User`s guide

16 | Dome CXtra
You can also calibrate the internal RightLight photometer to generate
a new calibration data file whenever the white level is outside the
tolerances set for the display configuration. Calibrating the white
level requires an external photometer.
For Dome Ex, Dome GX4MP, and Dome Sx color displays, you can set
color coordinates for panel appearance in one of two ways:
Select an option for panel appearance with blue, clear, or default.
Use the LCD II photometer to perform color calibration
when you specify color coordinates by color temperature
(Kelvin scale) or by CIE xy coordinates.
NOTE: Setting nondefault color coordinates can change the target
luminance range.
CXtra Services menu, and commands of the RightLight Conformance submenu