Item Brochure

DeVilbiss Healthcare 100 DeVilbiss Drive Somerset, PA 15501 USA
800.338.1988 814.443.4881 www.DeVilbissHealthcare.com
Safety Features
To reduce service and maintenance, alarms do not require a battery. Alarms indicate:
— Power failure
— High gas temperature
— High pressure
— Low/high flow
— Low oxygen
— Service required
• Pressure-compensated flow meter to ensure accurate flow display
• Pressure-relief valve and thermal protection on the compressor
• Flame-retardant cabinet
• Double insulated unit and two-prong plug
©2017 DeVilbiss Healthcare LLC 06.17 All Rights Reserved.
DeVilbiss
®
and OSD
®
are registered trademarks of DeVilbiss Healthcare. LT-2328 Rev A
• Convenient top and side handles for easy transport
• Lockable flow meter
• Accommodates up to 50 ft of oxygen tubing plus 7 ft of cannula (15.3 m)
• Durable and reliable design
Unit tested in high temperature and high humidity for extended periods of time
— Proven rotary valve design
• Maintains oxygen purity up to 5,000 ft
• Three-year provider service interval
• Auxiliary Oxygen Port
Performance and Design Features (continued)
Specifications 1025DS
Outlet Pressure 20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Oxygen Percentage 2-10 LPM=87%-96%
Power Consumption 120 vac, 60HZ: 639 watts average
Weight 42 lbs. (19 Kilograms)
Dimensions 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm)
Warranty 3 years
Alarms Audible and visual high pressure, low/high flow, low oxygen, high gas
temperature, power failure, Oxygen Sensing Device (OSD)
DeVilbiss 10-Liter Series Concentrators