Health Care refrigerator RA 122 H, MR 60 Operating Instructions 207.4489.
Please read these operating instructions carefully before putting the refrigeration unit into operation. If you later sell or dispose of it, please ensure that the new owner receives these operating instructions. Please follow the installation details below carefully Guarantee is valid for products installed as described only. Thank you for choosing our appliance. We are sure it will provide you with trouble-free use.
CONTENT S 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Unpacking View of the appliance Cleaning Positioning the appliance 4.1. Installation instruction Using the appliance 5.1. Temperature regulation Changing the feet Thermometer Useful suggestions Defrosting, cleaning and maintenance Customer service Putting into operation, technical data 11.1.
1. Unpacking After removal from the cardboard packaging, make sure the appliance is not damaged. If you find damage to the appliance resulting from transport, report it immediately to the transportation firm. 2. View of the appliance RA 122 H MR 60 Figure 1. 1. door lock 3. thermostat knob 2. thermometer 4. evaporator 5. hinge plate DOMETIC RT DO NOT ACCEPT LIABILITY FOR LOSS OR DAMAGE OF KEYS. HOWEVER WE WOULD BE HAPPY TO SUPPLY A COMPLETE LOCK AND KEY REPLACEMENT.
3. Cleaning Clean the refrigerator thoroughly at intervals, as necessary. Remove the shelves; these, the cabinet interior and the door should be washed with a warm solution of sodium bi-carbonate. Finally wipe over with a cloth rinsed in warm water only, and dry with a clean cloth. Do not wash any plastic parts in water that is more than hand warm, and do not expose them to dry heat. The outside of the cabinet should be wiped down regularly with a clean, damp cloth.
3. Ventilation must be provided as shown in alternatives A, B, C or D (fig 2-3) 4. The ventilation duct must at least measure 105mm x the width of the refrigerator 5. Only the entire cooling unit must project into the duct as shown 6. Air passing through the duct must not be preheated by any source of heat 7. Ventilation grills, if used, must have openings of at least 200cm2 each During installation, please ensure the mains plug is accessible at all times. 5.
6. Changing the feet Should you wish to install your Health Care refrigerator, you will not need the 40-mm feet. After having removed the 40-mm feet, you can then re-attach the 12-mm feet, located inside the 40-mm feet, to the front and rear fixing points. The 40-mm feet and the 4 remaining screws should be retained for future use. 7. Thermometer (Fig. 5) Refer to the separate instruction leaflet supplied with the thermometer.
10. Customer service Before you call a service engineer, carry out the following simple tests: Is there a power failure? Check the fuses. Is the plug firmly connected to the socket and is the socket switched on? Check the socket by plugging in another appliance, e.g. a table lamp. Is the thermostat at the correct setting? (normally 3 or 4).
11.1 Connecting to electrical mains The appliance may only be operated from nominal voltage mains as shown on the data plate. The appliance mains plug may be connected to a mains socket earthed in accordance with regulations. Any electrical work required to install this appliance should be carried out by a qualified electrician or competent person. The manufacturer declines any liability should these safety measures not be observed.
12. Environmental protection information The appliance does not contain any CFCs/HCFCs. Ammonia (natural hydrogen and nitrogen compound) is used as a refrigerating agent in the refrigeration unit. The ozone-friendly cyclopentane is activated as a motive agent for the PU foam insulation. Sodium chromate is used for corrosion protection (less than 2 weight % of the coolant). 13. Recycling After unpacking the appliance, the packing materials should be delivered to a local collection site.
Printed by Xerox Hungary Ltd. Rev.: 2003. 08. 01.