22 708-DS / 709-DS Operator’s Manual
3 Setting Up the 708-DS / 709-DS
Environmental Requirements for Installation
• Humidity: max relative humidity 80% for temperatures up to 31 ºC
decreasing linearly to 50% relative humidity at 40ºC
• Indoor use only
• Pollution Degree: 2
• Installation Category: II
• Altitude: 2000m
• Temperature: 5 ºC to 40 ºC
• Power: 115/230 V ac, 50/60 Hz, 2.5 A
Main supply voltage fluctuations are not to exceed ± 10% of the
nominal supply voltage
Clearance
1 Prepare the area where the equipment is to be located.
2 Ensure a minimum clearance of 30 cm above the unit (when fully
raised) and 10 cm at the rear and on both sides of the Agilent 708-DS
/ 709-DS. Approximately 82 cm x 82 cm total bench space is
required per apparatus.
3 Orient the Agilent 708-DS / 709-DS and all other system
components appropriately.