Product Manual
.......................................................................................................................
DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000
.......................................................................................................................
.......................................................................................................................
DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000
.......................................................................................................................
5
Setup and Operation
1) Positiontheconcentratorinalocationwhichwillpermitunrestrictedairintake,benear
anelectricaloutlet,andbeinaroominwhichyouspendmostofyourtime.Makesure
the device is at least 6 inches (15 cm) from walls, furniture and/or curtains, is not in a
closed confined space such as a closet, and is at least 5 feet (1.5 meters) from hot
sparkingobjects,aheatsource,oranopename.
2) Verifytheairlter,locatedonthebackofyourunit,iscleanandproperlyinstalled[See
Cleaning and Maintenance on Page 6].
3) After reading and understanding this manual, plug the power cord directly into an
electrical outlet.
Oxygen Cannula Connection without Humidification Bottle
4a) Connect the oxygen cannula to the Oxygen Outlet Barb
[FigureC].SkiptoStep9.
Oxygen Cannula Connection with Humidification Bottle
4b) Connect the Connection Adapter to the Oxygen Outlet Barb
[Figure D]
5) Fill the humidifier bottle with distilled water.
6) Usingthestrap,rmlysecurethehumidierbottletothe
concentrator [Figure E].
7) RuntheConnection Adapter under the concentrator handle
and connect to the humidifier bottle according to the bottle
manufacturer’s instructions [Figure E].
8) Connect the cannula to the humidifier bottle per the bottle
manufacturer’s instructions.
9) Press the Power Switch to the ON position [Figure F]. During
start up, the Green LED Operations Indicator Light will flash.
Full oxygen purity will be reached within 6-10 minutes.
10) ChecktheFlow Meter to ensure, that at eye level, the center
of the ball is in line with the prescribed LPM flow rate
[FigureG].
a. To increase the flow rate, turn the Flow Meter Knob
counter-clockwise.
b. To decrease the flow rate, turn the
Flow Meter Knobclockwise.
11) Checkthatoxygenisowingthroughthe
cannula.
a.RefertotheTroubleShootingGuideon
page 8 if there is no oxygen flow.
12) Put on the cannula as directed by your
equipment provider.
13) Whennotinuse,turnotheconcentrator.
Figure C
Figure D
Figure E
Figure F
FigureG