Technical data
Intervals before readiness for operation
1st measurement about 6 s
after terminating a result of a previous test in range:
(breath alcohol unit)
0.25 mg/L 120 s
0.50 mg/L 140 s
1.00 mg/L 180 s
> 1.5 mg/L 120 s
after terminating a result of a previous test in range:
(blood alcohol unit)
0.50 ‰ 120 s
1.00 ‰ 140 s
2.00 ‰ 180 s
> 3.00 ‰ 120 s
Intervals for other measuring units selected can be calculated
accordingly (see page 34).
Calibration
Calibration interval depends on required accuracy,
typically: 6 months
Permissible calibration range for unit of measuring value:
mg/L 0.160 to 0.750 mg/L
µg/L 160 to 750 µg/L
g/210L 0.035 to 0.150 g/210L
µg/100 mL 16 to 75 µg/100 mL
µg% 16 to 75 µg%
‰ 0.35 to 1.50 ‰
% 0.035 to 0.150 %
g/L 0.35 to 1.50 g/L
mg/100 mL 35 to 150 mg/100 mL
mg% 35 to 150 mg%
mg/mL 0.35 to 1.50 mg/mL
Standard calibration value when using Dräger Ethanol-Standard
solution for unit of measuring value:
mg/L 0.476 mg/L
µg/L 476 µg/L
g/210L 0.100 g/210L
µg/100 mL 47.6 µg/100 mL
µg% 47.6 µg%
‰1.00 ‰
% 0.100 %
g/L 100 g/L
mg/100 mL 100 mg/100 mL
mg% 100 mg%
The use of these standard calibration values corresponds to a
conversion factor of 2100 between breath and blood alcohol
concentration-values.
Dimensions 142 mm x 75 mm x 36 mm (H x W x D)
Weight approx. 195 g
Electrical supply
Batteries 2 x 1.5 V Type Alkaline
(Mignon, LR6, AA)
approx. 1500 measurements
CE marks Electromagnetic compatibility
(Directive 89/336/EEC)