless the lid is closed and latched • Remov
Specifications: 12-Place Speed Range (Model 755V): 755V w/12-Place Horizontal Rotor: 500 to 3,200 (± 100) RPM* 755V w/Fixed-Angle Rotors: 500 to 4,000 (± 100) RPM* Max Speed (Model 755): (This speed is set by the factory, see page 7) with 12-Place Horizontal Rotor 3,600 (± 100) RPM with Fixed-Angle Rotors 3,600 (± 100) RPM Maximum capacity: 12-Place Horizontal Rotor: 180 mL (12 x 15 mL) 20-Place Fixed-Angle Rotor: 300 mL (20 x 15 mL) Overall Dimensions (H x W x D): Centrifuge Motor: Nominal Acceleration Tim
Specifications: Microplate Rotor Speed: Force: Maximum capacity: 1500RPM (±100) RPM 322 (±50) xg 200g Overall Dimensions (H x W x D): Centrifuge Motor: Nominal Acceleration Time: Nominal Deceleration (Breaking) Time: Protection Breaker: Timer (electronic): Current Requirement: 9 in. x 14.5 in. x 17 in. ½ H.P. Brushless DC 65 seconds 50 seconds 4 Amp. re-settable 1 to 30 min ±1% 2.0 A @ 115 VAC or 1.0 A @ 230 VAC 115 or 230 (±10%) Volts AC 50 / 60 Hz 39 lbs.
Control Panel: Lid Safety Switch ‘RUN’ Indicator ‘LATCH’ Indicator Speed Control (755V Only) ‘UNLOCKED’ Indicator ‘START’ Button ‘OPEN/ EMERGENCY STOP’ Button Cycle Counter Power (On/Off) Button ‘RUN’ Indicator Lights up when the machine is in operation, (power is being applied to the motor). ‘LATCH’ Indicator Lights up when the lid has been closed and latched properly.
Setup Location: 1. Unpack the centrifuge and verify that all of the supplied equipment is present. 2. Choose a setup location which meets the following criteria: a) A bench top clearance height of 24” is required in order to open the lid. b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a setup location which will allow for a clearance envelope of at least 30” x 30” (with the centrifuge at the center).
(Continued) 9. 10. 11. 12. 13. 14. 15. 16. 17. If this is a variable speed version, turn the speed control to MAX. Turn the centrifuge on by pushing the ‘START’ button. The ‘RUN’ indicator light will illuminate. The unit will accelerate to its set speed. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there are any loud or unusual sounds, stop the centrifuge by pushing the ’OPEN / EMERGENCY STOP’ button immediately and refer to page 12 on troubleshooting.
2. Change the Run-Time Preset: NOTE: The centrifuge must be plugged in and the lid must be opened to access programming functions! a. Push and hold both the ‘START’ and ’OPEN / EMERGENCY STOP’ buttons for approximately three (3) seconds. The ’LATCH’ indicator light will begin to flash; indicating program mode. Release the two buttons. b. Press the ‘START’ button once for each minute of desired run time. The ‘RUN’ indicator light will flash once for each entered minute. c.
BALANCED LOADS Your centrifuge must contain a balanced load in order to work properly. To ensure that the load is balanced, keep these rules in mind when inserting test tube samples: 1. 2. 3. Opposing tube carriers must be identical. Opposing tube carriers must be empty or loaded with an equal number of equally weighted samples. If an odd number of samples is to be spun, use a water-filled tube to mate with the unpaired one.
Tube Holder Configurations: 12-Place or Fixed Angle only Your Horizon Model 755 centrifuge can spin various test tube sizes ranging from 1.5mL to 15mL, (up to 125mm, (4.9 in.) in length), with the appropriate accessories. Use the following chart and drawing to determine which tube holder and cushion combination should be used with your application. DIRECTIONS: 1. Compare the tube to be spun with the four boxes shown below. 2. Find the box that most closely matches the tube’s length.
Force Chart: Force Chart: Model755 755 Relative Centrifugal ForceForce Chart Chart Model Relative Centrifugal 12 Place Horizontal 20 Place Fixed Angle 24 Place Horizontal 2200 2000 1800 1600 RCF (xg) 1400 1200 1000 800 600 400 200 0 0 400 800 1200 1600 2000 2400 2800 3200 3600 4000 Spee d (RPM) This chart will allow youset to your set your Model centrifuge a desired g-force This chart will allow you to Model 755755 centrifuge to atodesired g-force byby providing the appropriate speed for
Care and Preventative Service: With proper care and maintenance, your Horizon centrifuge will provide years of laboratory service. For proper care, the following steps should be taken: 1. Provide Adequate Ventilation: For cooling purposes, the Horizon draws in ambient air through the rear of the lid and exhausts this air out the rear of the base. Do not block the rear of the centrifuge as this will not allow the machine to properly ventilate itself. 2.
(Continued) 7. Remove Accessories Before Moving: All tube holders, samples, and caps must be removed from the rotor chamber before transporting or storing the centrifuge to prevent damage and injury. Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. See page 15, Step 5, for the recommended cleaning solutions.
The Horizon model 755 complies with all requirements of UL standard 61010A-1, 61010A-2-20; Can/CSA C22.2 No’s 1010.1; 1010.2.20. Safety: Horizon Lid Safety Switch: The Horizon lid is secured to the top of the cabinet by a latching knob and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the cabinet opening and prevents the lid from opening. A mechanical stop positions the pawl and prevents it from rotating completely.
Troubleshooting: 1. Problem: The rotor does not spin freely. Solutions: - Make sure nothing has fallen into the rotor chamber. - If there is nothing obstructing the rotor, contact your authorized dealer or The Drucker Company for further assistance. 2. Problem: Excessive noise when the machine is running. Solutions: - Check to see that the load is balanced. - Make sure that nothing has fallen into the rotor chamber.
Replacement Parts: 755– 24-Place, 755V– 24-Place Part No. 7728052 7751068 7786022 7735016 7717039 7751043 7760002 7714101 7714103 7712263 7713029 7724071 7732018 7732019 7713027 Description Foot, rubber Switch, lid safety Rotor, twenty-four place Motor, 1/2 H.P.
Replacement Parts: 755– 12-Place, 755V– 12-Place Part No. 7728052 7751068 7786024 7786042 7786044 7735016 7717039 7751043 7760002 7714101 7714103 7712263 7713029 7724071 7732018 7732019 7713027 7713032 7713031 7713033 Description Foot, rubber Switch, lid safety Rotor, twelve-place horizontal Rotor, twenty-place fixed-angle Rotor, six-place fixed angle Motor, 1/2 H.P.
Providing Quality Centrifuges since 1932 WARRANTY The Drucker Company warranties that this centrifuge is free from defects in materials and workmanship for 5,000 cycles or 2 years, whichever comes first. Should the centrifuge require warranty or out-of-warranty service please contact: 200 Shadylane Drive • Philipsburg, PA 16866 Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.