Product Manual
Owner's((Manual(|(REV(001( (
(
(
8(
DECLARATION - ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer's declaration - Electromagnetic emissions
The Drive Medical #15901C, #15902C is intended for use in the electromagnetic environment
specified below. The customer or the user of this bed should ensure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF Emissions CISPR 11
Group B
The Drive Medical #15901C/15902C must emit
electromagnetic energy in order to perform its
intended function. Nearby electronic
equipment may be affected.
RF Emissions
CISPR 11
Class 1
The Drive Medical #15901C/15902C is suitable
for use in all establishments including domestic
establishments and those directly connected to
the public power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not
Applicable
Voltage fluctuations/
flicker emissions
IEC 6100-3-3
Guidance and manufacturer's declaration: Electromagnetic Immunity
Recommended separation distances between portable and mobile
RF communications equipment and the Drive Medical #15901C & #15902C
The Drive Medical #15901C/15902C is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the Drive Medical #15901C/15902C can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the bed as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
𝑑 = 1,2 𝑃
80 MHz to 800 MHz
𝑑 = 1,2 𝑃
800 MHz to 2,5 GHz
𝑑 = 2,3 𝑃
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
0.1
0.38
0.38
0.73
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) accordable to the transmitter manufacturer.
NOTE I At 80 MHz and 800 MHz the separation distance for the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.