D M E D I C A L Oxylog 3000 Emergency and transport ventilator MT-153-2001 Instructions for Use Software 1.n This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
How to use these Instructions for Use Preparation Charging the battery ● When the external supply is available: 1 the green lamp » N «, lights up regardless of whether the ventilator is switched on or off. The battery is being charged.
Contents For Your Safety and that of Your Patients 5 Intended Medical Use 7 Operating concept 9 Preparation 15 Checking readiness for operation 25 Operation 31 Fault – Cause – Remedy 53 Care 57 What's what 65 Accessories 71 Service Mode 75 Abbreviations and symbols 83 Technical Data 87 Description 95 Order List 102 Index 104 This "Instructions for Use" manual is published for public use and is only meant for your information.
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
For Your Safety and that of Your Patients For Your Safety and that of Your Patients For Your Safety and that of Your Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 This "Instructions for Use" manual is published for public use and is only meant for your information.
For Your Safety and that of Your Patients For Your Safety and that of Your Patients Strictly follow the Instructions for Use Any use of the apparatus requires full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here. Maintenance The apparatus must be inspected and serviced by trained service personnel every two years. Repair and general overhaul of the apparatus may only be carried out by trained service personnel.
Intended Medical Use Intended Medical Use Intended Medical Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Intended Medical Use Intended Medical Use Oxylog® 3000 is a time-cycled, volume-constant and pressurecontrolled emergency and transport ventilator for patients with a tidal volume from 50 mL upwards. With the following ventilation modes — IPPV/IPPVAssist (CMV/CMVAssist*) Intermittent Positive Pressure Ventilation Controlled and assisted volume-constant ventilation with PEEP for CPPV.
Operating concept Operating concept Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Key for switching on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Ventilation Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Selecting the ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating concept Operating concept Key for switching on/off 1 To switch on, briefly press the »O« key. 1 To switch off, hold down the »O« key for about 3 seconds and 2 confirm the switch-off prompt = press rotary knob.
Operating concept Selecting the ventilation mode 1 Hold down the appropriate key for the ventilation mode for about 3 seconds or 1 press the appropriate key briefly and 2 confirm. The selected ventilation mode will now be activated. 3 The actual ventilation mode is displayed in the top left-hand corner of the screen. For detailed instructions on setting the ventilation modes, see pages 32 onwards.
Operating concept Structure of the screen windows Status and alarm messages window Measured values display window Curves and measured values window Settings and alarms window Information window ➀ ➁ ➃ ➂ 05437171 1 2 3 4 5 ➄ "Values" screen window ➅ ➆ ➇ 05837171 6 Line displaying all the measured values in the current ventilation mode.
Operating concept ● Activate parameter for setting. The active parameter appears light on a dark background. ● Set parameter and confirm. To change to the next page: ● Press »Set. ∆∆ « key. The pages are displayed consecutively. "Alarms" screen window 1 Setting menu for setting alarm limits and alarm parameters. For detailed operating instructions, see "Setting alarm limits" on page 44. Alarms 1/2 : First of two pages in the menu. To change to the next page: ● Press »Alarm ∆∆ « key.
Operating concept Oxylog 3000 2 3 1 06237170 To change to the next page: 1 Press »Curves « key. The pages are displayed consecutively. To select other screen pages: ● Press the appropriate keys, e.g. 2 screen page »Set. ∆∆ « or 3 »Alarms ∆∆ «. To return to the pressure curves or flow curves main page: 1 Press »Curves « key.
Preparation Preparation Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Assemble reusable hose set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Connect disposable hose set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Connecting power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation Preparation Reusable or disposable ventilation hose sets can be used. See Order List on page 102. The Carrier System 3000 (optional) – see page 72 – is available for transport under operating conditions. The onboard equipment holder 3000 (optional) – see page 74 – is available for positioning the Oxylog 3000 in the vehicle.
Preparation When using a bacterial filter or HME (Heat Moisture Exchanger) ● Connect the bacterial filter or HME to the angled connector. Bacterial filters increase the breathing resistance and dead-space volume of the ventilation equipment. 02037170 The flow measurement function on the patient side does not depend on the use of an HME. 1 Connect ventilation hose to socket of breathing valve.
Preparation Connect disposable hose set — instead of the reusable hose set. Do not use disposable hose sets other than those in the order list. The minute volume may be measured incorrectly and the device malfunction if other disposable hose sets are used for ventilation. The service temperature range of the device is limited to –10 oC to 50 oC when using the disposable hose set. 1 Connect blue flow measuring hose to the blue socket, 2 and the transparent flow measuring hose to the other socket.
Preparation Connecting power supply Oxylog 3000 is designed to operate on power supplies with different voltages: Internal supply — with rechargeable battery (specified Smart Battery, see "Technical Data", page 88) Additional external power supply To recharge the battery and to extend the electrical operation time when using a rechargeable battery. — DC voltage from the on-board power supply via DC/DC converter or — with AC/DC power pack.
Preparation Charging the battery ● The ambient temperature must be between 0 and 35 oC when charging the batteries! Indicators » N « and » J « remain off while the ventilator is being operated from the internal battery. Oxylog 3000 2 1 02437170 When the external supply is available: 1 the green lamp » N «, lights up regardless of whether the ventilator is switched on or off. The battery is being charged.
Preparation External power supply with DC/DC converter The DC/DC converter should be used to connect the Oxylog 3000 to on-board supplies of different voltages (12 V, 24 V, 28 V DC). The voltage of the on-board supply may fluctuate, depending on the amount of power required for various purposes with the result that the supply voltage falls below or exceeds the range permitted for the Oxylog 3000. The on-board voltage is converted into a constant DC voltage of approx.
Preparation Connecting gas supply Take care when handling O2: ● Secure O2 cylinders so they cannot fall over and keep away from excessive heat – risk of explosion! ● Do not grease or lubricate O2 fittings, such as cylinder valves and pressure reducers, and do not handle with greasy hands – fire risk! ● Only open or close cylinder valves by hand and rotate smoothly. Do not use tools. ● No smoking and no naked lights.
Preparation Determining the approximate pneumatic operating time for Oxylog 3000 Example for supply of medical gas: Cylinder pressure measured on the pressure gauge of the pressure reducer: 200 bar Liquid capacity of the O2 cylinder: 2.5 L Supply of medical gas: 2.5 L x 200 bar = approx. 500 L Example for pneumatic operation time: IPPV mode, frequency 10 1/min, V T = 1 L, O2 = 100 % Minute volume = 10 1/min x 1 L = 10 L/min Operation time = Medical gas supply [L] (MV +0.
Before using for the first time Supply from a piped medical gas system 1 Screw O2 medical gas hose into Oxylog 3000 and 2 plug gas probe into O2 terminal unit until it has engaged twice and the supply of O2 is assured. 1 02637170 2 Before using for the first time ● Ensure that batteries are fully charged, page 20. 24 This "Instructions for Use" manual is published for public use and is only meant for your information.
Checking readiness for operation Checking readiness for operation Checking readiness for operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Connecting test lung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Perform device check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Error messages during device check . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Checking readiness for operation Checking readiness for operation — whenever the ventilator has been prepared or the ventilation hoses changed — at the latest every six months. The following functions are checked with the menu-based test: — Gas supply present — Hose system / breathing valve connected and OK — Alarm functions OK — Ventilation functions OK — Monitor functions OK. Oxylog 3000 interrupt the test if a fault is detected. The relative fault is indicated on the screen.
11937171 Checking readiness for operation ● Select »Device check« in main menu and confirm. 04137171 The device check can be ended at any time by pressing the »Alarm Reset« key. 04237171 ● Ensure that the gas supply has been connected. ● Ensure that the correct gas type (O2 or medical air) has been set and confirm. If the wrong gas type has inadvertently been set: ● Press »Alarm Reset« key to cancel device check.
Checking readiness for operation ● Ensure that the test lung has been connected. 04337171 Oxylog 3000 automatically checks whether a test lung has been connected. The device check is aborted if a test lung is not detected within one minute. The check is continued when Oxylog 3000 detects the test lung. ● Ensure that the configured hose system has been 04537171 connected – either: — the disposable hose set or — the reusable hose set and confirm.
Checking readiness for operation Oxylog 3000 successively activates the acoustic and visual alarm signals and prompts the operator to acknowledge each signal. 04937171 ● Confirm acoustic and visual alarm signals. The device check proceeds automatically. 05037171 During the automatic test sequence, Oxylog 3000 checks the flow, pressure build-up and alarm signals. Corresponding sounds are heard. The bar graph shows the progress made by the check.
Checking readiness for operation Error messages during device check Message Cause Remedy No communication control- / charge-board Device defective. Call DrägerService. System leakage Leak in ventilation hose and/or test lung. Check hoses, breathing valve, flow sensor and test lung for leaks and replace if necessary. Internal leak in system. Call DrägerService. No testlung Test lung not connected or major leakage. Connect test lung.
Operation Operation Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Starting operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Preparing ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 IPPV (CMV), IPPVAssist (CMVAssist) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation Operation Check readiness for operation, page 26. Starting operation 4 2 3 1 00437170 Switching on 1 Briefly press the »O« key. Oxylog 3000 carries out the self-test. ● Wait for the 5 second self-test to be completed. Upon expiry of the self-test, the ventilator automatically starts ventilation with the default settings. Manufacturer's default settings: — Ventilation mode IPPV (CMV) — Ventilation time ratio I:E = 1:1.
Operation IPPV (CMV), IPPVAssist (CMVAssist) IPPV – Intermittent Positive Pressure Ventilation Volume-controlled ventilation with fixed mandatory minute volume MV, set with tidal volume V T and frequency Freq. For patients without spontaneous breathing, see details on page 96 onwards. Paw Plateau time Tplat Pmax Pplat PEEP t IPPVAssist – Intermittent Positive Pressure Ventilation Assisted For patients with partial spontaneous breathing. For synchronisation with the patient's spontaneous breathing.
Operation IPPV (CMV) can be extended to include the trigger function IPPVAssist (CMVAssist): Trigger (IPPVAssist) For synchronisation with the patient's spontaneous breathing efforts. The mandatory ventilation strokes are synchronised with the patient's spontaneous breathing efforts when the trigger is activated and the trigger sensitivity set. The actual frequency may be higher than the set ventilation frequency Freq. in this case.
Operation 12337171 For heart-lung resuscitation The airway pressure Paw is limited to the set Pmax value by Oxylog 3000 without ending inspiration prematurely (pressure-limited, inconstant-volume ventilation when Pmax is reached). Pmax should be set to maximum in order to apply the maximum possible minute volume. Setting alarm limits, page 44. This "Instructions for Use" manual is published for public use and is only meant for your information.
Operation SIMV, SIMV/ASB (SIMV/PS)* Synchronised Intermittent Mandatory Ventilation Assisted Spontaneous Breathing Fixed mandatory minute volume MV set with tidal volume VT and ventilation frequency Freq. The patient can breathe spontaneously between the mandatory ventilation strokes and thus contribute to the total minute volume. Spontaneous breathing can be assisted with ASB.
Operation BIPAP (PCV+)*, BIPAP/ASB (PCV+/PS)* Biphasic Positive Airway Pressure Assisted Spontaneous Breathing Pressure-controlled ventilation combined with spontaneous breathing throughout the breathing cycle and variable pressure support at CPAP level. The mandatory portion of the total minute volume MV is set via the inspiratory pressure Pinsp, PEEP and ventilation frequency Freq. Used for patients without spontaneous breathing, to spontaneously breathing patients shortly before extubation.
Operation CPAP, CPAP/ASB (CPAP/PS)* Continuous Positive Airway Pressure Assisted Spontaneous Breathing For patients with adequate spontaneous breathing. Spontaneous breathing at an elevated pressure level to increase the functional residual capacity FRC. Spontaneous breathing can be assisted with ASB.
Operation Apnoea ventilation For automatically switching over to volume-controlled mandatory ventilation (SIMV) in the event of an apnoea – only effective in ventilation mode CPAP. When an apnoea occurs, the device simultaneously outputs an alarm signal and starts volume-controlled mandatory ventilation with the parameters frequency »fApnoea«, tidal volume »VTApnoea« and the maximum airway pressure »Pmax«. upon expiry of the set alarm time "TApnoea". The ventilation time ratio I:E is invariably set to 1:1.
Operation To end apnoea ventilation: ● Press the »Alarm Reset« key. The ventilator continues operation with the original ventilation mode and with the original ventilation parameters set (CPAP). The manufacturer's settings fApnoea = 12 1/min and VTApnoea = 500 ml can be changed in "Customer Service Mode", see page 77. Apnoea ventilation can only be activated in ventilation mode CPAP without NIV. Apnoea ventilation is not available in any of the other pressure-controlled ventilation forms.
Operation NIV – Non-invasive ventilation Mask ventilation NIV can only be activated as a supplementary function in the pressure-controlled ventilation modes BIPAP (PCV+), BIPAP/ASB (PCV+/PS), CPAP, CPAP/ASB (CPAP/PS). Mask leakages are detected by the device, compensated and included in the displayed flow curve and measured values for V T and MV. The measured values for V T and MV will be incorrect during ventilation with leakages without active NIV function.
Operation O2 concentration without optional "O2 blending" 1 The O2 concentration can be set to 60 % (comparable to "Air Mix" on Oxylog 2000) or 100 % (comparable to "No Air Mix" on Oxylog 2000). Other settings are not possible. Oxylog 3000 1 100 O2 [%] 08637170 60 O2 concentration with optional "O2 blending" (40 % to 100 %*) 60 40 This is a calculated value and not based on one which has been measured by an O2 sensor for the inspiratory O2 concentration FiO2 ! 42 O2 blending is optional.
Operation When patients who are spontaneously breathing, the achievable O2 concentration will depend on the profile of the inspiratory flow. Even if this profile is changed, the message "Check settings O2" may appear after some time. In toxic surroundings: ● the patient must be ventilated with 100 % O2 in order to ensure that toxic constituents are not entrained into the breathing gas.
Operation Setting alarm limits Upper alarm limit for Paw Pressure limitation with Pmax Regardless of the set ventilation mode, the airway pressure is controlled by the ventilator and limited to the set maximum inspiratory pressure Pmax. Pmax appears in the pressure curve as a dashed line. When this dashed line is reached, Oxylog 3000 outputs a "!!! Paw high" alarm. The volumecontrolled stroke cannot be applied completely (ventilation with inconstant volume).
Operation In the Event of an Alarm Oxylog 3000 3 1 2 00737170 1 the red lamp flashes or 2 the yellow lamp flashes. 3 The alarm message appears on the right of the top line on the screen. Oxylog 3000 assigns corresponding priority to the alarm message, highlights the text with the appropriate number of exclamation marks and generates different tone sequences for the respective alarms. !!! = Warning !! = Caution ! = Advisory Warning An alarm with top priority 1 Red alarm lamp flashes.
Operation Advisory Low-priority alarm. 1 Yellow alarm lamp lights up. Advisory messages are identified by one exclamation mark. Example: ! Settings not confirmed Oxylog 3000 generates a two-tone alarm sequence which only sounds once. ● Refer to the list "Fault – Cause – Remedy" on page 53 for Oxylog 3000 3 1 2 information on how to remedy the faults. 01037170 When the fault has been remedied the alarm tone is cancelled.
Operation Displaying curves and measured values The main page displays the airway pressure curve Paw(t) or flow curve Flow(t) and two relevant measured values. To display a different curve 1 Press »Curves « key. Oxylog 3000 10637170 2 1 13137171 Example: airway pressure curve Paw(t) Example: flow curve (t) Displaying other measured values 2 Press »Values ∆∆ « key: the next page is displayed on the device. 13237171 The following pairs of measured values are displayed in the default setting: 1.
Operation Special functions Manual inspiration / Inspiration hold Inspiration Hold for volume-controlled strokes: Regardless of the time at which it is started, an automatic ventilation stroke can be extended for up to max. 15 seconds (in IPPV, IPPVAssist, SIMV, SIMVASB). Or manual inspiration: A ventilation stroke can be started manually between two automatic strokes and held for up to max. 15 seconds. The pattern of the manually started ventilation stroke corresponds with the set ventilation mode.
Operation O2 inhalation (optional) The O2 inhalation function is not a ventilation mode! ● It may only be used for patients with spontaneous breathing who receive a constant O2 flow of between 0 and 15 L/min via a mask. If stenosis occurs, the flow is interrupted by the ventilator for 500 ms at an airway pressure of 30 mbar and the airway pressure is reduced to 0 mbar. The "!!! Paw high" alarm is active. ● The spontaneously breathing patient may only be connected to the device via an inhalation mask.
Operation Calibration The pressure sensors and flow measurements are automatically calibrated by the device at regular intervals. The saved calibration values are retained even when the device is switched off. Screen brightness 13337171 The screen brightness levels can be defined independently of the ventilation mode for battery and mains operation on the last page of the "Setting" menu: — The brightness setting is active in both mains and battery operation when making settings on the ventilator.
Operation Displaying configuration and information The following settings can be made for the application concerned via »Configuration and information«: — Select langugage — Select gas supply (O2 or medical air) The settings made under "Configuration" are retained even when the ventilator is switched off. The following ventilator data can be displayed via »Configuration and information«: — Identification No.
Set gas supply ● Press key »Set. ∆∆ « to select the menu »Configuration and information 1/2«. ● Select and activate line »Gas supply«. O2 or medical air can be set. ● Set and confirm the required gas supply. When medical air is selected, Oxylog 3000 will deliver a concentration of 21 % regardless of the set O2 concentration. However, the control »O2« can be set to 40 %, for example. In this way, Oxylog 3000 will use less medical air by additionally drawing in ambient air.
Fault – Cause – Remedy Fault – Cause – Remedy Fault – cause – remedy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Messages in the Alarms window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Messages in the information window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 This "Instructions for Use" manual is published for public use and is only meant for your information.
Fault – cause – remedy Fault – cause – remedy Oxyog 3000 classifies error messages according to three priority levels and identifies these accordingly with the aid of exclamation marks: !!! Warning = Message with top priority !! Caution = Message with medium priority ! Advisory = Message with low priority The messages are listed in alphabetical order. The following list is intended to assist in identifying and rectifying the underlying cause of any faults triggering an alarm.
Fault – cause – remedy Message Cause Remedy !! High frequency Patient breathes at a high spontaneous Check patient's condition, check rate. ventilation pattern, correct alarm limit fspn if necessary. !! Int. battery charging inop Technical defect. Call DrägerService – only restricted operation is now possible. !!! Int. battery discharged The operating time for operation with the internal battery has expired and an external DC supply has not been connected.
Fault – cause – remedy Message !!! Paw low !! Paw measurement inop Cause Remedy No pressure difference >5 mbar between inspiration and expiration or set pressure level is not achieved. Leak in cuff. Inflate cuff and check for leaks. Leakage or disconnection. Check hose system for leaking connections. Ensure that the breathing valve has been fitted correctly. Fault in measurement hoses for flow measurement on patient side.
Care Care Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Disassemble reusable ventilation set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Remove disposable hose set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Cleaning and disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Care Care ● Clean breathing valve, flow sensor, angled connector and ventilation hoses of the reusable hose set whenever they have been used. ● The disposable hose set must always be disposed of correctly after use. ● Clean ventilator and medical gas hoses if heavily soiled. Disassemble reusable ventilation set 1 Disconnect ventilation hose from socket. 2 Disconnect flow measuring hoses from sockets. 3 Unscrew medical gas hose from Oxylog 3000.
Care 1 Disconnect flow sensor from breathing valve – do not twist or use force on the hose nozzles, as this can damage the flow sensor. 4 2 Carefully detach flow measuring hoses from flow sensor, pulling in the axial direction of the hose nozzles. 3 Detach angled connector from flow sensor. 1 ● Do not allow any objects to enter the flow sensor. Do not purge with compressed air. The wind vane inside may be damaged and cause measuring errors! 4 Detach ventilation hose from breathing valve.
Care Remove disposable hose set 1 Disconnect flow measuring hoses. 2 Disconnect ventilation hose. ● Correctly dispose of the complete disposable hose set. ● The disposable ventilation hose set must not be sterilized: it cannot withstand high temperatures and may be damaged! 1 07637170 2 Cleaning and disinfecting To ensure material compatibility, use disinfectants based on: — aldehydes — alcohols — quaternary ammonia compounds.
Care Users in the Federal Republic of Germany are recommended to use only disinfectants on the current DGHM list (DGHM: German Society for Hygiene and Microbiology). The following disinfectants on the DGHM list are recommended: — Dismozon pur — Incidur — Sekusept Powder — Trichlorol The DGHM list (published by: mhp-Verlag, Wiesbaden) also specifies the active ingredient in each disinfectant.
Care Sterilising reusable hose sets Disassemble the breathing valve, flow sensor and angled connector. Dismantle the breathing valve. The disassembled parts of the breathing valve, the flow sensor, the angled connector, the flow measuring hoses and the ventilation hose ● can be sterilized in hot steam at 134 °C in accordance with EN 285 (Sterilization – Steam sterilization – Large-scale sterilization) for at least 3 minutes.
Care Batteries Batteries: ● Do not burn, risk of explosion! ● Do not open forcibly, risk of caustic burns! Batteries are special waste: ● They must be disposed of in accordance with the local waste disposal regulations. Device disposal — at the end of its useful life. ● Oxylog 3000 must be disposed of correctly in consultation with the relevant waste disposal companies. ● The statutory regulations must be observed.
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
What's what What's what What's what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Front panel – standard version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Front panel with all options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Side view, right . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
What's what What's what Front panel – standard version 1 2 3 4 5 6 7 8 19 9 18 10 17 16 15 1 Screen with screen pages for the specific application 2 Key »Alarms ∆∆ « for setting and displaying alarm limits 3 Key »Set.
What's what Front panel with all options 1 2 3 4 5 6 7 8 9 21 10 11 20 12 13 18 17 1 Screen with screen pages for the specific application 2 Key »Alarms ∆∆ « for setting and displaying alarm limits 3 Key »Set.
What's what Side view, right 1 2 3 4 5 6 Screw for securing the battery compartment cover Sockets for flow measuring hoses Socket for ventilation hose or inhalation mask Connector for medical gas hose Socket for DC supply Window for IrDA interface Note Instructions for Use 1 2 3 6 4 MT-156-2001 5 Rear view 7 Filter cartridge for ambient air 8 Rating plate 8 68 MT-158-2001 7 This "Instructions for Use" manual is published for public use and is only meant for your information.
What's what Reusable hose set 1 2 3 4 5 Breathing valve Ventilation hose Flow measuring hoses Angled connector Flow sensor 1 2 5 3 04937170 4 Disposable hose set Breathing valve Ventilation hose Flow measuring hoses Angled connector Flow sensor 1 2 5 4 08137170 1 2 3 4 5 3 This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Accessories Accessories Carrier System 3000 (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Onboard Equipment Holder (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Carrier System 3000 (optional) Carrier System 3000 (optional) Intended use Carrier System 3000 – transport unit for Oxylog 3000, with oxygen cylinder, pressure reducer and ventilation hose set for emergency care of patients, with optional choice of O2 supply. Preparing for use ● Connect the marked gas inlet of the Carrier System 3000 to a high-pressure oxygen supply with a pressure range between 3 and 6 bar. ● Slowly open the cylinder valve on the full O2 cylinder.
Carrier System 3000 (optional) When mounted on a tube or rail, the Carrier System is only held by its own weight. The ventilator must be secured additionally, particularly when it is being transported. Vibrations may cause the ventilator to drop and injure the patient or user. After use ● Disconnect the high-pressure supply and close the valve on the O2 cylinder. Technical data Operating temperature Storage temperature Width x Height x Depth Weight with maximum equipment and 2.
Onboard Equipment Holder (optional) Onboard Equipment Holder (optional) Intended use Onboard Equipment Holder – mounting unit for Oxylog 3000 which can be inserted, locked and subsequently removed by the user and which can be pivoted horizontally. The onboard equipment holder is designed for mounting on the roof or wall of a vehicle. Preparing for use ● Secure the onboard equipment holder to the wall of the vehicle via the three mounting holes with M5 screw connections provided for this purpose.
Service Mode Service Mode Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Customer Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Set startup settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Select hose type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Mode Service Mode Customer Service Mode In service mode, the ventilator performs function tests, outputs status information and permits configuration of parameter settings. Displays in service mode appear in English and cannot be changed to any other language.
Service Mode 11737170 Settings in service mode Select the required function with the cursor (asterisk). ● Select parameter = turn rotary knob. ● Activate parameter = press rotary knob. ● Set value = turn rotary knob. ● Confirm value = press rotary knob. 11837170 Quit test: ● Select line »EXIT« = press rotary knob and confirm. The set values are saved and remain effective whenever ventilation is started after switching on.
Service Mode To restore the manufacturer's defaults: ● Select and confirm line »Set factory default«. Range of settings: Parameter Tinsp I:E PEEP ∆ ASB Pinsp Ramp TApnoea Tplat Trigger MV-high MV-low Frequ.-high Sound Brightness NIV O2-Flow Range 0.2 to 10.0 s 3.0/1.0 to 1.0/4.0 0 to 20 mbar 0 to 35 mbar 0 to 55 mbar SLOW,STANDARD, FAST 0 to 30 s 0 to 50 % 0 to 15 L/min 2.0 to 41 lpm 0.
Service Mode Set measured values display window 11637170 The arrangement of measured value pairs on the individual pages of the measured values display window can be varied. Each measured value can be freely selected in any position and is only displayed at that position. ● Start configuration on page 1/5 and continue through to 5/5. Enter activation code 11437170 The activation codes for options can be entered. The activated options are then displayed.
Service Mode Test loudspeaker, buzzer, LEDs and display Tests the loudspeaker, buzzer, all LEDs and the display. Select the required test ● Start the test. Each function is tested by the device. To test the screen display (Test display): ● Turn the rotary knob; various test cards are displayed. 09537170 The selected test remains active until the rotary knob is pressed again.
Service Mode Display error and info logbook Any technical errors and/or special occurrences, such as activation of a software option, completion of the device check and device calibration, are listed in chronological order. Display (example): 10037170 Change over to the next page: ● Select line »Page«, confirm and turn rotary knob. Display settings logbook The operating phases with ventilator settings and time are listed in chronological order.
Service Mode Exit service mode ● Press key »O« for 3 seconds; its yellow lamp flashes. To switch ventilation on: ● Briefly press key »O«. To switch off: ● Press rotary knob. 82 This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Abbreviations and symbols Abbreviations and symbols Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 This "Instructions for Use" manual is published for public use and is only meant for your information.
Abbreviations Abbreviations Abbreviation Explanation Abbreviation Explanation ASB (PS) Assisted Spontaneous Breathing Pressure-assisted spontaneous breathing Paw Airway pressure PCV+ Pressure Controlled Ventilation plus BIPAP (PCV+) Biphasic Positive Airway Pressure Spontaneous breathing with continuous positive airway pressure and two different pressure levels PEEP Positive end expiratory pressure Pinsp Set value of the upper pressure level in BIPAP Biphasic Positive Airway Pressure Assist
Symbols Symbols Symbol Explanation , Inspiration is started and held manually Settings Display screen window "Settings" ∆∆ Alarms ∆∆ Display screen window "Alarms" Values Display screen window "Measured values" Curves Changeover between flow / pressure curve ∆∆ Suppress acoustic alarm for 2 minutes Alarm Reset Acknowledge alarms E Standby/Operation switch _ Lower and upper alarm limits > Upper alarm limit only < Lower alarm limit only ! Advisory message !! Caution message !!!
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Technical Data Technical Data Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Technical Data Technical Data Ambient conditions During operation Temperature Atmospheric pressure Rel. humidity –20 to 50 oC1) 570 to 1200 hPa 5 to 95 % During storage Ventilator without replaceable battery, with reusable ventilation hose set Temperature Atmospheric pressure Rel. humidity –40 to 75 oC 570 to 1200 hPa 5 to 95 % Disposable ventilation hose set Temperature Atmospheric pressure Rel.
Technical Data Pressure support ∆ASB Rise time for pressure support Performance data Control principle Max. inspiratory flow Device compliance with 1.5 m ventilation hose with 3 m ventilation hose Inspiration resistance Expiration resistance Dead space incl. flow sensor Supplementary functions Demand valve Relief valve Max.
Technical Data Frequency measurement Range Resolution Accuracy 0 to 99 /min 1 /min ±1 /min Curve display Airway pressure Paw (t) Flow (t) –10 to 100 mbar –120 to 120 L/min Monitoring Expiratory minute volume MV Alarm, upper alarm limit Range of settings Alarm, lower alarm limit Range of settings when the upper alarm limit has been exceeded 41 to 2 L/min when the level drops below the lower alarm limit 0.
Technical Data Charging times Nickel Metal Hybrid battery Lithium ion battery The specified charging times apply when recharging the battery fully after it has been exhausted. Approx. 4 hours Approx. 5 hours Permissible ambient temperature during charging 0 oC to 35 oC Indication of battery capacity in 25 % increments Accuracy of the capacity indication The indicated capacity is determined by the battery itself.
Technical Data Gas cylinders and pressure reducers Pressure reducer Gas consumption for internal control Accuracy of gas consumption indication must comply with national regulations and be officially approved. must have a vent valve on the output side to limit the delivery pressure to approx. 10 bar in the event of a fault. 0.1 to 0.5 L/min ±0.
Technical Data Materials used Housing, Oxylog 3000 Housing, AC/DC power pack Housing, DC/DC converter Touch sensitive keypad on ventilator Impact-proof acrylonitrile butadiene styrene (ABS) Impact-proof acrylonitrile butadiene styrene (ABS) Impact-proof acrylonitrile butadiene styrene (ABS) Polyester film Reusable ventilation hose set Ventilation hose, flow measuring hoses Flow sensor housing, breathing valve Vane in flow sensor Diaphragms in breathing valve Silicone rubber Polysulphone (PSU) Stainless
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Description Description Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Volume-controlled ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 IPPV (CMV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description Description Ventilation modes Volume-controlled ventilation IPPV (CMV) Volume-constant mandatory ventilation stroke The ventilation pattern is specified by the settings for tidal volume V T, frequency Freq., ventilation time ratio I:E and PEEP. At the end of the flow phase, the expiration valve remains closed until the end of the inspiration time Tinsp. This phase, the inspiratory pause, can be identified as the plateau Pplat in the curve Paw (t).
Description SIMV Paw Unsynchronised mandatory ventilation stroke Synchronised mandatory ventilation stroke PEEP Trigger window Tinsp Spontaneous breathing time Set mandatory 1 cycle time Freq. t Spontaneous breathing time + T Set mandatory 1 cycle time Freq. This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Description ASB (PS) The machine then produces an increase in pressure up to the preselected ASB pressure ∆ ASB above PEEP, which is adjustable to the breathing requirement of the patient. The time for this pressure increase (»Ramp«) is adjustable: — In case of rapid increase in pressure Oxylog 3000 supports the insufficient spontaneous breathing of the patient with a high peak flow. — In case of slow increase in pressure Oxylog 3000 begins gently with regular inspiratory flow.
Description BIPAP (PCV+) Biphasic Positive Airway Pressure (Pressure Controlled Ventilation plus) Paw Mandatory BIPAP stroke Spontaneous breathing Synchronized BIPAP stroke Pinsp ramp PEEP Triggerwindow t Tinsp 1 Freq. This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Description 100 Paw Mandatory BIPAP stroke Spontaneous breathing Synchronized BIPAP stroke Pinsp ramp PEEP Triggerwindow t Tinsp 1 Freq. 03137171 Using BIPAP As with SIMV, the time pattern is set using the basic setting parameters of frequency Freq. and inspiration time Tinsp. The lower pressure level is set with the PEEP parameter, while the upper level is set with Pinsp.
Description Functional description Microprocessor system Gas blender V1 – 3 F1 V10 DR NV SV S3 P E S8 F2 V7 P E P E S9 S7 The variable pneumatic actuators in the Oxylog 3000 are controlled by the microprocessor system via digitized electrical test signals. Gas supply The supply gas O2 (or compressed medical air in exceptional cases) is purified by filter F1 and adjusted to a constant pressure by pressure regulator DR. Ambient air is taken in via filter F2 as required.
Order List Order List Part name Part No. Part name Part No. ZV*-O2/Air Workstation Oxylog 3000 2M 86300 Basic unit Oxylog 3000 2M 86955 Accessories required for operation Power supply: AC/DC power pack 100-240 V/50-60 Hz 2M 86 730 Available power cables: Air connecting hose with angled connector for central supply and DIN connection Length 1.5 m M 29 285 Length 3.0 m M 29 245 Length 5.
Order List Part name Part No. Oxylog 3000 emergency rescue units Portable system, complete 2M 86 970 For supplying the Oxylog 3000 from a piped medical gas supply or cylinder via an O2 manifold, comprising: Oxylog 3000 Ventilation hose with measuring leads Breathing valve Flow sensor Angled connector O2 cylinder, 2.5 L O2 pressure reducer O2 manifold Equipment bags This "Instructions for Use" manual is published for public use and is only meant for your information.
Index Index 100 % O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Advisory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Air Mix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Alarms screen window . .
Index Primary care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 PS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Replacing the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Reusable hose set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, Routine and additional functions . . . . . . . . . . . . . . . . . . . . . . RTCA DO–160D . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
This "Instructions for Use" manual is published for public use and is only meant for your information. We do not guarantee the accuracy of these instructions with respect to your specific device type or status. Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
These Instructions for Use apply only to Oxylog 3000 with Serial No.: If no Serial No. has been filled in by Dräger these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device. ç Directive 93/42/EEC concerning Medical Devices Dräger Medical AG & Co. KGaA Germany z Moislinger Allee 53 – 55 D- 23542 Lübeck y +49 451 8 82- 0 x 26 80 70 FAX +49 451 8 82- 20 80 ! http://www.draeger.com 90 37 171 - GA 5503.