Operating the 708-DS / 709-DS 4
708-DS / 709-DS Operator’s Manual 55
The Manual Operation screen allows you to run a manual method. Press
the rectangular area following the text description to enter parameters.
The Data Entry screen displays, allowing you to enter the data.
• Press Cancel to return to the Select Start Type screen without
saving the test information.
• Press Accept to save the entered values and return to the Main
screen.
• Press Run in the Manual Operation screen. The Main screen
displays and the following actions take place:
• The Manually Insert Dosage(s) dialog box displays (“Load
Dosage Prompt” on page 105).
• If DDMs are installed, tablets are automatically dropped into the
vessels simultaneously or sequentially based on the DDM
Increment. For example, a DDM Increment of 5 seconds drops the
first tablet immediately, the second tablet 5 seconds after the first,
and so on until all the tablets have been dropped. If the increment
is zero, all tablets are dropped simultaneously.
• Elapsed time is reset to zero and incremented once per second
while the test runs.
• The spindles start rotation at the set speed.