User's Manual
16 / 24
11.Parameters of LED light source
12.Data shall be refreshed within two seconds. The average time of data is set in accordance with 3.5.7 and
data in set average time will be selected and averaged.
Light Source
Central Wavelength
Radiant Power
Red light
660nm
<15mW
Infrared light
905nm
<15mW
5.2 List of Functions
List of Functions
OX201
Measurement of blood oxygen saturation
√
Measurement of pulse rate
√
Low battery indication
√
Display of pulse histogram
√
Display of perfusion intensity
√
Turn off automatically in eight(8) seconds
if there is no finger
√
Bluetooth connectivity
√
Waveform display
√
5.3 Performance Index
1. Measurement scope and accuracy of blood oxygen saturation Measurement range: 1%~100%. Within
measurement range of 70%~100%, the measurement accuracy is ±2%
2. Measurement range: 25 bpm~100bpm. The measurement accuracy is ±2bpm.
Measurement range: 101 bpm~250bpm. The measurement accuracy is ±2%bpm.
3. Measurement scope and accuracy of respiratory measurement The measurement scope of respiratory rate
is between 4rpm~70rpm with measurement accuracy not specified.
6 Maintenance
Regular maintenance is of great importance to ensure the normal operation of equipment.
Caution:
All safety inspection or maintenance work of equipment to be disassembled shall be carried out by
professional maintenance personnel, and the non-professional personnel may cause equipment failure.
If there is any problem with the equipment, please contact the maintenance personnel or our company.
6.1 Cleaning and Disinfection
1. Clean surface of the Pulse Oximeter before it is used for patients.