Product Manual

Owner's((Manual(|(REV(001( (
(
(
7(
General information
This medical bed is a class II type B IP54 medical device. The recommended environment for
operation of the bed is listed below:
Ambient temperature: 10°C ~ 40°C (50°F~104°F)
Relative humidity range: 30% ~ 75%
Atmospheric pressure: 86KPa~ 106Kpa
The recommended environment for storage/transportation is listed below:
Ambient temperature: -10°C ~ 50°C (14°F~122°F)
Relative humidity range: 10% ~ 93%
Atmospheric pressure: 86KPa~ 106KPa
NOTE: The information in this document is subject to change without notice.
Only items that have been specified as part of this bed or that have been specified as being
compatible with this bed shall be connected to it.
Caution- This bed frame complies with EMC requirements of IEC 60601-1-2. Radio transmitting
equipment, cell phones or similar electronic devices, used in proximity of the bed, may affect
the beds performance.
There are no known contra-indications for use of this product.
LEFT/RIGHT REFERENCE GRAPHIC
ELECTROMAGNETIC EMMISSION AND IMMUNITY
This MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in the table below.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
WARNING
The use of ACCESSORIES, transducers and cables other than those specified, with the exception
of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as
replacement parts for internal components, may result in increased EMISSION or decreased
IMMUNITY of the EQUIPMENT or SYSTEM.