Dome® Sx Display User’s Guide 0197 Dome S3c Display Dome S10 Display
© 2014 NDS Surgical Imaging. All rights reserved. Information in this document has been carefully checked for accuracy; however, no guarantee is given to the correctness of the contents. This document is subject to change without notice. Go to web site ndssi.com for the latest documentation. NDSsi provides this information as reference only. Reference to products from other vendors does not imply any recommendation or endorsement. This document contains proprietary information protected by copyright.
Contents Display Package Contents . . . . . . . . . . . . . . . . . . . . . . . . . . 4 About the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Display Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Installing the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Package Contents The Dome S10 display, shown here for illustrative purposes. Your display package may contain a Dome S3, Dome S3c, Dome S6c, or Dome S10 unit. DVI cable DC power adapter Power cord DisplayPort cable Mini USB cable Quick reference NOTE: DVI-to-DVI and DisplayPort cables ship in the Dome S3 and Dome S3c display packages. DVI-to-DVI cables ship in the Dome S6c display package. DVI and mini USB cables ship in the Dome S10 display package.
About the Display The Dome® Sx display systems, available in 3, 6, and 10 megapixels, consist of a TFT LCD panel and a graphics board and driver.The display’s thin film transistors, in a transmissive-type display, use an integrated cold cathode fluorescent tube (CCFT) backlight system. The new model of the Dome S3c display and the Dome S6c display both use an LED backlight system.
Related Documents ITEM ITEM NUMBER DESCRIPTION Dome CXtra Quick Reference 020-0914-xx Quick reference booklet that covers installation of CXtra calibration software and describes DICOM, AAPM TG18, DIN, and MQSA quality control functions. (Printed material in Dome CXtra CD 015-0081-xx envelope.) Dome CXtra User's Guide 020-0911-xx In-depth user's guide that covers installation and configuration of CXtra calibration software and use of DICOM, AAPM TG18, DIN, and MQSA quality control functionalities.
Display Components Review the following illustrations to identify controls and ports on the front and back panels of the Dome Sx displays. Front sensor on Dome S3/Dome S3c MQSA button on Dome S10 The sensor on the front of the Dome S3, Dome S3c, and Dome S6c displays allows you to run an auto-conformance test without using the photometer. On the S3 series the sensor is located on the side bezel as shown in the preceding illustration.
Dome S3/Dome S3c connector plate Dome S6c connector plate Dome S10 connector plate The Dome Sx displays include the following components on the connector plate. • Power input. – On Dome S3, drives power to the display via an 8-pin DIN connector. The power input is 12V ±5% (130 W). – On Dome S3c, Dome S6c, and Dome S10, drives power to the display via a 6-pin DIN connector. The power input is 24V ±5% (150 W). • USB port. – Dome S3/Dome S3c uplink B port. Usage to be defined at a future time.
• Video input. Drives the data to the display. – Dome S3 and Dome S3c displays each offer two video connector options: a DVI connector or a DisplayPort connector. – Dome S6c and Dome S10 displays each offer two DVI connectors. (NOTE: For the Dome S6c display, a DisplayPort 1.1 video output port and a DisplayPort 1.2 video input port are future features.) • Reset button. Restores the display configuration to default setting. • LED lights. Provides information on the status of the display.
Installing the Display First, turn your computer off. Next, leave the power cord plugged into the grounded outlet. Then install the Dome Sx display using only the power adapter and video cable that shipped with the unit. WARNING No modification of this equipment is allowed. In locations where 240V outlets are used, connect the Dome Sx display only to a center-tapped, 240V, singlephase supply (for Canada and the United States only).
Video Cable Installation on Dome S6c and Dome S10 Displays For both the Dome S6c and the Dome S10 displays, connect the primary output from the graphics board (the bottom port) to the primary input of the Dome S6c and the Dome S10 displays. (See following illustration for the location of the primary input port on the connector plate of each display.) Improper connection may cause VGA boot messages not to appear on the display screen.
Technical Information The design of the Dome® Sx digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display. Read all instructions before setting up the display. Classification Shock Protection: Class I. Degree of Protection Against Electric Shock: No applied part. Degree of Protection Against Harmful Ingress of Water: Ordinary equipment (IPX0).
Symbol explanations READ DOCUMENTATION. Follow operating instructions or consult instructions for use. MANUFACTURER. DATE OF MANUFACTURE. SERIAL NUMBER. AUTHORIZED REPRESENTATIVE. European Community. DANGEROUS VOLTAGE. Important precautions about electric shock. Read the accompanying text carefully, to prevent damage to display components and to guard your safety. DIRECT CURRENT. BAROMETRIC PRESSURE. Transport and storage 12,000 meters (39,400 feet), maximum in unpressurized container. RELATIVE HUMIDITY.
The Dome S series offers AMLCD units designed for viewing medical X-ray images. A Dome Sx display should not be used near patients and should be kept outside of 1.5 m perimeter and 2.29 m vertical. Intended use Dome Sx line, models Dome S3, Dome S3c, and Dome S6c are intended for use in displaying and viewing radiography images for review, analysis, and diagnosis by trained medical practitioners.
Safety precautions External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (EN 60601-1:2006 and IEC 60601-1:2005 series for medical electrical equipment). In addition, all such combinations (systems) must comply with the standard IEC 60601-1:2005, Safety requirements for medical electrical systems.
Safety tips • Never open the display case, even when the power is off. Dangerous voltage inside may cause electric shock or death. • To avoid damage to the display, use the grounded power supply and video cable supplied by NDSsi, or use certified replacements. • Be sure the display is electrically grounded. You must connect the third grounding pin on the US power cord to a grounded outlet. The European power cord does not have a third grounding pin, but it must be plugged into a grounded outlet.
Unpacking and handling tips The Dome Sx display is a precision instrument that requires proper care to maintain product operation and adherence to specification. Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel. • Use both hands to grasp the display case when lifting it from the shipping carton, but avoid touching the screen. • Do not apply pressure to the screen or touch the screen with bare fingers or objects.
Preventing fire and injury • Replace the power supply or cables if damaged. • Use only the power source indicated in this guide or listed on the display. • Do not plug the power supply into an overloaded AC outlet or extension cord. Overloaded AC outlets and cords can result in electric shock or fire. • Do not drop or push objects into the display case. Internal components contain high voltage. • Unplug the power cord from the wall outlet during thunderstorms.
WARNING! DISCONNECT SUPPLY BEFORE SERVICING AVERTISSEMENT COUPER L’ALIMENTATION AVANT L’ENTRETIEN ET LE DEPANNAGE Shipping/storing the display Keep the display in its shipping container until installation. Return the display to its original container whenever you need to store the unit, move it to another location, or return it for repair. The packaging supplied by the manufacturer protects the display while it is in transit. See environment specifications for more information.
Dome S3 Specification In locations where 240V outlets are used, connect the Dome S3 display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification LCD Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 541 mm (21.3 in.) Display resolution 1536 x 2048 pixels (portrait) 2048 x 1536 pixels (landscape) Pixel pitch 211.
Power Supply WARNING: Use only the power adapter supplied with the Dome S3 display unit; the BridgePower BPM130S12F02 (130W) model. Category Characteristic Item Specification Power input requirements Voltage selection Universal auto switching with medical approvals and PFC 100–240V AC 3.0 A (3.0–1.5 A) 50–60 Hz 12 V DC ±5% 10.8 A (130 W) 70 W (typical) 228.6 mm x 76.2 mm x 50.8 mm (9 in x 3 in x 2 in) 1.3 kg (2.
Dome S3c Specification In locations where 240V outlets are used, connect the Dome S3c display to a centertapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification LCD Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 541 mm (21.3 in.) Display resolution 1536 x 2048 pixels (portrait) 2048 x 1536 pixels (landscape) Pixel pitch 211.
Power Supply WARNING: Use only the power adapter supplied with the Dome S3c display unit; the BridgePower BPM150S24F05 (150W) model. Category Characteristic Item Specification Power input requirements Voltage selection Universal auto switching with medical approvals and PFC 100–240V AC 2.5 A (2.5–1.5 A) 50–60 Hz 24 V DC ±5% 6.25 A (150 W) 66 W LED backlight (typical) 81 W CCFL backlight (typical) 223.5 mm x 63.5 mm x 38.1 mm (8.8 in x 2.5 in x 1.5 in) 0.8 kg (1.
Dome S6c Specification In locations where 240V outlets are used, connect the Dome S6c display to a centertapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification LCD Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel 762 mm (30.0 in.
Power Supply WARNING: Use only the power adapter supplied with the Dome S6c display unit; the BridgePower Corp. BPM150S24F05 (150W) model. Note that adapter has a power switch. Category Characteristic Item Specification Power input requirements Voltage selection Universal auto switching with medical approvals and PFC 100–240V AC 2.5 A (2.5–1.5 A) 50–60 Hz 24 V DC ±5% 6.25 A (150 W) 85 W (typical) 223.5 mm x 63.5 mm x 38.1 mm (8.8 in x 2.5 in x 1.5 in) 0.8 kg (1.
Dome S10 Specification In locations where 240V outlets are used, connect the Dome S10 display to a centertapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification LCD Technology TFT AMLCD Dual Domain IPS, single 10-bit pixels Screen size, diagonal 761 mm (30.0 in.
Power Supply WARNING: Use only the power adapter supplied with the Dome S10 display unit; the BridgePower BPM150S24F05 (150W) model. Category Characteristic Item Specification Power input requirements Voltage selection Universal auto switching with medical approvals and PFC 100–240V AC 2.5A (2.5–1.5 A) 50–60 Hz 24 V DC ±5% Power output requirements Power consumption Physical Voltage Current Frequency Voltage Current Wattage Size 6.25 A (150 W) 90 W (typical) 223.5 mm x 63.5 mm x 38.1 mm (8.8 in x 2.
Windows Video Modes Resolutions are expressed in pixels, width by height (W x H). Display Resolution Orientation Dome S3 1536 x 2048 Portrait 2048 x 1536 Landscape 1536 x 2048 Portrait 2048 x 1536 Landscape 3280 x 2048 Landscape 1640 x 2048 (dual) Portrait 1536 x 2048 (dual)‡ Portrait 2x 2048 x 2560 (non-span mode) Portrait 4096 x 2560 (horizontal span mode) Landscape Dome S3c Dome S6c Dome S10 * † ‡ 32-bit palette is used for 256 and 1024 shades of gray.
Power Input Power is supplied to DomeS3c, Dome S6c, and Dome S10 displays via the 6-pin DIN port on the connector plate, and via an 8-pin DIN port on the Dome S3 display. See pin assignments on the following illustrations. For Dome S3c, Dome S6c, and dome S10, the power input is 24V ±5% (150 W); for Dome S3, 12V ±5% (130 W). IMPORTANT! Use only the power adapter that ships with the display unit.
LED Status Lights For Dome S3, Dome S3c, and Dome S10, the two LEDs on the back panel provide information on the status of the display. • LED A describes the digital-link status between the display controller and the interface. • LED B describes the display-panel status. it shows any faults currently in the panel.
The LEDs on the back panel of the Dome S6c display provides information about the status of video and power supply connections to the display. LED A (left) and LED B are located between the power port and the DisplayPort port on the connector plate The LED A set describes the status of the video connection between the display controller and the display interface. LED A Color Video Cable Connection Top indicator Green Primary DVI connector is plugged into the computer.
Troubleshooting Problem Possible Cause Solution No image appears on the screen Computer is OFF. Power on the machine. Power cord is not securely connected. Tighten power cord connection and turn on computer. Video cable connected incorrectly. Make sure the first display of a dualheaded system or the only display of a single-headed system is connected to the primary port on the display controller. (NOTE: Both DVI and DisplayPort functions cannot be used at the same time.
Regulatory Compliance Canada, European Union, United States This display has been tested and found to comply with IEC 60601-1:2005 and EN 60601-1:2006, and IEC/EN 60601-1-2 standards, and is certified to meet medical standard CAN/CSA C22.2 No. 60601-1-A and UL 60601-1 ANSI/AAMI ES60601-1:2005 (R2012).
We offer state-of-the-art displays suitable to the European market, CE-marked displays based on compliance with counsel directive 93/42/EEC—commonly referred to as the Medical Device Directive (MDD). The following summarizes our qualification of these displays as it relates to compliance with the MDD. The European Medical Device Directive requires that the intended use of the device be defined.
Furthermore, the applicable safety standards for an MDD Class IIa display are EN 60601-1:2006 and IEC 60601-1:2005. To help the medical device designer evaluate the suitability of these displays, NDSsi has also conducted EMC testing to IEC 60601-1-2 as it can be applied. The display with its power supply alone does not represent a functional medical device. Hence, NDSsi configured a minimal operating system to exercise the display. The resulting data are made available to interested parties.
EU Declaration of Conformity for Medical Application A Declaration of Conformity has been filed for this product. For additional copies of the Declaration of Conformity document, contact NDS Surgical Imaging.
EU Declaration of Conformity for Medical Application A Declaration of Conformity has been filed for this product. For additional copies of the Declaration of Conformity document, contact NDS Surgical Imaging.
U.S. FCC Compliance Statement Class A Digital Device: Dome S3, Dome S3c, Dome S6c, or Dome S10 Display This device complies with Part 15 of the FCC Rules. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
Model Dome S series needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS; Portable and mobile RF communications equipment can affect the Dome S series. 1. ELECTROMAGNETIC EMISSIONS Guidance and Manufacturer’s Declaration—Electromagnetic Emissions The Dome S series is intended for use in the electromagnetic environment specified below.
2. ELECTROMAGNETIC IMMUNITY Guidance and Manufacturer’s Declaration 211; Electromagnetic Immunity The model Dome S series is intended for use in the electromagnetic environment specified below. The customer or the user of the display should assure that it is used in such an environment.
3. ELECTROMAGNETIC IMMUNITY Guidance and Manufacturer’s declaration–Electromagnetic Immunity The model Dome S series is intended for use in the electromagnetic environment specified below. The customer or the user of the Dome S series should assure that it is used in such an environment.
4. RECOMMENDED SEPARATION DISTANCES Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Dome S Series The model Dome S series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Each Dome Sx display has one or more of the following certifications. China China Compulsory Certification regulating safety and EMC. GB4943 GB9254 GB17625 Class A Dome S3, Dome S3c, Dome S6c, and Dome S10 Germany TÜV Rheinland GM Mark for medical products. Japan Voluntary Control Council for Interference by information technology equipment sold in Japan. Class A Dome S3, Dome S3c, Dome S6c, and Dome S10 Class B Korea Korea Certification. Dome Sx Display User’s Guide, 020-0931-00 Rev.
NDSsi Dome® Display Products Warranty and Service Terms and Conditions STANDARD LIMITED WARRANTY COVERAGE: NDS Surgical Imaging, LLC (hereinafter "NDSsi") warrants this product to be in compliance with the specifications provided by NDSsi and to be free from defects in material and workmanship as defined in such specifications. Subject to the conditions set forth below, NDSsi agrees to repair or replace any defective part of the enclosed unit for the length of period indicated on the chart below.
EXCLUSIONS – WHAT IS NOT COVERED 1 Any product with a defaced, modified, or removed serial number. 2 Damage, deterioration or a malfunction resulting from accident, misuse, neglect, fire, water, lightning, or other acts of nature, unauthorized product modification, or failure to follow instructions supplied with the product. 3 Cosmetic damage including, but not limited to scratches, cracks, dents, markings, glue and adhesive residue. 4 Any damage of the product due to shipment.
NDSSI’S LIABILITY IS LIMITED TO THE COST OF REPAIR OR REPLACEMENT OF THE PRODUCT. NDSSI SHALL NOT BE LIABLE FOR THE FOLLOWING: 1 DAMAGE TO OTHER PROPERTY CAUSED BY ANY DEFECTS IN THE PRODUCT, DAMAGES BASED UPON INCONVENIENCE, LOSS OF USE OF THE PRODUCT, LOSS OF TIME, LOSS OF PROFITS, LOSS OF BUSINESS OPPORTUNITY, LOSS OF GOODWILL, INTERFERENCE WITH BUSINESS RELATIONSHIPS, OR OTHER COMMERCIAL LOSS, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
CONTACTING NDSSI CUSTOMER CARE PRODUCT LINE Dome® TOLL FREE (866) 961-9340 LOCAL (503) 620-3787 E-MAIL medicalsupport@ndssi.com HOURS 6 A.M. – 5 P.M. Pacific Time, Monday – Friday 1 Contact NDSsi Medical Support during the hours listed above. Have your NDSsi model number and serial number available. You may be required to provide proof of purchase as a condition for receiving warranty service. 2 Our representative may perform troubleshooting and diagnosis of the problem by telephone or e-mail.
PAYMENT FOR NONWARRANTY EVALUATIONS AND REPAIRS AND OTHER EVALUATION CHARGES: Nonwarranty repairs requested by customer will be performed at NDSsi’s then current rates for any nonwarranty repairs and will include applicable evaluation fees and the cost of return shipping (if customer requests that NDSsi ship the repaired product).
ITEM ITEM NUMBER DESCRIPTION Dome S3 display 933-0993 3-megapixel grayscale display. Dome S3c display 933-0992 3-megapixel color display. Dome S6c display 933-1014 6-megapixel color display. Dome S10 display 933-0989 10-megapixel grayscale display. Power adapter, 6-pin 902-0176 Power block (24V, 150W) with 6-pin DIN connector for use with the Dome S6c display. Power block (24V, 150W) with 6-pin DIN connector for use with the Dome S3c, Dome S6c, or Dome S10 display.
Index proper disposal 19 proper handling 17 recycling instructions 12 regulatory compliance 33 resolution 28 safety precautions 15 shipping/storing 19 system requirements 5 technical information 12 troubleshooting 32 warranty and service terms A about display 5 auto-conformance 7 C certifications 43 cleaning display 18 components, packaged 4, 49 connecting power cord 10 video cable 10 connector plate 8, 29 customer support 47 D DIN connector 29 display about 5 classification 12 cleaning 18 components 7
P S package contents, display 4 power adapter 21, 23, 27, 29 DIN connector 29 input 29 power cord, connecting 10 power supply 21, 23, 27 precautions 15 preventive measures 18 product handling 17 safety precautions 15 shipping/storing display 19 specification Dome S10 display 26 Dome S3 display 20 Dome S3c display 22 Dome S6c display 24 symbols, explained 13 system requirements 5 R technical information 12 tips, safety 16 troubleshooting 32 recycling instructions 12 regulatory compliance 33 Canada, EU,
NDS Surgical Imaging, LLC 5750 Hellyer Avenue San Jose CA 95138-1000 USA www.ndssi.