Front Sensor tab
Specify the front sensor conformance test(s) you want RightLight to
run. You can schedule testing to run automatically each day and
conduct testing manually.
RightLight alerts you of a failure in these three ways: by message, by
entries in the History log, and by status change of the Dome CXtra
icon in the system tray. If the CXtra-enabled workstation is connected
to Dome Dashboard, an alert is also sent to the Dashboard app or
SNMP console.
To set auto alert parameters
Click the selection box for the automatically scheduled and manual
tests you want to run.
• For the RightCheck White Level test, enter a value for Tolerance ±.
Then check the scheduled or manual selection box, or both, for test
alerts. The RightCheck White Level test is classified a failure if the
measured white level is outside the range defined by this percent,
applied to the Target Luminance defined on the Backlight tab.
Enter a value ranging from 0 (zero) to 25 (default = 10.00). Tests
producing a white level value outside of this percentage range
generate a warning message.
EXAMPLE: Target white level: 600 cd/m
2
Tolerance: ±10% Result: Values outside the range from 540 to 660
(600 ± 60 cd/m
2
) indicate a test failure.