3. The parameters of the oximeter can be set through the following menu.
1. Patient type setting. Including: adult children
2. Response speed setting. Including: Rapid,Normal,slow
3. Sensitivity setting. Including: low,Middle,High.
4. Delete all Historical Data
5. Device search list
5 Product Specification
5.1 Specification of host
1.The product uses internal power supply.
2.Power requirements: 2 AAA batteries
3.Liquid inlet protection degree: IP22
4.Operation mode: run continuously
5. Electrical safety classification: BF type application part
6.Not suitable to use in places with flammable anesthetic gas.
7.Normal working conditions: ambient temperature range: 10 ~ 40 ℃ (50 ~ 104℉); ambient humidity
range: 15% ~ 95% non condensing; atmospheric pressure range: 70kPa ~ 106kpa
8.Storage environment: environment temperature range: - 20 ~ 60 ℃ (-4 ~ 140℉); environment humidity
range: 15% ~ 95%, non condensing; atmospheric pressure range: 50kPa ~ 107.4kpa
9.Physical dimension: 61.8×34.2×33.9mm
10.Product weight: About 50g (including battery)