Product Manual
18 19
b.Set Contraction Time
The contraction time controls the time of stimulation. By pressing parameter
controls, the contraction time can be pre-set. The range is adjustable from 1
second to 30 seconds.
c.Set Relaxation Time
The relaxation time determines the time of relaxation. The stimulation ceases at
setting relaxation time and then re-start in a cycle pattern. The relaxation time of
both channels is changed by pressing the parameter controls. The range of it is
adjustable from 1 second to 45 seconds.
d. Set Ramp Time
This ramp time controlling the time of output current that increase from 0 to the
setting level, and from the setting value to 0. When the ramp time is set, each
contraction may be ramped in order that signals come on and come off gradu-
ally and smoothly. The ramp time is adjustable from 1 to 8 seconds by pressing
the parameter controls.
e. Set Pulse Rate
The pulse rate determines how many electrical impulses are applied through
the skin each second. By pressing the parameter controls, the number of cur-
rent impulses per second (Hz) for both channels can be continually adjusted.
The pulse rate is adjustable from 1 Hz to 150 Hz. Unless otherwise instructed,
turn the pulse rate control to the 70-120 Hz range.
f. Set Pulse Width
The pulse width determines the length of time. Each electrical signal is ap-
plied through the skin, which controls the strength and sensation of the
stimulation. The pulse width is adjustable from 2 to 250 uS. Press the but-
tons until the desired value is reached.
g. Set Timer
The treatment time is adjustable from 1 to 60 minutes.
10.Check/Replace the Battery:
Over time, in order to ensure the functional safety of DIGITAL EMS,
changing the battery is necessary.
1.Make sure that both intensity controls are
switched to off- position.
2.Slide the battery compartment cover and remove.
3.Remove the battery from the compartment.
4.Insert the battery into the compartment. Note the
polarity indicated on the battery and in the
compartment.
5.Replace the battery compartment cover and slide
to close
Chapter 18 : BATTERY INFORMATION
AGF-6X DIGITAL EMS can be used with 6F22 rechargeable battery when
necessary. If you use rechargeable batteries, please follow the instructions.
PDF created with FinePrint pdfFactory trial version http://www.pdffactory.com