User`s guide

Dome Sx Display User’s Guide, 020-0931-00 Rev. F | 15
Safety precautions
External equipment intended for connection to signal input, signal
output, or other connectors, must comply with the relevant IEC
standard (EN 60601-1:2006 and IEC 60601-1:2005 series for medical
electrical equipment). In addition, all such combinations (systems)
must comply with the standard IEC 60601-1:2005, Safety
requirements for medical electrical systems.
Equipment not complying to IEC 60601-1:2005 must be kept outside
the patient environment, as defined in the standard as at least 1.5
meters from the patient or the patient support.
Any person who connects external equipment to signal input, signal
output, or other connectors has formed a system and is therefore
responsible for the system to comply with the requirements of
IEC 60601-1:2005. If in doubt, speak with a qualified technician.
The power cord is the only disconnect device that allows you to cut
power from the AC main. Position the display in order to operate the
disconnect device without difficulty.
Never touch the display and the patient at the same time.
WARNING! No modification of this equipment is allowed.