Product Information and User Instructions Models CH5000S and CH5000 Caution – This device is only to be used by or on the prescription of a physician.
Table of Contents Introduction to Your Pure Oxygen Concentrator Page 2 How Your Pure Oxygen Concentrator Operates Page 2 Descriptions of Parts and Controls Page 3 Warnings and Cautions Page 4 Setup and Operation Page 5 Cleaning and Maintenance Page 6 Models, Parts, and Accessories Page 6 Alarms and Troubleshooting Page 7-8 Specifications Page 9 Separation Distances & Symbol Key Page 10 Warranty Page 11 Intended Use The intended use of the Pure Oxygen
Descriptions of Parts and Controls lement Please be sure the following items are included: • User Manual • Oxygen Concentrator • Connection Adapter for humidifier bottle NORMAL OXYGEN WARNING LOW OXYGEN ! lement Insert 1 NORMAL OXYGEN 1 6 2 3 ALARM SERVICE REQUIRED Insert 2 11 7 8 I/O 5 15 4 ALARM SERVICE REQUIRED 12 13 I/O ONLY REF. 14 Figure A Figure B 9 Pure Oxygen Concentrator CH5000S 120V ~ 60Hz 3.0A Max O2 : 93% +/- 3% Use No Oil Or Grease Inovo, Inc 401 Leonard Blvd N.
Warnings and Cautions • • • • • • • • • • • • • • • • • Under certain circumstances, oxygen therapy can be hazardous. Seeking medical advice before using an oxygen concentrator is advisable. Proper operation of this device requires unobstructed ventilation of the rear filters and ventilation slots. Keep the device a minimum of 6 inches (15 cm) from walls, furniture, curtains and/or any other object which may impede airflow and ventilation slots.
Setup and Operation 1) 2) 3) Position the concentrator in a location which will permit unrestricted air intake, be near an electrical outlet, and be in a room in which you spend most of your time. Make sure the device is at least 6 inches (15 cm) from walls, furniture and/or curtains, is not in a closed confined space such as a closet, and is at least 5 feet (1.5 meters) from hot sparking objects, a heat source, or an open flame.
Cleaning and Maintenance Unplug the concentrator before cleaning. Do not use oil or grease. Concentrator Periodically, with a damp cloth, wipe the outside of the concentrator using a solution of hot water and dishwashing detergent. If any medical cleaning solutions are used, be sure to follow the manufacturer’s directions. Filter Door Filters The Foam Intake filter is a washable filter. Rinse and wash as needed. 1) Remove the filter cover. 2) Remove both the Foam Intake filter and the HEPA filter.
Alarms and Troubleshooting Alarm Table – Oxygen Sensor Model Number CH5000S Light Indication Sounds Cause Blinking Green None Normal Operation - Unit in Start-up Phase Solid Green None Normal Operation Solid Yellow None Oxygen Purity Between 72% and 81%. Device can still be used but service provider should be contacted. Solid Red Solid Audible Tone Oxygen Purity Below 71% and/or Low/No Flow (0 to 0.3 LPM). Service Required - Call Provider.
Basic Troubleshooting Symptom Probable Cause Potential Action A. Unit does not operate. No lights are seen on the front panel when the power switch is “On”. An audible alarm is heard (intermittent beep). 1. Power cord not properly inserted into wall outlet. Check power connection at the wall outlet. 2. Issue with power supplied to wall outlet. Check your home circuit breaker and reset if necessary. Try using a different wall outlet if the situation occurs again. 1. Air filter is blocked.
Specifications Electrical Requirements 120 VAC +/- 10%, 60 HZ Rated Current 3A Power 390 W @ 3 LPM Typical Sound Level 45 dBA Altitude 13,123 ft (4000 m) above sea level - concentration level at 93% Operating Temperature 41°F - 104°F (5°C - 40°C) Operating Humidity 20 - 60% RH, Non-condensing Storage Temperature -13°F to 158°F (-25°C to 70°C) up to 93% RH Oxygen Output Purity 87% to 96% for flows ranging 0.5 to 5 LPM Outlet Pressure 5.5 PSI +/- 0.5 PSI Flow Range 0.
Separation Distances Recommended separation distances between portable and mobile RF Communications equipment and the Pure Oxygen Concentrator The Pure Oxygen Concentrator is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
Warranty The Pure Oxygen Concentrator has been carefully manufactured and inspected in our Lehigh Acres, FL USA facility and is warranted to be free from defects in workmanship and materials. Under this warranty, Drive Medical and its subsidiaries obligation shall be limited to the replacement or repair of any such units or parts that prove, by Drive Medical’s inspection to be defective within three years from original date of purchase.
Drive Medical Design & Manufacturing 99 Seaview Blvd Port Washington, NY 11050 Toll-free 877-224-0946 www.drivemedical.