Operation Manual
15
Technical data
Operating voltage: 400 V 3Ph 50 Hz
Fuse protection: 3 x 16 A
Power input: 3900 W
Dimensions L/W/H in cm: 210/66/28,5
Weight: approx. 80 kg
Emitter type: UV 800 W
Filter type: Diamond glass
Irradiator according to DIN 5050/EN 60
335-2-27 / UV-Typ 3
Spare parts
If necessary only the spare parts given
by us may be used. Burners and filters
are co-ordinated. Assign for repair work
only an authorized customer service.
Burner Code-No: 20013535
Filter disk Code-No: 20013567
Acryl disk Code-No: 20013555
Timer Code-No: 606600
Fan HDB Code-No: 20013534
Main-fan Code-No: 20011681
Error tracing
One or more burners go out after short
time:
The overtemperature protection of the
respective fan responds.
The fan does not run duly.
The whole equipment precipitates after
short time:
The current control relay of the two main
fans (acryl disk fans) responds.
The ambient temperature is too high.
The fan does not run duly.
english
A further filter disk is in the disk
framework. Loosen the screw of the
attachment sheet metal with
abeklapptem disk framework. The filter
disk can be taken now.
Caution! with a mounting sheet
metal two filter disks are held.
The whole device is without function:
The automatic circuit breaker of the
device initiated.
Open the middle
cover sheet at
the top of the
device.
Examine the
automatic circuit
breaker.
automatic circuit
breaker