Operator`s manual

27
Manual (LOCAL) Operation
4. NOTE: this operation “puts the matrix’s switches to sleep”
thereby rendering the matrix inoperable during the process.
ENTER or CLEAR or rebooting returns the matrix to normal.
5. NOTE: proper performance of Find Switch ID relies on the
behavior of Dow-Key Switch firmware revision 4 and above –
revision 3 and below does not allow “putting the switch to
sleep”. So, this feature is best executed on a matrix that
does not contain switches with firmware revision 3 or less, or
those switches will respond with their ID’s as well as the
switch in question. The revision of firmware of any switch
configured to the matrix may be learned with Main
Menu>System Settings>Switch Information (see below).
Temperatures
View current values of a maximum of 4 temperature sensors,
and set thresholds at which an Over Temperature alarm should
occur. Setting all 4 alarm thresholds to 0° Celsius disables Over
Temperature alarms and causes the current temperatures to
read out 0° as well; this is the recommended setting for Matrix
Products that contain no temperature sensors.
Most standard Models do not feature temperature
sensors.
Switch Closure Counts
View the number of times any position of any switch configured
to the matrix has been closed upon, to a maximum of
1,000,000.
Switch Information
View the Part Number, Serial Number, Firmware Number,
Firmware Revision Level, maximum number of positions, its
PCBA Code (factory), and Coil Delay Time, of any switch
configured to the matrix.
Default Settings
This password protected option is used during the assembly
process and is not intended for Operator use.