The Manual Operation screen allows you to run a manual method. Press
the rectangular area following the text description to enter parameters.
The Data Entry screen displays, allowing you to enter the data.
Press Cancel to return to the Select Start Type screen without
saving the test information.
Press Accept to save the entered values and return to the Main
screen.
Press Run in the Manual Operation screen. The Main screen
displays and the following actions take place:
The Manually Insert Dosage(s) dialog box displays (“Load
Dosage Prompt” on page 105).
If DDMs are installed, tablets are automatically dropped into the
vessels simultaneously or sequentially based on the DDM
Increment. For example, a DDM Increment of 5 seconds drops the
first tablet immediately, the second tablet 5 seconds after the first,
and so on until all the tablets have been dropped. If the increment
is zero, all tablets are dropped simultaneously.
Elapsed time is reset to zero and incremented once per second
while the test runs.
The spindles start rotation at the set speed.