Product Manual

Specification
Electrical Requirements
115VAC/60Hz
Power Consumption
75W
Dimensions
4.1” (105mm) High
×5.7” (145mm) Wide
× 7.9” (200mm) Deep
Weight
1.6kg (3.53lbs)
Sound Pressure Level
58dB(A)
Max. Comp. Pressure
30 Psig or greater
Neb. Operating Pressure
10~14 Psig
Liter Flow Range
4 - 8 LPM
Particle Size Less Than
5 microns
Operating Temperature
10 ~ 40 (50 to 104 )
Operating Humidity
Up to 95% non-condensing
Storage/Transport Temperature Range
-20 ~ 60 (-4 to 140 )
Storage/Transport Humidity
Up to 95% non-condensing
(1). Conditions may vary based on altitude above sea level, barometric pressure and
temperature. Specification subjects to change without notice.
(2). This product complies with the following standards: C22.2 No 601.1-M90,
UL Std. No 2601-1 and IEC/EN 60601-1-1.
(3). For electromagnetic compatibility the standards are based as follow: CSA 601.1-2
and IEC/EN 60601-1-2.
(4). Protection against electric shock is classified as Type II.
(5). Degree of protection against electric shock is classified as Type BF.
(6). Mode of operation belongs to continuous mode.
(7). Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
(8). Degree of protection against ingress of water is rated as IPX0.
Warning:
To reduce the risk of infection:
1. The Nebulizer kit with a mark on it should not be reused.
2. To prevent the risk of cross infection, different persons should never use the same
nebulizer kit.
3. If you are using a reusable nebulizer kit, cleaning of the nebulizer is recommended after
each aerosol treatment. Disinfecting is recommended once a day. Since nebulizers vary
in type and construction, we recommend that you follow the cleaning instructions
provided with the nebulizer. If there is no cleaning instruction accompanying with the
nebulizer, follow the recommended cleaning procedures described on page 6.
4. Use only the FDA or CE approved nebulizer kits.
R
3
4
This compressor nebulizer, 3050-2, is designed to efficiently deliver the prescribed
saline solution to treat your respiratory disorders, such as asthma, allergies, and
bronchitis. The SportNeb2 will convert the saline solution into an aerosol mist that
penetrates through your lungs.
Read this manual thoroughly before using nebulizer and save for future reference.
1 mark on it should not be reused.
Introduction
R
8. NEVER block the air openings of the product or place it on a soft surface,
9. Avoid the operation in wet or damp locations.
10. Unplug the unit before filling the nebulizer.
such as a bed or couch, where the air opening may be blocked.