Operating the 708-DS / 709-DS 4
708-DS / 709-DS Operator’s Manual 73
From the Method Properties screen:
• Press Cancel to exit without saving the values and return to the
Select Method screen.
• Press Back to save all values to permanent memory and return to
the previous Method Properties screen.
• Press Save to save all values to permanent memory and return to
the Select Method screen.
• Press Next to save all values to permanent memory and proceed
to the next Method Properties screen.
Enable Sample
Point Alarm
Enabling this option activates the sample notification dialog as well as
activating the audible and visual alarms. This alarm draws the operator
to the machine at the timepoint for manual sampling scenarios.
Alarm Lead Time The Time Point Approaching screen (“Time Point Approaching” on
page 102) is available to notify the user of upcoming time points. The
time entered is how long before the time point the Time Point
Approaching screen displays. This alarm draws the operator to the
machine at the timepoint for manual sampling scenarios.
Table 4 Method Properties Screen 2
Option Function