Product Manual
.......................................................................................................................
DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000
.......................................................................................................................
6
Cleaning and Maintenance
Unplugtheconcentratorbeforecleaning.Donotuseoilorgrease.
Concentrator
Periodically, with a damp cloth, wipe the outside of the concentrator using a solution of hot
water and dishwashing detergent. If any medical cleaning solutions are used, be sure to
follow the manufacturer’s directions.
Filters
TheFoamIntakelterisawashablelter.Rinseandwashasneeded.
1) Removetheltercover.
2) RemoveboththeFoamIntakelterandtheHEPAlter.
3) SeparatetheFoamIntakelterfromtheHEPAlter.
4) Wash in a solution of warm water and dishwashing detergent.
5) Thoroughly rinse and dry.
6) ReplaceFoamIntakelter,HEPAlter,andFilterCover.
TheHEPAIntakeFilterisareplaceablelter.Replacementis
recommended approximately every 12 months, or as needed.
Please contact your equipment provider for a replacement.
Cannula, Tubing, and Humidifier Bottles
Clean and replace as recommended and directed by the
manufacturer or your equipment provider.
Model, Parts and Accessories
Model Description
CH5000S Pure Oxygen Concentrator, with Oxygen Sensor
CH5000 Pure Oxygen Concentrator, without Oxygen Sensor
Parts Description
PL-2001844 Hooded Swivel Caster 1/each
PL-2001853 HEPAIntakeFilter 1/each
PL-2001854 ReusableFoamIntakeFilter 1/each
PL-3000706 FilterIntakeDoor 1/each
The following humidifier bottle is recommended for use with the Pure Oxygen Concentrator.
Accessories Description
HUM001 HumidierBottle 50/case
HUM-TUB HumidierAdapterTubing 50/case
CON550 RidgedOxygenTubingConnector 50/case
SOFT225 Non-kinkingCozyNasalCannula 25'25/case
SOFT250 Non-kinkingCozyNasalCannula 50'20/case
TUBNK25 Non-kinkingOxygenTubing 25'25/case
TUBNK50 Non-kinkingOxygenTubing 50'20/case
Filter Door
HEPA FoamIntake