Operator`s manual
Table Of Contents
- Contents
- Figures
- Tables
- Safety
- Introduction
- Setting Up the 708-DS / 709-DS
- Operating the 708-DS / 709-DS
- Main Screen - Options
- Main Screen - Display Parameters
- Main Screen - Sample Temperatures
- Main Screen - Run
- Main Screen - Stop
- Main Screen - Drive Unit Up / Down
- Manual Drive Unit Up / Down
- Manual Sampling
- Main Screen - Manifold Up / Down
- Main Screen - Lock
- Main Screen - Unlock
- Main Screen - Remote Control
- Main Screen - Alarms
- Main Screen - Maintenance Due Icon
- Menu Screen - System Menu
- General 708-DS / 709-DS Conventions
- Agilent 708-DS / 709-DS Notifications
- Main Screen - Options
- Maintenance and Troubleshooting
- Index

98 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
Parameter Limits / Ranges
Table 1 4 Parameter Limits / Ranges
Parameter Range Resolution Configured
Spindle Speed 10.0 - 250.0 RPM 0.1 RPM “Main Screen - Display Parameters” on page 48
“Run - Manual Operation (708-DS)” on page 54
“Method Properties Screen 1” on page 69
“Menu Screen - Diagnostics Screen 1” on page 90
Temperature
(Bath or Vessel)
25 - 55 ºC 0.1 ºC “Main Screen - Display Parameters” on page 48
“Method Properties Screen 1” on page 69
“Menu Screen - Diagnostics Screen 1” on page 90
Serial Number 15 characters Factory
Profile Print
Interval
000:01:00 to
024:00:00
1 second “Run - Manual Operation (708-DS)” on page 54
“Method Properties Screen 2” on page 70
Test Duration 000:00:10 to
999:59:59
1 second “Run - Manual Operation (708-DS)” on page 54
“Method Properties Screen 3” on page 72
Final Spin
Duration
000:00:10 to
001:00:00
1 second “Method Properties Screen 2” on page 70
Tester ID 12 characters “Menu Screen - Instrument Settings Screen 1” on page 82
Comm ID 01 to 99 01 “Menu Screen - Instrument Settings Screen 1” on page 82
Media Volume 100 to 2000 mL 1 mL “Run - Manual Operation (708-DS)” on page 54
“Method Properties Screen 1” on page 69
DDM Increment 000:00:00 to
001:39:59
1 second “Run - Manual Operation (708-DS)” on page 54
“Method Properties Screen 2” on page 70
Temp Error
Tolerance
0 to ±0.5 ºC 1 second “Menu Screen - Instrument Settings Screen 2” on page 84
Note: A value of 0 disables the tolerance check.
Speed Error
Tolerance
0 to 1.0 RPM 1 second “Menu Screen - Instrument Settings Screen 2” on page 84
Note: A value of 0 disables the tolerance check.
Manifold Down
Time
000:00:00 to
001:00:00
1 second “Method Properties Screen 3” on page 72