Product Manual
....................................................................................................................
Drive Fish Pediatric Compressor Nebulizer Model 18090-FS and 18091-FS
....................................................................................................................
12
3. Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does
nothavetobewashedbecauseonlylteredairpassesthroughit.
NOTE The Drive Disposable Nebulizer will last at least 15 days and possibly
longer, depending on usage. Proper cleaning will also help extend the life
of the nebulizer. Because it is disposable, we recommend that an extra
nebulizer be kept on hand at all times. The Drive Reusable Nebulizer is
dishwasher safe and may be cleaned and reused for up to one year.
Compressor Cleaning
1. With power switch in the “Off ” position, unplug power cord from wall outlet.
2. Wipe outside of the compressor cabinet with a clean damp cloth every few days
to keep dust-free.
Do not submerge in water; doing so will result in compressor damage.
Filter Change
1. Filter should be changed every 6 months or
sooneriflterdiscolors.
2. Removelterbygraspingitrmlyandpulling
outoftheunit.Discardoldlter.
3. ReplacewithanewDrivelter.Additionallters
should be purchased from your Drive provider.
4. Pushnewlterintoposition.
Donotreuseorwashlterorsubstituteanyothermaterialsuchas
cottonforaDriveair-inletlter.Itwillresultincompressordamage.
Maintenance
AllmaintenancemustbeperformedbyaqualiedDriveproviderorauthorizedservice
center.
Electric shock hazard. Do not remove compressor cabinet. All
disassemblyandmaintenancemustbedonebyaqualiedDrive
provider.
WARNING!
DANGER!
WARNING!
Filter
cover