User`s guide
52 | Dome CXtra
Configuring the Warning Light Indicators
NOTE: The Dome S10 display has LED warning light indicators to
notify you of the need to run MQSA quality control. Each indicator
glows in one of two colors—yellow or blue—by one of four
patterns—slow blink, double blink, fast blink, or solid.
The LEDs are built into the MQSA button located on the lower right
underside, beneath the display ID. Both the MQSA button and the
functionality for setting the warning lights are available only for
Dome S10 display installations.
A yellow or blue glowing pattern means a test is due. No glowing of
either color means all quality control activity is completed until the
next due date.
The LEDs are ranked by priority of tests. For example, if two categories
of tests are due and using the same indicator color, the indicator of
the higher ranking test activates. Upon completion of that test, the
indicator activates for the lower ranking test.
First Quality Control for Medical Physicist— Quarterly,
Biannually, or Annually
Second Monthly
Third Daily
LEDs built into the MQSA button
the Dome S10 display glow by
preset color and pattern to indicate
a quality control test is due