Dome CXtra ® User’s Guide ™
© 2014 NDS Surgical Imaging, LLC. All rights reserved. This document contains proprietary information of NDS Surgical Imaging. It is exclusive property of NDS Surgical Imaging. It may not be reproduced or transmitted, in whole or in part, without a written agreement from NDS Surgical Imaging. No patent or other license is granted to this information. The software, if any, described in this document is furnished under licensed agreement.
Contents What’s New in Version 6.
DICal Service 81 Configuring the DICal Service 82 Using DICOM Response Function as Default Calibration Choosing Display Calibration 84 Performing a Custom DICOM Calibration 85 Reviewing Calibration Events 87 Performing a Conformance Test 88 Reviewing Conformance Test Data 88 Reviewing Conformance Test Events 92 Advanced Options 94 Test Patterns Service 96 AAPM TG18 Test Patterns 96 Miscellaneous Test Patterns 96 Displaying Test Patterns 99 Adding Custom Test Patterns 100 Modifying or Deleting Custom Test Pa
What’s New in Version 6.x Series Dome Displays Dome CXtra support has been added for the Dome S6c display, a 6-megapixel color unit with DVI video connectivity, and the Dome GX24 display, a 2-megapixel unit with DisplayPort and DVI video connectivity.
MQSA Test Service You can set up test reminders for consistent quality control activity and record the results of the quality control tests electronically. NOTE: Applies to Dome E5 and Dome S10 displays only. Expansion of Manual Conformance Testing With the addition of the front sensor on some models of Dome displays, the Manual Conformance functionality has been bifurcated into the External Sensor Conformance testing (using the photometer device) and the Front Sensor Conformance testing.
Windows Platform Support Support is available for the Microsoft Windows 7 (x86 or x64) and Windows Server 2008 R2 (x64 only) operating systems. Support for the Microsoft Windows XP x64 operating system was added to the Dome CXtra 6.x series in an earlier version. CXtra Utilities Users can start, restart, and stop all CXtra services and application, and run CXtradiag with utilities. Access the tools via the Windows Start menu. NOTE: Windows administrator privilege is required to run the utilities.
NOTE: Privileged users and administrators can calibrate the white level if a display fails the White Level test during the manual conformance testing. Only administrators can calibrate the white level via the White Level Calibration service. Support for 10-Bit Displays You can now run the Dome CXtra software on Dome 10-bit grayscale displays. You can check for 10-bit capability (both current and maximum supported) on the Properties tab of the Reporting dialog.
Getting Started The Dome CXtra software adds essential functionality to Dome medical imaging display systems in diagnostic and referral settings. By using the Dome CXtra suite of value-added services, you can calibrate and maintain displays, run acceptance tests, and monitor displays remotely. While the services operate as independent plugins, they each use the Reporting service for listing properties, logging events, and saving history.
CXtra Services The following services constitute the Dome CXtra software. RightLight™ – Assures backlight stabilization for your display panel, calibrates your panel to the DICOM Grayscale Standard Display Function, automatically tests the calibration for conformance, allows manual conformance testing, and supports DICOM calibration at various color coordinates on Dome color displays. (Manual external sensor conformance testing requires the LCD or the LCD II photometer.
Privilege – Password-protects operating parameters of the display system. Reporting – Collects, displays, and prints configuration data, characteristics, and event logs created by most other services. Enterprise Management – Manages displays from a central location. Requires SNMP console (not provided) for remote viewing. The use of CXtra services is display-dependent.
System Requirements The Dome CXtra software requires one of these operating systems: • Windows 7 (x86 or x64) • Windows Server 2008 R2 (x64 only) • Windows XP Professional (x86 or x64), Service Pack 3 Installing the Dome CXtra Software NOTE: Install the Dome CXtra software to a local disk drive only. The software may not work properly if any components are installed on a network drive. To install the Dome CXtra software 1 Log onto Windows with administrator privileges.
UPGRADE INSTALLATION: To upgrade to package 6.0.1 or later, insert the installation CD and run setup.exe. Follow the wizard instructions to complete the installation. Services that have been modified since the previous release are upgraded. The application starts when the installation is finished. The Dome CXtra icon appears on the Windows taskbar. The check mark indicates that the application is running successfully.
Installing the Photometer To use the RightLight and DICal custom calibration and conformance test features, you must purchase and install the optional LCD or LCD II photometer. (Only the LCD II photometer is RoHS compliant.) NOTE: The LCD II photometer is required for use with Dome color displays. Use the following instructions to install the photometer correctly per your operating system. To install the photometer on Windows 7/Windows Server 2008 R2 1 Log onto Windows with administrator privileges.
To install the photometer on Windows XP 1 Log onto Windows with administrator privileges. 2 If necessary, install the Dome CXtra software. (See page 8.) 3 Connect the photometer to a USB port on the computer. The operating system detects the new hardware and opens the Found New Hardware Wizard. NOTE: If you connect the photometer to a USB hub and the Found New Hardware Wizard does not appear, connect the photometer directly to a USB port on the computer.
Starting Dome CXtra The Dome CXtra application starts automatically when you run your computer. If you exit the application or if it stops running for any reason, restart it by selecting Windows Start > All Programs > NDSsi > CXtra. A Dome CXtra status icon appears on the Windows taskbar to indicate application status. The Success icon shows that the display is operating normally. The Wait icon appears the application is starting up or restarting from a stop.
Opening the CXtra Services Menu All Dome CXtra functions are accessible from the services menu. This graphic shows the default menu entries for standard services. To open the menu, click the Dome CXtra icon on the Windows taskbar. If the Dome CXtra icon is not visible on the Windows taskbar, open the menu by Windows Start > All Programs > NDSsi > CXtra. The menu is expanded to show the White Level Calibration service when the user is logged on as administrator.
CXtra Utilities The following utilities have been added to the application. CXtra Start Launch the application and all CXtra services CXtra Diag Create Cxtradiag package CXtra Restart Resume application operation and all CXtra services CXtra Stop Terminate the application and all CXtra services To access a utility, click Windows Start > All Programs > NDSsi > CXtra Utilities, then select one of the four options. NOTE: Windows administrator privilege is required to run the utilities.
RightLight Service The RightLight service performs backlight stabilization, DICOM calibration, automatic conformance testing, and manual conformance testing via an external photometer for Dome Ex, Dome GX4MP, and Dome Sx displays. Manual conformance testing via the display front sensor is available for Dome Sx units. With RightLight, you can set a target white level, equalize the black level across the displays, set a desired response function, and specify ambient light in the viewing area.
You can also calibrate the internal RightLight photometer to generate a new calibration data file whenever the white level is outside the tolerances set for the display configuration. Calibrating the white level requires an external photometer. For Dome Ex, Dome GX4MP, and Dome Sx color displays, you can set color coordinates for panel appearance in one of two ways: • Select an option for panel appearance with blue, clear, or default.
Configuring the RightLight Service If the Privilege service has been installed, you must log on as a privileged user to change the RightLight configuration. (See page 138.) To configure the RightLight service 1 Open the CXtra Services menu and select RightLight Panel Configuration.
Backlight tab For displays selected, you can view the values of these parameters: • Black level, in cd/m2 • White level, in cd/m2 • Backlight use, in hours • Brightness, as a percentage of maximum effective brightness In the White Level field, type the target white level value (in cd/m2) or use the arrow keys to change the value shown. Click Apply or OK. NOTE: White Level values are limited to the upper and lower ranges specified per display.
NOTE: If you set the white level too close to the maximum brightness capability of the display, frequent warnings on high white levels may appear. To correct this problem, reduce the white level setting on the Backlight tab until the warnings stop appearing or use the CXtra Preference dialog to disable the warning popups. If you have a requirement to exceed the maximum limit, you can do so but you void your display warranty. Contact NDSsi Customer Care for instructions.
Calibration tab Select one of these response functions for auto-calibration: • DICOM Response Function (default) – Calibrates to the DICOM Grayscale Standard Display Function. See DICOM PS 3.14-2000 (Digital Imaging and Communications in Medicine (DICOM), Part 14: Grayscale Standard Display Function). • Exponential (log linear) Response Function – Calibrates to an exponential (log linear) response function. The logarithm of the luminance is a linear function of the digital driving level.
Ambient tab Specify the ambient light in the viewing area. Select the ambient light description that best fits the display location: • Low (reading room - 0.20 cd/m2) • Moderate (typical office - 0.80 cd/m2) • High (brightly lit office - 1.50 cd/m2) • Custom – Fill in a value ranging from 0.00 (default) to 5.00 cd/m2. The RightLight service allows you to enter luminance in cd/m2 resulting from the diffuse reflection of ambient illuminance. Since typical Dome panels have a reflectance of 1.
Color tab Specify the panel appearance of Dome Ex, Dome GX4MP, and Dome Sx color displays, with or without calibration. Appearance. Use options in the Appearance field to specify blue, clear, or default when you do not color-calibrate the display panel. Calibrated color coordinates. Use options in the Calibrated Color Coordinates field for precise specification of color coordinates.
Front Sensor tab Specify the front sensor conformance test(s) you want RightLight to run. You can schedule testing to run automatically each day and conduct testing manually. RightLight alerts you of a failure in these three ways: by message, by entries in the History log, and by status change of the Dome CXtra icon in the system tray. If the CXtra-enabled workstation is connected to Dome Dashboard, an alert is also sent to the Dashboard app or SNMP console.
• For the DICOM LUM test, enter a value for Maximum allowed. Then check the scheduled or manual selection box, or both, for test alerts. The DICOM LUM test is classified a failure if it produces a value greater than the maximum allowed. Enter a value ranging from 0 (zero) to 2 (default=1.00). A value of 0 (zero) indicates perfect conformance with the DICOM Grayscale Standard Display Function. The DICOM LUM Test assesses the standard deviation of JNDs (Just Noticeable Differences) per digital interval.
Rear Sensor tab Specify the conformance test(s) you want RightLight to run. Every 5 minutes, RightLight automatically performs the tests selected on this tab. For each test you select, you can specify a threshold value that defines a test failure. RightLight alerts you of a failure in these three ways: by message, by entries in the History log, and by status change of the Dome CXtra icon in the system tray.
For the DICOM LUM test, check the selection box. Then enter a value for Maximum allowed. The DICOM LUM test is classified a failure if it produces a value greater than the maximum allowed. Enter a value ranging from 0 (zero) to 2 (default=1.00). A value of 0 (zero) indicates perfect conformance with the DICOM Grayscale Standard Display Function. The DICOM LUM Test assesses the standard deviation of JNDs (Just Noticeable Differences) per digital interval. EXAMPLE: Maximum allowed: 1.
External Sensor tab Specify the external sensor tests that you want to complete manually. For each manual test that you select, you can specify a threshold value that defines a test failure. When a failure occurs, the CXtra icon is set to a warning state. In addition, if the CXtra-enabled workstation is connected to Dome Dashboard, an alert is also sent to the Dashboard app or SNMP console. Selections made on this tab determine operating parameters for the Conformance Wizard, which guides the manual test.
2 Select the number of luminance measurements you want to make during each manual test. Note: For the external sensor, the default number of measurements is 15. When you select All, 256 measurements are taken on 8-bit display systems; 1024 measurements are taken on 10-bit systems.
Checking Conformance via the Front Sensor Front sensor conformance tests are run per criteria you input for auto-scheduled or manual testing. If you have the Privilege service installed, you must log on as a privileged user before you run a front sensor conformance test. (See page 138.) To run the Front Sensor Conformance Test on demand, open the CXtra Services menu, select RightLight Conformance > Front Sensor Conformance Test. The test dialog appears onscreen while the conformance check is in progress.
Auto Conformance Testing via the Rear Sensor Automatic conformance testing shows how well your display conforms to the DICOM standard. RightLight automatically performs the tests selected on the Rear Sensor tab every 5 minutes. See page 25 to view or modify the test parameters. If you have the Privilege service installed, you must log on as a privileged user before you set up auto conformance testing. (See page 138.
To run an external sensor conformance test 1 Open the CXtra Services menu. Select RightLight Conformance > External Sensor Conformance Wizard. 2 Click Next to continue. 3 Click Initialize RightLight Panels. Upon completion of the process, click Next to proceed.
4 Put the photometer on the target and click Measure. If the White Level is unaligned with the target by 15 percent or more, a warning dialog appears. You can choose to perform White Level calibration to the target, ignore the warning and continue the measurement of conformance data, or direct the Conformance Wizard to skip the display panel, redo the initialization, or close.
5 Upon successful completion of the conformance measuring, as indicated by the Success icon, click Next. You have the option to view the report via this dialog. 6 Click Finish to close the wizard.
After you close the Conformance Wizard, RightLight enters the event in both the Reporting History log and the Conformance History log. All RightLight history events from conformance tests are recorded in the Reporting History log. RightLight provides additional information on conformance events in graphical and tabular formats in the Conformance History log.
Checking Current Conformance Status To view the results of the latest manual (external sensor or front sensor) conformance test, open the CXtra Services menu. Select RightLight Conformance > Current Conformance Status. In the same dialog, you can also view the results of the latest automatic (rear sensor) conformance test.
Reviewing Conformance Events Dome CXtra polls RightLight automatically for its status every 5 minutes, which causes the user-selected tests to run. The system makes an entry in the Reporting History log by the specified logging interval on the Auto Alerts tab and whenever it detects a status change. NOTE: When you make changes on the Rear Sensor tab, conformance tests run immediately. When you perform a manual conformance test, the system also makes an entry in the Reporting History log.
Reviewing Manual Conformance Data For manual conformance tests run via the front or external sensor, you can view data in graphical or tabular format by selecting sensor type, display model, and date. Not the following: • You must be using a Dome Sx display to view test results from both front and external sensors; select one or both check boxes on the dialog. With a Dome Ex display, all data result from conformance testing with the external (or manual) sensor.
Graph type • Response graph. Plots the measured luminance at various driving levels (P-values). Luminance is shown along the Y axis using a scale that you select from three options: Linear / Lum, Linear / Log Lum, and Linear / JND Lum. The graph consists of the luminance values specified by the DICOM standard and the luminance values measured during the conformance test. The better the data points align with the curve, the better the DICOM conformance.
• Conformance graph. Plots Just Noticeable Differences (JNDs) per luminance interval. The graph shows the mean value for JND/ luminance interval, the DICOM Luminance Uniformity Metric (LUM), and the line of best fit (FIT). These values are useful in quantifying the quality of conformance with the DICOM Grayscale Standard Display Function. The closer the JND values are to the mean line, the better the DICOM conformance.
Tables tab To view or print a conformance test, choose a table type, display, and selected test date and time to view or print the tabular data. When using multiple displays, choose the individual display for which you want test results from the Display pull-down list. Select a particular conformance test by date and time from the Date pull-down list. Table type • Response table. Lists the luminance value measured for each luminance interval in a selected conformance test. • Conformance table.
• Color table. Lists the following data values by Kelvin scale or CIE xy coordinates.
Disabling RightLight Popup Warnings You can disable all warning message popups related to the RightLight service. NOTE: Disabling of warning message popups is not recommended. Read the notice on page 43 before disabling any popups. To disable RightLight popup warnings 1 Open the CXtra Services menu and select CXtra Preference. 2 Check the box in the Disable RightLight Popups field. 3 Click OK.
Important notice on disabling RightLight popup messages WARNING: The Dome CXtra software includes "popup" warning dialogs (popups), which, unless disabled, appear on the operator's display screen under various operational conditions. These warnings include important messages about the performance of the display which may impact the integrity and reliability of the information displayed, such that the accuracy of the radiologist's or other medical user's diagnosis may be compromised.
White Level Calibration Service IMPORTANT: Only administrators can calibrate the white level via the White Level Calibration service and restore factory calibration. NOTE: White level calibration is available for use on all Dome Ex, Dome GX4MP, and Dome Sx displays, except the Dome S6c display. Restoration of the factory white level calibration is available for these same displays, except the Dome S6c display.
Privileged users can calibrate the white level during manual external sensor conformance testing. For more information, see “To run an external sensor conformance test” on page 31. When you log on as administrator, the CXtra Service menu expands to include two functions, White Level Calibration Wizard and Restore Factory White Level Calibration. (See “Logging On as Privileged User/ Administrator” on page 138.
Running the White Level Calibration Wizard IMPORTANT: White level calibration via the White Level Calibration service is available for use on all Dome Ex, Dome GX4MP, and Dome Sx displays, except the Dome S6c display. Restoration of the factory white level calibration is also available for these Dome Ex, Dome GX4MP, and Dome Sx displays, except the Dome S6c display. To calibrate the white level First, open the CXtra Services menu and click White Level Calibration Wizard.
Restoring Factory White Level Calibration IMPORTANT: Restoration of the factory white level calibration is available only if you have previously calibrated the white level using the White Level Calibration service or via RightLight external sensor conformance testing. Factory calibration can be restored for Dome Ex, Dome GX4MP, and Dome Sx displays, except the Dome S6c display. To restore factory calibration Open the CXtra Services menu and select Restore Factory White Level Calibration.
MQSA Quality Control Tests Regular inspection and testing of diagnostic imaging displays designated for primary image diagnosis in mammography ensure display calibration and quality control. Only the Dome E5 and Dome S10 displays can be used for image diagnosis as regulated by the Mammography Quality Standards Act, or MQSA.
• Calibrated external photometer • Calibrated luxmeter with a minimum resolution of 1 lux • Mini-USB cable (Dome S10 display installations only) You can conduct three categories of tests—daily, monthly, and quality control for medical physicists (quarterly, biannually, annually options included)—to maintain the display. First, configure the software for test reminders.
Configuring MQSA Test Reminders Quality Control tests are simple checks that must be performed regularly to ensure the Dome display is operating to design standards. These tests are meant to detect any changes that might compromise image quality, as well as any deterioration in equipment performance that may occur over time. Setting test reminders facilitates the MQSA quality control activities. You must log on as a privileged user to set up or modify the MQSA test configuration.
To configure the MQSA test reminders 1 Open the CXtra Services menu and select MQSA Test Configuration. 2 Optional: Check the selection box to enable Expert Mode. This mode allows radiological technicians and medical physicists already familiar with evaluation of the SMPTE test pattern to quicken visual inspection of the display screen. 3 Check the selection box to enable the functionality for MQSA test reminders. NOTE: Reminders are issued only for the test(s) that you enable on the dialog.
Configuring the Warning Light Indicators NOTE: The Dome S10 display has LED warning light indicators to notify you of the need to run MQSA quality control. Each indicator glows in one of two colors—yellow or blue—by one of four patterns—slow blink, double blink, fast blink, or solid. The LEDs are built into the MQSA button located on the lower right underside, beneath the display ID.
To configure the warning lights for test reminders 1 Open the CXtra Services menu and select MQSA Test Configuration. 2 To set the indicator for a test reminder, click the warning light button in that field. (NOTE: The default or previously selected pattern appears on the button face. For Dome S10 display only: When you hold the mouse over the warning light button, you can preview the selected light pattern via the MQSA button.) 3 Choose a color and select a pattern for that color.
MQSA Test Wizard When you configure the MQSA test reminders, you can see due status of tests to perform via LED indicators in the MQSA button (on Dome S10 displays) and by the test wizard accessible via the CXtra Services menu. Depending upon the display you are using or how you access the test wizard—by the MQSA button (only on the Dome S10 display), the CXtra Services menu, or the Windows Start menu—the appearance of the wizard may vary.
MQSA Test Wizard that appears via the MQSA button (Dome S10 display only) and via the Windows Start menu (Dome E5 and Dome S10 displays) The Ignore Tests/Reset Ignore Tests toggle button is visible only when one or more tests are due After you run a test, that test button is grayed out on the wizard accessible via the MQSA button and the Windows Start menu. You can rerun the test before the next due date by accessing the wizard via the taskbar.
Running the Daily Quality Control Test To ensure optimal image viewing from day to day, inspect the display to keep the screen free of dust, fingerprints, and other marks. Frequency. Clean the display daily or on days when clinical image reviews are scheduled. Materials required. You need only a dry, soft, lint-free cloth or cleaning tissue. If necessary, the cloth or tissue may be lightly dampened with water or a mild cleanser before use. Never apply water or other fluids directly to the display.
3 Use a dry, soft, clean cloth to wipe off the residue. NOTE: Always wipe off water and other stains immediately to avoid damage to the screen. Do not exert excessive pressure on the display screen as this can damage the LCD module. 4 Check the screen to verify that it is free of dust, fingerprints, and other marks. The screen must be free of dust, fingerprints, and other marks.
3 Click the selection box to record the clean status of the display(s). Click the Check All Clean button to select the Clean check boxes simultaneously for a multiple display workstation. Click Uncheck All Clean button to remove the clean status or to be able to select the clean status for individual displays. NOTE: Hover the mouse over the italic text to read the tooltips for materials required and procedures.
If any conditions are the daily quality control test are not met, this dialog appears when you click Submit. Return to the quality control test dialog and follow procedures to remedy any deficiency. Or, click Ignore if you know for sure that the deficiency does not impede display quality or performance.
Running the Monthly Quality Control Test The monthly display calibration check is to ensure that the display is calibrated and the brightness and contrast settings are at an appropriate level for the reading of mammographic images on the Dome display. The calibration check was designed to be performed with the SMPTE pattern built into the Dome CXtra software.
To complete the monthly control test record 1 Click Monthly Quality Control Test on the wizard. The SMPTE pattern appears on each display connected to the system. 2 Examine the display screen. Click the Check All button if all conditions are met. Click Next to inspect another display.
3 Repeat the procedure for any other display(s). Then click Submit to finish. 4 From either the Normal Mode or the Expert Mode, enter tester’s name to complete process. Click Submit to finish.
5 Click View Report or Close. You can view the monthly quality control test report on the screen or print it. Click Close when finished. Then clear the Test Completed dialog from the screen.
If any conditions are the daily quality control test are not met, this dialog appears when you click Submit. Return to the quality control test dialog and follow procedures to remedy any deficiency. Or, click Ignore if you know for sure that the deficiency does not impede display quality or performance. Expert Mode When you select the Expert Mode option on the MQSA Test Configuration dialog, you can quicken the visual inspection of the monthly quality control test. The following dialog appears.
Running the Quality Control Tests for Medical Physicists The Quality Control tests listed in this section must be performed by the medical physicist to ensure that the Dome display provides a high level of mammographic image quality.
Site and System Summary Start by recording the site and system data for the facility. To record the site and system data 1 Click the Site and System Summary button on the wizard. 2 Fill in the facility ID, address, and display location information. 3 Use the pull-down arrows to enter the date of installation and the date of survey. 4 Enter the name of the medical physicist. 5 Click Submit to finish.
Return to the wizard to start the next test. Viewing Condition Check and Setting Compare the reading room configuration established as optimal with the actual one. If differences exist between this configuration and the actual one, adjust the room configuration to match that shown in the drawing. For example, if desk lights were described to be turned off, turn them off. Or, if the curtains were described to be closed, close them. You need a calibrated luxmeter with a minimum resolution of 1 lux.
NOTE: Look for the nominal value of the ambient light. This is a record of the most recent measurement and provides a comparison value for the next medical physicist survey. If the reading room configuration has changed, the source of the problem must be identified and corrective action taken before any further mammographic images are reviewed or interpreted using the Dome display. To record the viewing conditions 1 Click Viewing Condition Check and Setting button on the test wizard.
Return to the wizard to start the next test. Display Calibration Check The calibration check is to ensure that the display is calibrated and the brightness and contrast settings are at an appropriate level for the reading of mammographic images on the Dome display. Conduct the display calibration at least annually. NOTE: You must power on the Dome display at least 30 minutes before doing any measurements. This is mandatory to ensure that the backlight heating has reached steady state.
To check display calibration 1 Click the Display Calibration Check button on the test wizard. 2 Click Next to access the External Sensor Conformance Wizard. 3 Click Next to launch the wizard. Be sure the photometer is already installed.
4 Place the photometer on an opaque surface, then click Initialize. When the initialization is completed, click Next when the button is enabled. 5 Place the photometer on the target, and click Measure. After the measurement have been taken, click Next when the button is enabled.
6 Upon successful test completion, click Next. If the conformance test fails, click the selection box to correct the warning. The wizard repeats.
Once you have completed a successful conformance test, return to the MQSA Test Wizard, and conduct the next test. Image Quality of SMPTE Pattern The quality control of the SMPTE pattern is to ensure the display is calibrated and the brightness and contrast settings are at an appropriate level for the reading of mammographic images on the Dome display. Conduct an image quality of the SMPTE pattern at least annually.
To inspect display screen quality When the SMPTE pattern appears on the display screen, examine the image carefully for the following features: • Verify that the 0%-5% contrast is visible. • Verify that the 95%-100% contrast is visible. • Verify that each gray level step from 0% to 100% can be distinguished from the adjacent squares. Verify that you can, for example, distinguish the 0% square from the 10% square or the 90% square from the 100% square.
To complete the control test record 1 Click Monthly Quality Control Test on the wizard. The SMPTE pattern appears on each display connected to the system. 2 Examine the display screen. Click the Check All button if all conditions are met. Click Next to inspect another display.
3 Repeat the procedure for any other display(s). Then click Submit to finish. 4 From either the Normal Mode or the Expert Mode, enter tester’s name to complete process. Click Submit to finish.
Return to the MQSA Test Wizard, and conduct the next test. Analysis of Screen Defects The visual inspection of screen defects is to determine if artifacts observed during system inspection can be attributed to the Dome display. Conduct the visual inspection during system acceptance, at least annually, and when necessary to isolate artifacts. To inspect the display screen for defects Follow these steps to inspect the display screen for defects.
• At a normal viewing distance (approximately from 13 to 20 inches) inspect each panel for visible white dots and other foreign debris. Record those that are not admissible for mammography reading. • Place the cursor anywhere on the desktop. Right-click to open the popup menu and select Properties. • Select the color White from the pull-down color palette. Click OK. All screens should display White.
The quality control testing for medical physicists is completed for the current cycle. Click Finish. Click View Report or Close.
You can view the quality control test report on the screen or print it. Click Close when finished, then clear any remaining dialog from the display screen.
DICal Service You can use the DICal service to adjusts the characteristic curve of the Dome GX24, Dome GX2MP Plus, and Dome GX2MP displays to conform to the DICOM Grayscale Standard Display Function. The configuration of the service provides an option for default DICOM calibration (for viewing diagnostic images), which is adequate for many applications.
Configuring the DICal Service The desktop display ships with brightness and contrast values already set (defaulted) for DICOM compliance, as shown in this sample table.
Using DICOM Response Function as Default Calibration Open the DICal Panel Configuration dialog to the Configuration tab. To activate DICOM grayscale calibration at each system startup, click the check box for DICOM response function. The default DICOM calibration is applied to your display settings whenever you select Desktop DICOM Calibration from the services menu. No other steps are needed to apply the DICOM calibration after installation.
Choosing Display Calibration Besides DICOM calibration, you can choose to use exponential, gamma, or original display calibration. Open the CXtra Services menu, then select DICal Panel Configuration and Calibration tab. Choose one of these response functions for auto-calibration: • DICOM Response Function – Calibrates to the DICOM Grayscale Standard Display Function. Refer to DICOM PS 3.14-2000 (Digital Imaging and Communications in Medicine (DICOM), Part 14: Grayscale Standard Display Function).
NOTE: Upon software installation, only Desktop DICOM Calibration and Desktop Original Calibration are default entries on the services menu. On the Calibration tab, select or deselect check boxes in the Menu field to customize calibration entries on the services menu. To switch to another calibration Open the CXtra Services menu. Choose from the display calibrations you have added to the services menu. Otherwise, go to the Calibration tab of the DICal Panel Configuration dialog to make a choice.
To perform a custom calibration 1 If you have one of the display models listed on the configuration dialog, follow instructions provided with the display to verify that the contrast is set properly. This sample table shows the contrast percentage for a set of the display models.
Reviewing Calibration Events The Dome CXtra software records an event in the History event log each time you perform a calibration. To review a calibration event 1 Open the CXtra Services menu and select Reporting. 2 Click the History tab to see a list of events. 3 Click a DICal service calibration event, then click View Detail to see detailed information about the event. NOTE: Only calibration and test status events have detailed information available.
Performing a Conformance Test You can use DICal service to verify that your display conforms to the DICOM standard. NOTE: This procedure requires installation of the optional LCD or LCD II photometer. (See page 10.) To perform a conformance test 1 If the Privilege service has been installed, log on as a privileged user. (See page 138.) 2 Open the CXtra Services menu and select DICal Conformance > Conformance Wizard.
Using the Graphs Tab Use the controls on the Graphs tab to select a conformance test and change how the system shows the data for a selected test. When using multiple display panels, select a particular panel in the Display drop-down list. Select a particular conformance test by selecting the data and time of the test in this drop-down list. Use Print Graph to output the graph currently on the screen. Graph Type • Response. Presents the luminance values at various luminance intervals (P-values).
• Conformance. Presents values of Just Noticeable Differences (JND) per luminance interval and the mean value for JND/luminance interval. These values include the Luminance Uniformity Metric (LUM - DICOM metric) value and the FIT line expression used to quantify the degree of conformance. In general, the closer the JND values are to the mean line, the better the DICOM conformance. In addition, the closer the calculated LUM metric is to 0 (zero), the better the DICOM conformance.
Using the Tables Tab The Tables tab shows response and conformance numerical data for a given conformance test. When using multiple display panels, select a particular panel in the Display drop-down list. Select a particular conformance test by selecting the data and time for the test in this drop-down list. Use Print Table to output the data currently on the screen. Table Type • Response. Presents the luminance value measured for each luminance interval in a selected conformance test. • Conformance.
Reviewing Conformance Test Events The Dome CXtra system records an event in the History event log each time you complete a conformance test. To review a conformance test event 1 Open the CXtra Services menu and select Reporting. 2 Click the History tab to see a list of events.
3 Click a conformance test status event, then click View Detail to see a summary of information about that test. NOTE: Only test events and calibration events have detailed information available.
Advanced Options When you connect a third-party monitor or unsupported display to your system, you see that the Dome CXtra software labels it “unknown” and “not” characterized in the DICal Panel Configuration dialog. The DICal service needs to characterize or recognize a display before it performs calibration. When the third-party monitor or unsupported display lacks a characterization data file, run the characterization wizard, then save the panel type with a unique name.
Select the data file to export, then store it in the Dome CXtra dat folder on the other system. Enabling DICOM Calibration on Third-Party Monitors After characterization, operate the third-party monitor with Desktop DICOM Calibration enabled. You can undo the Panel Type selection for third-party monitors using Edit Type. Choose and click OK.
Test Patterns Service You can view a variety of test patterns for assessing image quality using the Test Patterns service. Two categories of test patterns are installed with the service: • AAPM (American Association of Physicists in Medicine) TG18 • Miscellaneous patterns such as SMPTE, horizontal and vertical ramps, and geometric distortions You can also import custom test patterns in .DCM, .DC3, .JPG, or .TIF formats, as well as perform visual and interactive evaluations.
AAPM TG18 Test Patterns Set Series Type Images Description Multi Purpose (1K and 2K) TG18-QC Vis./Qnt 1 Resolution, luminance, distortion, artifacts TG18-BR Visual 1 Briggs pattern, low contrast detail vs. luminance TG18-CT Visual 1 Luminance response TG18-LN Quant. 18 DICOM grayscale calibration series TG18-UN Visual 2 Luminance and color uniformity and angular response TG18-UNL Quant.
98 | Dome CXtra Set Type Images Description Resolution, cont. TG18-LPH Visual 3 Horizontal bars at 1 pixel width, 1/16 modulation, 3 luminance levels TG18-LPV Visual 3 Vertical bars at 1 pixel width, 1/16 modulation, 3 luminance levels TG18-AFC Visual 1 4 AFC contrast-detail pattern, 4 CD values TG18-NS Quant. 3 Similar to RV/RH, 5 uniform regions for noise evaluation TG18-GV Visual 2 Dark spot pattern with low contrast object TG18-GQ Quant.
Displaying Test Patterns To select and display a test pattern 1 Open the CXtra Services menu and select Test Patterns > View Test Patterns. The Display Test Patterns dialog opens. 2 Click the plus sign (+) beside a test pattern set to see variations of the test pattern. NOTE: • Some AAPM TG18 patterns have variations such as 10%, 50%, and 80%. • On display systems configured for 10-bit imaging, you can view the horizontal and vertical ramps to see the differences in 8- and 10-bits per pixel.
3 Click a pattern or variation name to show that pattern. To invert the pattern (change black to white and white to black), click the Invert. (Some test patterns cannot be inverted.) 4 If necessary, click Hide Dialog to provide a clear view of the pattern. To restore the dialog, press the space bar. 5 To exit the test pattern and the dialog, click Close.
3 Click Add to open the Add Custom Test Pattern dialog. 4 Using the Look In drop-down list, navigate to the folder containing the pattern you want to add. 5 Select a .dcm, .dc3, .jpg, .tif, or .tiff pattern file. 6 If appropriate, select a view option: – Zoom to fit (Select a sizing method of Bilinear or Nearest – Neighbor if you choose this option.) – Ramp grayscale 7 Click Open. The selected pattern appears in the Add, Modify, and Delete a Custom Test Pattern dialog.
Modifying or Deleting Custom Test Patterns To modify or delete a custom pattern 1 If the Privilege service has been installed, log on as a privileged user. (See page 138.) 2 Open the CXtra Services menu and select Test Patterns > Manage Custom Test Patterns. The Add, Modify, or Delete a Custom Test Pattern dialog opens. 3 Select the test pattern to be modified or deleted. 4 Click one of the following buttons as needed: – Delete. Removes the custom test pattern. – View Options.
When invoked, the Test Patterns Evaluation dialog allows you to run the Dome Visual Test or the Dome Interactive Test or both. Use Setup to configure test parameters.
Run Dome Visual Test Use this test to display a visual pattern that you have selected (on the Setup dialog) and make a response to the resolution, uniformity, and visibility of details present in the pattern. The pass/fail status of the test is determined by your responses to the test questions. Run Dome Interactive Test With the interactive test, you can decipher textual strings that are shown on a background with very little contrast difference, 5 percent.
When a scheduled test pops up, the Preparing to Show Test Pattern dialog prompts you to select one of three options: taking the test, canceling the test, or postponing the test for a later time. If you decide to take the test later, the dialog pops up again after a time specified on the General tab of the Setup dialog. NOTE: Scheduled test do not pop up if any of the Test Patterns dialogs are open. Forced.
Patterns. Open this tab to configure the pattern type and location for the Dome Interactive Test and the pattern type for the Dome Visual Test. NOTE: Random is the default selection for all parameters of each test. Four types of patterns are supported by the Dome Interactive Test.
When Random pattern location is used by default or choice, you see the pattern in a random location each time the test is invoked. These six patterns are supported by the Dome Visual Test: • SMPTE • TG18-GV • TG18-QC • DIN-1 • TG18-CT • DIN-2 When Random pattern type is used by default or choice, you see a randomly selected pattern each time the test is invoked. Open this tab to configure these three parameters: • Prompt for Tester name: Enter your name after test completion.
Backlight Saver Service The Backlight Saver service allows you to specify when the display system turns off the panel backlights to extend the display’s clinical life. This service allows you to specify an inactivity value in minutes; the panel backlights turn off whenever a period of inactivity exceeds this value. The service lets you define one or more “saver periods” during each day of the week. You can specify a different inactivity or “sleep after” value for each saver period you define.
2 Select the day or days for which you want to create one or more saver periods. Click a single day name in the Selected Day column or hold down the Control key while clicking multiple days. For example, you might specify the same saver periods for weekdays (Monday through Friday) and different saver periods for weekends. 3 Click Add Period. 4 Define a saver period by entering values for Start Time and End Time. Entries must be in a 24-hour format. NOTE.
To delete an existing saver period 1 Open the CXtra Services menu and select Backlight Saver. 2 Click the associated data line and click Delete Period. To view the saver periods for a week 1 Open the CXtra Services menu and select Backlight Saver. 2 Click View Week to see the schedule. 3 Click OK. Notes You can overwrite the current Backlight Saver period by changing the Windows power options within any defined period.
DIN Test Service The DIN Test service tests your Dome displays for conformance with German standard DIN 6868-57. A successful acceptance test such as the one performed by DIN Test service is mandatory before DIN 6868-57 application category A and B devices* can be accepted in a hospital or other diagnostic setting. For more information about DIN acceptance criteria, obtain a copy of the DIN 6868-57 standard through the following web site: http://www2.din.de.
Acceptance or constancy testing requires a photometer for measuring the luminance of the display. The DIN Test service automatically reads luminance measurements when you use one of these photometers: • Wellhofer LXplus • Minolta LS-100 NOTE: The NDSsi photometer is also read automatically, but it is a contact device and is less suited for use in DIN testing.
To set up the display for testing 1 Open the CXtra Services menu and select DIN Test. The DIN Test service dialog appears. 2 Click Setup. The DIN Test Setup dialog appears, open to the Location tab.
To set up entries on the Location tab Enter descriptive text and click Apply.
To set up entries on the Display tab 1 Complete the Display and Performance Class fields. 2 Select Yes or No for the three installation questions, select the display matrix require, and click Apply. NOTE: The descriptive information in the middle of the display is read-only and cannot be changed.
To set up entries on the Photometers tab 1 Select a photometer from the pull-down list. Select Other if your photometer is omitted from the list. NOTE: All photometers must adhere to section 7.1 of the DIN specification. When you select Other, you must manually enter the measured values in the test wizard dialogs. 2 If you selected Other in step 1, type the name of the photometer manufacturer in the Manufacturer field. 3 Enter the photometer serial number in the Serial Number field.
To set up entries on the Medical Image tab To use an image other than the default image (achest.dcm), click Browse and select an image. Click Apply. NOTE: A landscape image, chestah.dcm, is available in the installation folder’s dat folder.
To view information on the Reference Values tab Open the Reference Values tab to view information about the most recent successful DIN acceptance test. Click Apply. NOTE: This information is read-only.
To set up entries on the Constancy tab 1 Open the Constancy tab to set DIN constancy test parameters. The tab shows the previously saved constancy settings for the display identified in the Display field.
2 To use the settings in the DIN constancy specification (recommended), click Restore Defaults. NOTE: Using the default settings is recommended to assure conformance with the DIN specification. The default values for Veil Test Interval and the Contrast Ratio Test Interval depend on whether Luminance Control and Contrast Stability are shown as Present or Not Present. They also depend on whether you enable Constant Ambient Illumination.
To specify language on the Reporting tab 1 Open the Reporting tab and select your language, English or German. 2 Click Apply or OK.
Running a DIN Acceptance Test If the Privilege service is installed, you must log on as a privileged user before you run a DIN acceptance test. (See page 138.) To run a DIN acceptance test 1 Open the CXtra Services menu and select DIN Test. The DIN Test service dialog appears. 2 Click Run Acceptance Test. 3 Using the photometer identified in the setup procedure, follow the instructions provided by the test wizard.
Running a DIN Constancy Test You must successfully run a DIN acceptance test (described in the previous section) on the device being tested before you can run the constancy test described below. If the Privilege service is installed, you must log on as a privileged user before you run a DIN constancy test. (See page 138.) To run a DIN constancy test 1 Open the CXtra Services menu and select DIN Test. 2 Click Run Constancy Test to open the test selection dialog.
5 After completing the last wizard dialog, click Finish. 6 In the Test Complete dialog that appears, type your name in the Name field and enter the name and location of the organization requesting the test. 7 Click Save Report. NOTE: Click Save As to store the report as an RTF file at a specified location on your hard drive. In this format, you can use common word-processing applications such Microsoft Word or Wordpad to view the report, or attach the file to e-mail transmissions.
Viewing DIN Test Results When you click Save Report after running a DIN acceptance or constancy test, the results are recorded in the history log. You can use the Reporting service to view the results at any time with the following steps: 1 Open the CXtra Services menu and select Reporting. 2 Click the History tab. 3 Click the line corresponding to the test results you want to view, and then click View Detail.
AAPM TG18 Test Service The AAPM TG18 Test service supports the testing of Dome Ex, Dome GX4MP, and Dome Sx displays in accordance with the recommendations of the American Association of Physicists in Medicine Task Group 18. The evaluations supported by this service are fully described in the AAPM TG18 document. Obtain a copy of the document at this URLS: http://deckard.mc.duke.edu/~samei/tg18. If you have the Privilege service installed, you must log on as a privileged user to use this service.
WARNING: Successful completion of a TG18 Test frequently requires preparing the display and the environment in accordance with applicable sections of the TG18 document. The wizard used by the TG18 Test for each evaluation provides only an outline of the basic test steps. It does not describe all prerequisite conditions and detailed information required to perform the evaluation. You can define the test that you want to complete for acceptance, plus routine daily, monthly/quarterly, and annual tests.
AAPM TG18 Test Setup You must complete at least part of the setup procedure before you use the TG18 Test to run a test on any display. During initial setup, specify a number of operating parameters within these categories on the Setup Information dialog: • Photometers • Configuration • Location • Displays • Reminders Review and change the operating parameters at any time to meet changes in equipment or required testing.
2 Click Setup. The Photometers tab of the TG18 Test Setup Information dialog appears. To set up entries on the Photometers tab 1 If you have a supported photometer installed, select the model name on the Photometer pull-down list. Otherwise, select Other. If you select a supported photometer, the system automatically lists the manufacturer’s name in the field provided. 2 If you selected Other, type the manufacturer’s name. 3 Enter the photometer serial number.
To set up entries on the Configuration tab 1 Open the Configuration tab to view and modify default evaluations for each test listed. You can replace the defaults for any test with any combination of listed evaluations. The number sequence following each evaluation name is the paragraph number that describes the evaluation in the TG18 document. 2 Select a test in the Select Test to Configure list.
• Click Select/Deselect All to toggle check marks for all evaluations. Click Restore Defaults to replace current selections with the AAPM Task Group recommended evaluations. Before replacing current selections, the system shows a warning message. You can proceed to replace current evaluation selections. CAUTION: Clicking Restore Defaults replaces custom evaluation selections for all four tests, not just the currently selected test. 3 Repeat steps 1 and 2 for each test. 4 Click Apply.
To set up entries on the Displays tab Review the characteristics of each display panel connected to the computer. You can indicate how each panel is used by selecting a performance class as described below. 1 Click the Displays tab. 2 Select a display in the Display pull-down list. The system supplies values for the attributes shown, if known. 3 Designate the selected display as Primary or Secondary performance class by clicking the appropriate radio button.
To set up entries on the Reminders tab Schedule optional reminders for completing tests other than acceptance. For example, if you specify a reminder for the daily test suite, the system automatically generates a reminder message for each display that you have not tested. Dome CXtra provides reminders for only those tests that are enabled. The routine tests (daily, monthly, annual) are not enabled until an acceptance test has been completed for a given display and the report saved for a given display.
3 Select the type of reminder message you want to receive — Informational or Warning. Selecting the Informational reminder adds a reminder event to the History log for each display at the intervals specified. Selecting the Warning reminder causes a message box to appear on your screen. In addition, the reminder event changes the Dome CXtra icon in the system tray and adds an entry into the History log for each display. If the reminder is issued as a warning, a status change trap is sent to the SNMP console.
To run a test 1 Open the CXtra Services menu and select AAPM TG18 Test. 2 Select a display panel in the Select Display list. 3 Select a test in the Select Test list box. 4 Click Run Test to access the wizard, which prompts you to complete each evaluation in the test. Where necessary, you enter measurement or evaluation data as you complete each evaluation. NOTE: Some evaluations require you to complete multiple wizard dialogs.
Viewing AAPM TG18 Test Results You can view the test results for a TG18 Test immediately after completing all evaluations. The Test Info dialog appears upon completion of a test. Enter your name in the Tester’s Name field and your organization name in the Tester’s Company field. Then pick among these options for viewing or saving the report: • Click View Report for a summary of evaluations completed. • Click Save Report to store the information.
When you click Save Report, Dome CXtra records the test event in the History log. You can then use the Reporting service to view events in the History log. Select the TG18 Test event in the log and click View Detail to review saved test results.
Privilege Service The Privilege service allows only privileged users and administrators to change the operating parameters of the display system. All others, called non-privileged users, are limited to viewing status information and the current operating parameter values. Logging On as Privileged User/Administrator NOTE: Passwords are case-insensitive. You can change the password by following the instructions on page 139.
4 To start Dome CXtra in the privileged user mode when the computer starts and without entering a password, click the “Start as privileged user” box. To start Dome CXtra in the non-privileged user mode when the computer starts, leave the box unchecked. 5 To allow non-privileged users to exit from the Dome CXtra application, click the “Allow non-privileged users to exit” box. 6 Click OK to close the dialog.
Reporting Service The Reporting service records and displays properties and events related to CXtra services installed for your display system. Properties include information about which services and versions are installed. Events are status messages transmitted at regular intervals by each CXtra service. Events are recorded in the History log. The log contains routine events and error messages and events useful in troubleshooting system error conditions.
Properties Tab The Properties tab lists the CXtra services installed and information about the display system components and operating parameters. Click the plus sign (+) beside any heading to display subordinate items containing detailed information. Check the display properties under the RightLight service to see both the current and the maximum supported bits per pixel.
History Tab The History tab displays events recorded in the History event log. The events appearing in the History log are dependent on each service and typically record changes that may be useful for capturing history or troubleshooting an error condition.
Sorting the History Event Log Each event is represented by a row of information with data shown in the following columns: Status. An icon indicating the event type: Information, Success, System Error, or Warning. Service. The CXtra service associated with the event. Date. Record of date and time of the event. Summary. A brief event description. Click a column heading to sort the event list based on the contents of that column.
Using CXtra Preference With the CXtra Preference dialog you can specify the retention period for event information stored in the History log, suppress all or some alerts, and disable the beeping sound. To view or change the retention period Open the CXtra Services menu and select CXtra Preference. Click Retain to save events for the specified number of days. Events older than the specified number of days are deleted unless Dome CXtra is in a warning or error status.
NOTE: If you are using Dome CXtra in conjunction with Dome Dashboard, you can exercise best practice by limiting history retention to 365 days (one year). To retain event information indefinitely, select Retain Complete History. When this option is selected, event information is never deleted and the History log continues to lengthen as long as the Dome CXtra application is operating.
Enterprise Management Service The Enterprise Management service allows you to manage medical imaging displays from a central location using Dome Dashboard software or standard Simple Network Management Protocol (SNMP) console applications such as Castle Rock™ SNMPc. You must install the Enterprise Management service on each computer with a flat panel that you want to monitor and then use the information in this chapter to configure the SNMP service running on the computer.
SNMP Management Console You can use Dome Dashboard software or a console application on your computer to obtain the following types of information from each display running the Enterprise Management service in conjunction with the SNMP service: • Hardware identification • Current operating parameters • Test results • Status messages From your management console, you can also reconfigure a display by changing selected operating parameters.
Configuring the SNMP Service Use this section to configure the SNMP service on a computer running the Enterprise Management service. Be sure the SNMP service is installed and operating on your computer. To confirm, open the Windows Control Panel, select Administrative Tools > Services and make sure SNMP service appears in the list of services. WARNING! SNMP Service Configuration Community names are case sensitive and must be identical with the case used for the Dome Dashboard client settings.
To open the Service Properties dialog 1 Open Control Panel > Administrative Tools > Services to see a list of installed services. 2 Right-click the SNMP service name and select Properties on the popup menu to open the General tab of the SNMP Service Properties dialog.
To enter traps information 1 Click the Traps tab in the SNMP Service Properties dialog. 2 Create the community name used by the SNMP console application. Type that name in the Community name field then click Add to List. (NOTE: Community names are case sensitive. Be sure the community name matches exactly the SNMP trap community entered in the Dashboard Edit Workstation dialog > Network Configuration tab.) 3 Create a list of trap destinations. Click Add to open an SNMP Service Configuration dialog.
In most cases you should enter the identifier for the host running your SNMP console application as one list item. You may also run other trap monitoring applications. Dome Dashboard users can define one or more consoles as trap destinations. 5 Click Apply to save your entries on the Traps tab. To enter security information 1 Click the Security tab in the SNMP Service Properties dialog.
3 Select READ CREATE in the Community rights drop-down list. This allows the Dome Dashboard or SNMP console application to read and set Dome CXtra parameters on this computer. 4 Type the community name used by your SNMP console application in the Community Name field and click Add. (NOTE: You can enter different community names for read, read/write, and read/create, but using the same name is recommended.) 5 Choose either Accept SNMP packets from any host or Accept SNMP packets from these hosts.
Display Agent Traps Systems running the Enterprise Management service for Dome CXtra generate only the following traps: • trapStatusObserverConnectionFailed • trapStatusChanged If the local Dome CXtra system status changes to Warning or Failure, the display agent sends a trapStatusChanged trap. For example, if an external application changes the calibration from DICOM Calibration to Original Calibration, the DICal service changes the Dome CXtra system status to Warning and sends the trapStatusChanged trap.
Available Traps Item Description trapStatusObserverConnection- Status observer connection failed. Failed Data Items: historyStatusStatus, historyStatusRecord, historyStatusServiceTest, historyStatusDateTime, historyStatusSummary trapStatusChanged Status changed. Data Items:(none) trapFPConnectionFailed (obsolete in version 3.0) Connection to RightLight service failed. (obsolete in version 3.0) Flat-panel white point cannot be achieved. trapFPLUMTest Flat-panel LUM test failed.
Troubleshooting Problem Possible Cause Solution Blue screen crashes. System instability. Install the BIOS, firmware, and driver upgrades provided by the manufacturer of your hardware. Displays dropping out or changing resolution unexpectedly. Dome CXtra not starting properly (such as showing the clock icon continuously). Display Assignment. If a different display is selected as the primary, CXtra must be restarted. (NOTE: Applies to all operating systems.) Fast User Switching.
Appendix Silent Install The servicesToInstall file specifies which services get installed during a CXtra installation. If this file is present in the same folder as the setup.exe of the CXtra package, any selections or deselections that the user makes with the select/deselect feature will be ignored. Components not listed below are always installed. Starting with CXtra 6.2.2, SNMP service configuration for Dome Dashboard is supported.
Service Format [,
If is RightLight, the options shown in the following table are allowed.
Example #2: Install Selected Services Install TestPatterns, Privilege, SNMP, and RightLight services, setting the number of measurements during conformance, no popups, Dome Dashboard Server and Community, and the ambient brightness: BEGIN FILE... rightlight, ambient, low, measurements, 32, dynamicrange, equalize, tolerance, 10, targetwhitelevel, 500, disablepopups, disablerightlightpopups testpatterns snmp, dashboardserver, ndsddv2, dashboardcommunity, ddv2community privilege ...
Index A AAPM TG18 test patterns 96 acceptance test 126 alerts auto 25 manual 27 manual/scheduled 23 suppressing 144 Ambient tab, RightLight 21 auto alert parameters, setting 23, 25 B Backlight Saver service, description 108 Backlight tab, RightLight configuration 18 brightness 18 C calibration 15, 20 custom DICOM 85 events, reviewing 87 Calibration tab, RightLight configuration 20 characterizing display 94 checking current conformance status 35 cleaning screen 56 closing Dome CXtra 14 Color tab, RightLi
D, cont.
P, cont.
S, cont.
NDS Surgical Imaging, LLC 5750 Hellyer Avenue San Jose CA 95138-1000 USA www.ndssi.