104 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
Warning Screens (Yellow)
Table 1 6 Warning Screens (Yellow)
Instrument Calibration Due
This Service Due warning displays when the instrument is due for
calibration (“Menu Screen - Calibration Screen 1” on page 87). Press
Ok to clear the warning.
Instrument Preventative Maintenance Due
This Service Due warning displays when the instrument is due for
preventative maintenance (“Preventive Maintenance” on page 116).
Press Ok to clear the warning.
Temperature Probe Dwell Time Warning
This Method Warning screen displays when an Auto Sampling Manifold
is installed and enabled (“Method Properties Screen 3” on page 72). If
the sum of the Manifold Down Duration and Manifold Lead Time
(“Method Properties Screen 3” on page 72) are less than 30 seconds, this
screen displays. Press Ok to clear the warning.
Invalid Initial Temperature Warning
This Method Warning screen displays only when Apparatus 1 (basket)
or Apparatus 6 (rotating cylinder) are installed and a Bath Temperature
Start is selected from the Start Options screen (“Method - Start
Options” on page 52). Press Ok to clear the warning and start the
method without the initial temperature being taken. Press Cancel to
abort the run and return to the Main screen.